核磁共振(Nuclar Magnetic Resonance) 在物理和化學領域的基礎研究上,提供了一 項無可替代的利器;核磁共振譜(Magnetic Resonance Spectroscopy) 在臨床醫學上
有著不可忽視的潛力,而核磁共振影像(Magnetic Resonancr Imaging)在臨床診斷上
更是逐漸普及而佔有一席之地。
本論文研究的目的主要在於發展圓極式(Quadrature, Circular Paolarization)的面
線圈(Surface Coil),以改善磁振影像或磁振譜的品質、提高信號對雜音比(Signal
to Noise Ratio) 並評估強化區與線圈尺寸之關係;同時可依需要設計成平面式,如
:脊椎線圈、眼部線圈,或者是圓弧式,如:膝部線圈、頸部線圈,較以往蝴蠂式
(Butterfly) 表面線圈更能適用於人體上不同部位之曲線。
圓極式線圈在原理上即替代了兩個互相垂直的線性線圈,本研究之圓極式線圈便是一
個線圈能夠同時接收兩垂直方向之變動磁場,在效果上,就相當於兩個垂直放置的線
性線圈。一般臨床診斷上為了得到較佳的影像品質,常需要花費較長的時間,將所得
之影像資料相加,才能提高信號對雜音比,而圓極式線圈在較短的掃描時間內(在相
同的參數及條件下)可以得到相同的信號對雜音比,如此節省掃描病人的時間,增加
掃描儀的使用率或在相同的掃描時間內改善畫質,以提高診斷的精確度。
在測試過程中,發現本論文之線圈尚可選擇適當的視界區(FOV,Field of View) ,視
界區變小,雜音自然就小,便可以特別加強此部位之信號對雜音比,得到更清晰的影
像,以利診斷。因研究的是圓極式之表面線圈,所以其信號對雜音比加強區並未涵蓋
整個視界區,但在接近線圈部份則有更高之信號對雜音比,可強調接近表面之組織,
而更符合了表面線圈所應具備之特性。
Magnetic resonance imaging (MRI) have been using prevailingly in medical
applications. The better quality of information from patients the higher
precision of diagnosis we can get. Radio-freqency (RF) coils take the jobs of
transimitter (transmit the radio-frequency electromagnetic wave to excite
spins) and receiver (receive the electromotive force from excited spins).
Improving the transmitting and receiving quality of RF coil is the direct and
easy way to improve information quality which is mainly affected by signal-to-
noise ratio (SNR),resolution,and contrast.
  Surface coil has much higher SNR than volume coil for superficial tissues,
but worse for the deep-seated tissues. The quadrature volume coil has 1.41
times SNR compared with linear one in almost the whole sensitive volume. Since
the magnetic field produced by the surface coil decays drastically moving
aways from the coil body, the quadrature surface coil is more complex
regarding its receiver ability and sensitive region of horizontal and vertical
signal due to its asymmetry.
  Both flat and curvature quadrature surface coils have been designed and
tested. The SNR enhancement (in percentage) is measured and calculated at the
different distance from coil center. Quadrature surface coil provides better
SNR (up to 150%) in the superficial region. Together with the right choice of
the size and shape of a coil body,the high SNR provided by the quadrature coil
serve to shorten the scan time and/or improve information quality, which means
better patient throughput and better diagnosis. The high SNR coil is useful
for magnetic resonance spectroscopy (MRS),too.