電傳導法應用在抽脂術前脂肪厚度測量之研究

Sub-cutaneous fat thickness measured by electrical conductivity before liposuction study

檔案

本條目沒有附加檔案。

類型

語言

中文

摘要

抽脂術流行於1980年代,提供了另一種改善身材曲線的方法。在手術進行前的一項步驟,就是用手或脂肪夾測量皮下脂肪厚度。然而由於每個人脂肪及皮膚的彈性不同,容易造成測量不準確,另外脂肪夾有超胖無法測量的限制,如此測量不準確的結果極易導致手術後遺症的產生。 本文的目的是想應用電傳導的方法來測量脂肪厚度,用以改進以上所提的缺點。我們設計了一個控制系統,利用電腦程式驅動步進馬達,經由傳動線和滾珠螺桿帶動雙極式探測電極刺入皮下組織,其上載有定電流,雙電極不斷前進直到負載電壓有急遽變化,送回授信號到電腦為止。程式計算電極前進且距離,即得此位置脂肪厚度。為了實用上的考慮,程式在執行多點量測後會自動在螢幕顯示區域性脂肪分布圖。
以豬肉作假體實驗,初步結果顯示用電傳導法測脂可行性相當高。我們希望在不久的將來能據此發展出一套有用的工具來幫助醫生手術進行順利。
The introduction of suction assisted lipectomy (SAL) or liposuction in the 1980's, a technique popularized and provided another method to improve body contour in patients. Before liposuction, sub-cutaneous fat thickness was usually measured by skinfold calipers or hands. However, they are likely to be inaccurate since the varying elastic properties of both fat and skin, and there is a limit to the size of the skinfold they can pinch. Such inaccuracy maybe caused some sequelae after liposuction. For the purpose of improving those disadvantages, we tried to apply electrical conductivity to measure fatness. In this study, we designed a control system using the computer to drive a stepping motor, via a transmission line and a ballscrew, bipolar electrodes with a constane current of 4 uA at a frequency of 20 KHz were sent into sub-cutaneous tissue. Bipolar electrodes were advanced until the load voltage showed a large change and then sent a feedback signal to the computer. The computer estimated the forward distance, that can be as the fat thickness of this position. For practical considerations, the graphics of local fat distributed were showed on the monitor of computer, when the measurements of eight points on the curve suctioning were finished.
For a test of pork, primary results showed fat thickness measured by electrical conductivity seems available. We wish to develop a useful tool based on electrical conductivity in th future.
None
顯示完整詮釋資料 統計資料

引用此條目

唐光輝 (1993). 電傳導法應用在抽脂術前脂肪厚度測量之研究. https://thesis.lib.nycu.edu.tw/handle/20.500.14371/17795

BibTeX RIS

引用文字依 APA 第 7 版格式自動產生,使用前請再核對。

分享

永久網址:https://thesis.lib.nycu.edu.tw/handle/20.500.14371/17795