客製化牙冠定位器應用於植體根尖放射線攝影
Customized crown-level position jigs for implant periapical radiography
摘要
植體根尖放射線攝影是植體治療中應用最廣泛的放射線檢查。文獻中發現使用未標準化的植體根尖放射線攝影測量植體周圍骨高度(peri-implant crestal bone level, piCBL)與植體周圍骨缺損深度會有誤差產生。先前為標準化植體根尖放射線攝影製作的植體定位器並不適用於黏著式固持(cement-retained)的植牙牙冠。另外,植體根尖放射線攝影上的植體長度被視為線性測量的校正依據,然而其正確性卻從未被驗證過。為解決以上問題,本研究提出以數位方式設計牙冠定位的植體軸向CLP來標準化植體根尖放射線攝影,並驗證植體長度作為校正依據的正確性。結果指出,沒有使用植體軸向CLP拍攝的植體影像變形量並無法預測,使用植體軸向CLP拍攝所得的植體長度,需依放大倍數進行校正,因為X光為線光源而非點光源。校正後的平均植體周圍骨高度落差(discrepancy of peri-implant crestal bone level, 𝚫D)可達到0.09毫米的正確度。另外發現植體軸向定位器可有效檢測出植體-支台體連接處50微米以上的間隙。
先前的植體贋復物與相鄰自然牙接觸喪失(interproximal contact loss, ICL)的文獻多使用牙線或金屬薄片來判定有無接觸喪失,且盛行率高達33% - 66%。過去文獻對於危險因子的分析多未提及植體贋復物相關因子的集群效應(clustering effect)。再者,植體贋復物與相鄰自然牙ICL的診斷工具也尚未標準化。因此,本研究先檢視固定式植體贋復物回診的根尖放射線攝影以計算植體贋復物與相鄰自然牙ICL於病人、植體贋復物與鄰牙三個層級的的盛行率,使用邏輯式迴歸與廣義估計方程式(generalized estimating equation, GEE)分析病人、植體贋復物與鄰牙三個層級的風險因子。多變項GEE的結果發現,較長的追蹤時間與缺乏牙弓完整性(arch integrity)是影響植體贋復物與相鄰自然牙ICL發生的危險因子,但因分析用的植體根尖放射線攝影並未標準化,所以計算出的盛行率偏低。因此,將牙冠定位器的標的轉至植體贋復物與鄰牙的接觸,製作出客製化接觸CLP。結果發現,在20-30微米的接觸距離下,ICL診斷有八成以上的真陽性率。
由此可知,使用客製化牙冠定位器可進行piCBL的精密測量、50微米的植體-支台體不密合與超過20微米的ICL精密檢測。未來可考慮設計一客製化牙冠定位器同時包含植體軸向與鄰間接觸的指示器(index),又可套用於咬合板(biteplate)上,依特定目的拍攝標準化植體根尖放射線攝影,作為長期追蹤使用。
先前的植體贋復物與相鄰自然牙接觸喪失(interproximal contact loss, ICL)的文獻多使用牙線或金屬薄片來判定有無接觸喪失,且盛行率高達33% - 66%。過去文獻對於危險因子的分析多未提及植體贋復物相關因子的集群效應(clustering effect)。再者,植體贋復物與相鄰自然牙ICL的診斷工具也尚未標準化。因此,本研究先檢視固定式植體贋復物回診的根尖放射線攝影以計算植體贋復物與相鄰自然牙ICL於病人、植體贋復物與鄰牙三個層級的的盛行率,使用邏輯式迴歸與廣義估計方程式(generalized estimating equation, GEE)分析病人、植體贋復物與鄰牙三個層級的風險因子。多變項GEE的結果發現,較長的追蹤時間與缺乏牙弓完整性(arch integrity)是影響植體贋復物與相鄰自然牙ICL發生的危險因子,但因分析用的植體根尖放射線攝影並未標準化,所以計算出的盛行率偏低。因此,將牙冠定位器的標的轉至植體贋復物與鄰牙的接觸,製作出客製化接觸CLP。結果發現,在20-30微米的接觸距離下,ICL診斷有八成以上的真陽性率。
由此可知,使用客製化牙冠定位器可進行piCBL的精密測量、50微米的植體-支台體不密合與超過20微米的ICL精密檢測。未來可考慮設計一客製化牙冠定位器同時包含植體軸向與鄰間接觸的指示器(index),又可套用於咬合板(biteplate)上,依特定目的拍攝標準化植體根尖放射線攝影,作為長期追蹤使用。
Implant periapical radiography was the most widely used radiographic examination in implant dentistry. Errors in measuring peri-implant crestal bone level (piCBL) and depth of peri-implant bony defects on non-standardized implant periapical radiographs were noted in literature review. Previously, fixture-level position jigs were recommended to standardize the implant periapical radiographs, but these were not suitable for cement-retained implant prostheses. Additionally, the implant length on periapical radiographs was viewed as a calibration reference for linear measurements, but its accuracy had never been validated. Therefore, innovative customized crown-level position (CLP) jigs that allow the film holder to be parallel to the implant axis were proposed in this study, and the accuracy of using implant length as a calibration reference for linear measurements was verified. The results revealed that the distorted implant projection on non-standardized implant periapical radiography was unpredictable. The implant length obtained from standardized implant periapical radiography should be calibrated for magnification, because X-ray is a line light source rather than a point one. Using a customized CLP jig to standardize the implant periapical radiographs achieved a discrepancy in piCBL with an accuracy of 0.09 mm after implant length calibration. Besides, implant periapical radiographs taken with utilizing a customized CLP jig were satisfactory for examine the 50μm misfit of implant-abutment connection.
Studies of interproximal contact loss (ICL) associated with implant-supported fixed prostheses (ISFPs) typically used dental floss or metal strips to determine ICL and showed a high prevalence of 33% to 66%. The clustering effect of implant prosthesis-related factors contributing to ICL had seldom been reported. Moreover, the diagnostic tool for ICL between ISFPs and the adjacent teeth had not been standardized. Thus, implant periapical radiographs were chosen to diagnose ICL between ISFPs and adjacent teeth to calculate the prevalence. Patient-, implant prosthesis- and adjacent tooth-related risk factors were analyzed by logistic regression and generalized estimating equation (GEE). Multivariate GEE indicated that a longer follow-up period and factors influencing arch integrity were risky for maintaining stable occlusion of implant prostheses. However, using non-standardized implant radiographs to determine ICL between ISFPs and adjacent tooth would underestimate the condition. Furthermore, a customized contact CLP jig was introduced as an objective diagnostic tool to detect the occurrence of ICL, achieving over 80% true positive rate when applied to evaluate a 20-30μm gap of ICL.
Overall, using a customized CLP jig enabled precise measurement of piCBL, assessment of the 50μm misfit of the implant-abutment connection, and distinction of ICL greater than 20μm as well. In the future, creating an index containing the implant axis and the orientation of proximal contact between the implant prosthesis and t adjacent tooth, and also compatible with the bite plate of Rinn XCP system is suggested. The index can be applied to take standardized implant radiographs for the specific aims in long-term follow-up of implant therapy.
Studies of interproximal contact loss (ICL) associated with implant-supported fixed prostheses (ISFPs) typically used dental floss or metal strips to determine ICL and showed a high prevalence of 33% to 66%. The clustering effect of implant prosthesis-related factors contributing to ICL had seldom been reported. Moreover, the diagnostic tool for ICL between ISFPs and the adjacent teeth had not been standardized. Thus, implant periapical radiographs were chosen to diagnose ICL between ISFPs and adjacent teeth to calculate the prevalence. Patient-, implant prosthesis- and adjacent tooth-related risk factors were analyzed by logistic regression and generalized estimating equation (GEE). Multivariate GEE indicated that a longer follow-up period and factors influencing arch integrity were risky for maintaining stable occlusion of implant prostheses. However, using non-standardized implant radiographs to determine ICL between ISFPs and adjacent tooth would underestimate the condition. Furthermore, a customized contact CLP jig was introduced as an objective diagnostic tool to detect the occurrence of ICL, achieving over 80% true positive rate when applied to evaluate a 20-30μm gap of ICL.
Overall, using a customized CLP jig enabled precise measurement of piCBL, assessment of the 50μm misfit of the implant-abutment connection, and distinction of ICL greater than 20μm as well. In the future, creating an index containing the implant axis and the orientation of proximal contact between the implant prosthesis and t adjacent tooth, and also compatible with the bite plate of Rinn XCP system is suggested. The index can be applied to take standardized implant radiographs for the specific aims in long-term follow-up of implant therapy.