不同的脊椎植入物對腰椎椎體與小面關節的影響

The effects of different spinal instrumentation composition on lumbar vertebral body and facet joint

檔案

日期

2006

類型

語言

中文

摘要

鄰近椎節的不正常受力或活動會造成椎間盤退化,嚴重者可能使椎節坍塌壓迫到神經產生疼痛。治療的方法其一是將坍塌的椎體復位,以椎間融合器搭配補骨植入椎間盤空間促使椎體融合穩定椎節。若腰椎坍塌,治療方法通常會使用後方腰椎融合術。但是,近幾年的臨床研究發現椎間融合器的沉陷與滑移造成融合失敗案例,然機轉尚未明瞭。因此,本研究以有限元素法分析腰椎後方融合術對椎間融合器、小面關節、和後方固定器的馮密西斯應力( von-Mises stress )分布效應,模擬三種不同的手術放置方法,分別為傳統的成對置放、單顆靠右置放和單顆斜向置放,置放的融合器頃角分別為零度和四度。此外也探討後方固定系統所造成的影響,有固定與沒固定的狀況所影響馮密西斯應力轉移的結果。
本研究是以CT 影像重建腰椎( L2~L3 ),依解剖特徵加入周邊韌帶。力量負荷條件是以400N 的預應力施加在腰椎第二節上,模擬椎體受到體重的作用。在腰椎第二節上施以10 N-m 的力矩,以模擬腰椎作前彎、後彎、側彎、和扭轉的動作。腰椎第三節的底部完全固定住。
研究結果指出融合器在前彎時會有較大的馮密西斯應力,且應力集中於融合器與椎體的界面。加上後方固定器後,在不同的動作下其融合器和小面關節的馮密西斯應力會較小,且在不同融合器擺放方式下其融合器的馮密西斯應力值是相近的。而椎體加上後方固定器後,仍發現小面關節的馮密西斯應力會隨著不同的動作而改變,表示小面關節仍會傳遞力量。基於以上力學發現,加上後方固定器後搭配單顆融合器植入於力學表現上是可行的,且手術中盡可能的保留小面關節完整。
Disc degeneration can be related to the result of abnormal intersegmental movement and abnormal stress. Severe disc degeneration may lead to collapse of vertebral segment in adjacent level, and may induce pain if the structure compresses the spinal nerve roots. Hence, an appropriate treatment is to reconstruct physical condition and cage insertion. Cage can stable vertebral segment. Bone graft packaged in the cage can induce bone fusion. If lumber spine collapses, posterior lumbar interbody fusion (PLIF) is usually used as a treatment. However, cage subsidence and migration causing fusion failure were appeared in recent year. The mechanism of subsidence and migration are unclear. Therefore, the purpose of the study was to investigate the effects of PLIF cage positioning that result in von-Mises stress distribution by finite element analysis. Cage positioning were including parallel cages insertion, singular cage right side insertion, and singular oblique cage insertion. 0° and 4° cages were used in the study. Beside cage positioning, the study also was to
investigate the effects of posterior fixator in von-Mises stress distribution. Thus, the results of study were comparing the stress of cage, posterior fixator, and facet joint.
The solid model of lumbar spine functional unit (L2~L3) was reconstructed in this study from computer tomographic scan images. Based on anatomy characteristics, the ligaments ware reconstructed. The loading condition was 400 N preload on the top of L2 vertebral body and 10 N-m in flexion, extension, lateral bending, and torsion on L2 vertebral body respectively. The bottom of L3 vertebral body was completely fixed.
The results showed that higher stress occur between cage and vertebral body interface in flexion then in other motions. Adding posterior fixation on the vertebral can decrease the stress of cage and facet joint while loaded, and also result in the similar stress on the cage in different cage position. And the stress on the facet joint changed in different vertebral motion. Hence, the load could transmit through the facet joint.
Based on the results, we think that adding posterior fixation with singular cage insertion may be feasible in mechanics, and reserving facet joint as possible during
surgery is suggested.