胸腰椎爆裂性骨折固定方式之有限元素法分析
Finite Element Analysis of Fixation Methods of Thoracolumbar Burst Fracture
摘要
臨床上對於胸腰椎骨折的處理,通常會以後方內固定器給予受傷的椎體進行固定,以求不穩定之椎體在術後初期可以藉由內固定器系統得到穩定的效果,增進骨融合的發生。但是臨床上常發現兩節的固定方式,植入物有鬆脫而導致椎體崩塌等失敗的併發症產生。似乎兩節式固定提供的穩定度不足以預防椎體的崩塌。因此,本研究以有限元素法來分析探討三節或是四節的固定方式對於預防椎體崩塌的效果。
本研究以電腦斷層掃瞄片(3mm)獲得胸腰椎幾何外型,並以ANSYS 8.1有限元素軟體建構T11-L3五節完整胸腰椎模型,包括韌帶、椎間盤、小面關節等脊椎元件,並與文獻數據驗證模型正確性。此外,在L1椎節處,以椎體切除術的方式模擬嚴重爆裂性骨折的發生,並以樑元素模擬內固定器系統,匯入有限元素模型中完成三種固定模型的建立並分析之。四種不同固定方式之模型包括兩節式固定(TSF)、上三節式固定(UTSF)、下三節式固定(BTSF)和四節式固定(FSF)。
在正常骨密度的模擬中,TSF模型在各種活動表現下的活動度(ROM)相較其他三組長節式固定皆呈現增加的趨勢。在前彎活動下,TSF的105%高於其他模型的表現(UTSF: 93%; BTSF: 95%; FSF: 79%)。在後彎活動下,TSF的10%略高於其他模型的表現(UTSF: 4%; BTSF: 4%; FSF: 0.5%)。在側向彎曲活動下,TSF的表現是20%大於其他模型 (UTSF: 9%; BTSF: 9%; FSF: 0.8%)。在扭轉活動下,TSF 56%的表現也是高於其他模型 (UTSF: 47%; BTSF: 50%; FSF: 44%)。在骨質疏鬆的模擬情形下,各模型相較於正常骨密度的模擬,皆呈現相同的增加趨勢。在前彎活動下,TSF的表現是109%高於其他模型(UTSF: 100%; BTSF: 100%; FSF: 89%)。在後彎活動下,TSF的13%略高於其他模型的表現(UTSF: 4%; BTSF: 6%; FSF: 1%)。在側向彎曲活動下,TSF的表現是38%大於其他模型 (UTSF: 16%; BTSF: 18%; FSF: 2%)。在扭轉活動下,TSF 64%的表現也是高於其他模型 (UTSF: 58%; BTSF: 61%; FSF: 50%)。此外前方椎體崩塌百分比(%ABC)的結果,四組模型分別為68.8%、70.6%、70.6% 和72.5%;在骨質疏鬆模型下,四組模型分別為66.6%、70.3%、70.3% 和 72.5%。
根據本研究的結果指出,在各種運動情形下,UTSF、BTSF和FSF的活動皆較TSF來得小,顯示長節式的固定對於相對於短節式固定,提供脊柱較大的穩定。尤其在前彎運動,不同骨密度處理下TSF的活動度皆超過正常胸腰椎活動度的表現,顯示兩節式的固定對於處理胸腰椎爆裂性骨折似乎不理想。相對於TSF模型,長節的固定方式因為多添加了一至兩節的固定裝置,而改變了結構的勁度值,增加的結構勁度將提高脊柱的穩定度。雖然FSF固定方式在各種活動表現下支活動度皆為最小,但是前方高度的壓縮比與其他長節式固定並沒有顯著的差異,顯示三組長節式固定對於預防椎體高度崩塌的效果是相近的。本研究也討論骨密度差異的影響,當骨密度減少至0.08 g/cm3去模擬骨質疏鬆的發生,四組模型的活動度相較於正常骨密度皆顯示增加的趨勢。
對於爆裂性骨折的處理,長節式的固定較兩節式固定來得理想,尤其是四節式的固定,因為兩節式固定的強度對於預防椎體的崩塌是不夠的。
本研究以電腦斷層掃瞄片(3mm)獲得胸腰椎幾何外型,並以ANSYS 8.1有限元素軟體建構T11-L3五節完整胸腰椎模型,包括韌帶、椎間盤、小面關節等脊椎元件,並與文獻數據驗證模型正確性。此外,在L1椎節處,以椎體切除術的方式模擬嚴重爆裂性骨折的發生,並以樑元素模擬內固定器系統,匯入有限元素模型中完成三種固定模型的建立並分析之。四種不同固定方式之模型包括兩節式固定(TSF)、上三節式固定(UTSF)、下三節式固定(BTSF)和四節式固定(FSF)。
在正常骨密度的模擬中,TSF模型在各種活動表現下的活動度(ROM)相較其他三組長節式固定皆呈現增加的趨勢。在前彎活動下,TSF的105%高於其他模型的表現(UTSF: 93%; BTSF: 95%; FSF: 79%)。在後彎活動下,TSF的10%略高於其他模型的表現(UTSF: 4%; BTSF: 4%; FSF: 0.5%)。在側向彎曲活動下,TSF的表現是20%大於其他模型 (UTSF: 9%; BTSF: 9%; FSF: 0.8%)。在扭轉活動下,TSF 56%的表現也是高於其他模型 (UTSF: 47%; BTSF: 50%; FSF: 44%)。在骨質疏鬆的模擬情形下,各模型相較於正常骨密度的模擬,皆呈現相同的增加趨勢。在前彎活動下,TSF的表現是109%高於其他模型(UTSF: 100%; BTSF: 100%; FSF: 89%)。在後彎活動下,TSF的13%略高於其他模型的表現(UTSF: 4%; BTSF: 6%; FSF: 1%)。在側向彎曲活動下,TSF的表現是38%大於其他模型 (UTSF: 16%; BTSF: 18%; FSF: 2%)。在扭轉活動下,TSF 64%的表現也是高於其他模型 (UTSF: 58%; BTSF: 61%; FSF: 50%)。此外前方椎體崩塌百分比(%ABC)的結果,四組模型分別為68.8%、70.6%、70.6% 和72.5%;在骨質疏鬆模型下,四組模型分別為66.6%、70.3%、70.3% 和 72.5%。
根據本研究的結果指出,在各種運動情形下,UTSF、BTSF和FSF的活動皆較TSF來得小,顯示長節式的固定對於相對於短節式固定,提供脊柱較大的穩定。尤其在前彎運動,不同骨密度處理下TSF的活動度皆超過正常胸腰椎活動度的表現,顯示兩節式的固定對於處理胸腰椎爆裂性骨折似乎不理想。相對於TSF模型,長節的固定方式因為多添加了一至兩節的固定裝置,而改變了結構的勁度值,增加的結構勁度將提高脊柱的穩定度。雖然FSF固定方式在各種活動表現下支活動度皆為最小,但是前方高度的壓縮比與其他長節式固定並沒有顯著的差異,顯示三組長節式固定對於預防椎體高度崩塌的效果是相近的。本研究也討論骨密度差異的影響,當骨密度減少至0.08 g/cm3去模擬骨質疏鬆的發生,四組模型的活動度相較於正常骨密度皆顯示增加的趨勢。
對於爆裂性骨折的處理,長節式的固定較兩節式固定來得理想,尤其是四節式的固定,因為兩節式固定的強度對於預防椎體的崩塌是不夠的。
Pedicle screw fixation has been a standard procedure to treat thoracolumbar burst fracture. The aim of surgery is to achieve a post-operation spinal stabilization and enhance bony fusion. There are good clinical results followed up with pedicle fixation, but some complications as pedicle screw loosening or body collapse were found clinically. Additionally, the major studies in pedicle screw fixation were
employed using cadaver and animal models. Few literatures are associated with finite element methods. The goal of this study is to establish a three-dimensional finite element model of the thoracolumbar (TL) spine T11-L3 and analyze the surgical methods of two-, three- and four segments fixation. In this study, we expect to find the most proper fixation method to manage thoracolumbar burst fracture.
In this study, a 3-D finite element model of TL spine was established that based on the CT images from a 26 y/o male. The software Image J 1.31V (Wayne Rasband National Institutes of Health, USA) was used to determine the geometry of the TL spine and built by the finite element software ANSYS8.1 (Swanson Analysis System Inc., Houston, TX, USA). The model included cancellous bone, cortical bone, posterior element, disc, endplate, and spinal ligament. The material properties were
adopted from previous literatures. After verifying the TL spine model, an L1 corpectomy model was performed to simulate serious burst fracture. The pedicle screws were implanted to the model and to analyze the influence of the different fixation methods. The four models included two segment fixation (TSF), Up-three segment fixation (UTSF), Below-three segment fixation (BTSF) and four segment fixation (FSF).
In normal bone density model, the TSF model increased the range of motion (ROM) with respect to all other fixed model under the different loading conditions. On the flexion motion, the TSF model was 105% greater than other model (UTSF: 93%; BTSF: 95%; FSF: 79%). On the lateral bending motion, the TSF model was 20% also greater than other model (UTSF: 9%; BTSF: 9%; FSF: 0.8%). The TSF model was 10% a little greater than other model (UTSF: 4%; BTSF: 4%; FSF: 0.5%) under the extension motion. On the torsion motion, the TSF model was 56% greater than other model (UTSF: 47%; BTSF: 50%; FSF: 44%). The osteoporosis model was the same tendency to toward the normal bone density model. On the flexion motion, the TSF model was 109% greater than other model (UTSF: 100%; BTSF: 100%; FSF: 89%). On the lateral bending motion, the TSF model was 38% also greater than other model (UTSF: 16%; BTSF: 18%; FSF: 2%). The TSF model was 13% a little greater than other model (UTSF: 4%; BTSF: 6%; FSF: 1%) under the extension motion. On the torsion motion, the TSF model was 64% greater than other model (UTSF: 58%; BTSF: 61%; FSF: 50%). Furthermore, the percentage of anterior body compression (%ABC) with “TSF”, “UTSF”, “BTSF” and “FSF” was 68.8%, 70.6%, 70.6% and 72.5% in the normal bone density model. %ABC with “TSF”, “UTSF”, “BTSF” and “FSF” was 66.6%, 70.3%, 70.3% and 72.5% in the osteoporosis model.
According to the results, the ROM of UTSF, BTSF and FSF were smaller than TSF under various loading condition. The mean was that the methods of long segment fixation provided the better stability for the spinal structure than the short segment fixation. Especially, on the flexion motion, the ROM of TSF were over the intact model in the different bone density models that seems the TSF method was imperfect to manage the thoracolumbar burst fracture. Because of the additional fixation systems were provide the stronger stiffness than TSF method. The additional stiffness provided the unstable spine more stability. Although the ROM of FSF were the smallest, but the %ABC were no significant differences during the long segment fixation that to prevent the collapse were to be similar. The study also to change bone mineral density to simulate the osteoporosis. When decreasing the BMD to 0.08 g/cm3 to simulate the osteoporosis. The ROM of four models were all increased to compare with the normal bone model.
The long segment fixation methods were better than TSF method to manage the thoracolumbar burst fracture, FSF method especially. The TSF method for burst fracture was difficult to prevent body collapse.
employed using cadaver and animal models. Few literatures are associated with finite element methods. The goal of this study is to establish a three-dimensional finite element model of the thoracolumbar (TL) spine T11-L3 and analyze the surgical methods of two-, three- and four segments fixation. In this study, we expect to find the most proper fixation method to manage thoracolumbar burst fracture.
In this study, a 3-D finite element model of TL spine was established that based on the CT images from a 26 y/o male. The software Image J 1.31V (Wayne Rasband National Institutes of Health, USA) was used to determine the geometry of the TL spine and built by the finite element software ANSYS8.1 (Swanson Analysis System Inc., Houston, TX, USA). The model included cancellous bone, cortical bone, posterior element, disc, endplate, and spinal ligament. The material properties were
adopted from previous literatures. After verifying the TL spine model, an L1 corpectomy model was performed to simulate serious burst fracture. The pedicle screws were implanted to the model and to analyze the influence of the different fixation methods. The four models included two segment fixation (TSF), Up-three segment fixation (UTSF), Below-three segment fixation (BTSF) and four segment fixation (FSF).
In normal bone density model, the TSF model increased the range of motion (ROM) with respect to all other fixed model under the different loading conditions. On the flexion motion, the TSF model was 105% greater than other model (UTSF: 93%; BTSF: 95%; FSF: 79%). On the lateral bending motion, the TSF model was 20% also greater than other model (UTSF: 9%; BTSF: 9%; FSF: 0.8%). The TSF model was 10% a little greater than other model (UTSF: 4%; BTSF: 4%; FSF: 0.5%) under the extension motion. On the torsion motion, the TSF model was 56% greater than other model (UTSF: 47%; BTSF: 50%; FSF: 44%). The osteoporosis model was the same tendency to toward the normal bone density model. On the flexion motion, the TSF model was 109% greater than other model (UTSF: 100%; BTSF: 100%; FSF: 89%). On the lateral bending motion, the TSF model was 38% also greater than other model (UTSF: 16%; BTSF: 18%; FSF: 2%). The TSF model was 13% a little greater than other model (UTSF: 4%; BTSF: 6%; FSF: 1%) under the extension motion. On the torsion motion, the TSF model was 64% greater than other model (UTSF: 58%; BTSF: 61%; FSF: 50%). Furthermore, the percentage of anterior body compression (%ABC) with “TSF”, “UTSF”, “BTSF” and “FSF” was 68.8%, 70.6%, 70.6% and 72.5% in the normal bone density model. %ABC with “TSF”, “UTSF”, “BTSF” and “FSF” was 66.6%, 70.3%, 70.3% and 72.5% in the osteoporosis model.
According to the results, the ROM of UTSF, BTSF and FSF were smaller than TSF under various loading condition. The mean was that the methods of long segment fixation provided the better stability for the spinal structure than the short segment fixation. Especially, on the flexion motion, the ROM of TSF were over the intact model in the different bone density models that seems the TSF method was imperfect to manage the thoracolumbar burst fracture. Because of the additional fixation systems were provide the stronger stiffness than TSF method. The additional stiffness provided the unstable spine more stability. Although the ROM of FSF were the smallest, but the %ABC were no significant differences during the long segment fixation that to prevent the collapse were to be similar. The study also to change bone mineral density to simulate the osteoporosis. When decreasing the BMD to 0.08 g/cm3 to simulate the osteoporosis. The ROM of four models were all increased to compare with the normal bone model.
The long segment fixation methods were better than TSF method to manage the thoracolumbar burst fracture, FSF method especially. The TSF method for burst fracture was difficult to prevent body collapse.