居家型心電圖診斷系統之發展
Development of a Electrocardiogram Interpretation System in Homecare
摘要
本論文主要是配合由國家衛生研究院的盧智成博士所開發出的心電圖機之功能,針對心電圖機的特性而發展一套心電圖即時診斷程式。心電圖機可以提供連續性的完整十二導程訊號及與個人電腦連接的特性,利用這個特性開發其專用的診斷分析程式,達到輕巧化、便利化。本論文也針對行動照護而發展一個可攜式的單一導程心電圖系統,透過無線傳輸將訊號送到智慧型行動電話或個人電腦做分析,即時診斷出心律不整的訊號,即透過電子郵件傳送出行動電話簡訊做緊急通報。
驗証本系統的準確性,比較在相同時間下與市售心電圖機所做出來的診斷結果。實驗是在臨床上及心電圖模擬器上取得90筆訊號,並用GE MAC 5000的結果為對照組。在量測訊號參數(心跳、QRS寬度、PR長度、QT間期和QTc間期)上,使用配對t檢定的結果顯示實驗組與對照組是沒有顯著差異(p<0.05)。在心律不整診斷上,在90筆訊號中挑選出47個有心律不整的病例來做分析。在47例中共有九種不同的心律不整訊號做診斷,正確性達87.5%以上,顯示本系統可應用於臨床或居家使用。
驗証本系統的準確性,比較在相同時間下與市售心電圖機所做出來的診斷結果。實驗是在臨床上及心電圖模擬器上取得90筆訊號,並用GE MAC 5000的結果為對照組。在量測訊號參數(心跳、QRS寬度、PR長度、QT間期和QTc間期)上,使用配對t檢定的結果顯示實驗組與對照組是沒有顯著差異(p<0.05)。在心律不整診斷上,在90筆訊號中挑選出47個有心律不整的病例來做分析。在47例中共有九種不同的心律不整訊號做診斷,正確性達87.5%以上,顯示本系統可應用於臨床或居家使用。
The aim of this study was to develop a real-time ECG analysis program. This program cope with the 12-lead ECG instrument designed by Dr. Chih-Cheng Lu in National Health Research Institutes. This instrument provided continuous and fully 12-lead ECG raw data. We developed an analyzing program to perform real-time measurement and interpretation in PC with Lu’s instrument. In addition, we developed a wireless single lead ECG system, which consists of wireless ECG sensor and real-time measurement program. This system can continuously monitor the senior citizens in the homecare region. The family will receive an alarm message via SMS (Short Message Service) when the program detects certain emergency signal.
In order to verify the analysis outcome, we had compared our results with commercial 12-ECG instrument (GE MAC 5000). We acquired data from clinics and ECG simulators to verify the program. In the measurement of vital parameters such as heart-rate, PR interval, QRS duration, QT interval, and QTc interval, there was no significant difference in the result between the GE MAC 5000 and present program. In the interpretation of arrhythmias, we tested nine kinds of arrhythmias, and the accuracy reached 87.5%. Thus, the proposed ECG analysis program was useful for the clinical diagnosis and homecare application.
In order to verify the analysis outcome, we had compared our results with commercial 12-ECG instrument (GE MAC 5000). We acquired data from clinics and ECG simulators to verify the program. In the measurement of vital parameters such as heart-rate, PR interval, QRS duration, QT interval, and QTc interval, there was no significant difference in the result between the GE MAC 5000 and present program. In the interpretation of arrhythmias, we tested nine kinds of arrhythmias, and the accuracy reached 87.5%. Thus, the proposed ECG analysis program was useful for the clinical diagnosis and homecare application.