利用正子電腦斷層掃描於小動物造影之特性評估

Feasibility Study to the Use of Clinical PET/CT on Small Animal Imaging

檔案

日期

2006

類型

語言

中文

摘要

微型正子電腦斷層掃描儀(microPET)主要是針對小動物(小鼠)體型所設計,學理上與臨床正子電腦斷層掃描儀(PET)比較應會得到較好的空間解析度(spatial resolution)與靈敏度(sensitivity)。目前結合正子電腦斷層與X光電腦斷層的掃描儀(PET/CT dual-modality)於臨床使用上已相當的成熟,二種掃描儀的結合除了能提供簡單且精確的影像融合平台,使PET像上具有放射活性聚積處,能藉由x-ray CT精細的解剖資訊定位出來外,還能提供快速且良好的衰減修正資訊(attenuation correction) 提升PET影像量化的精準度,但小動物(小鼠)體型的這類掃描儀目前大都還在研發中,此外,臨床用PET/CT的大照野(field of view)在小動物研究運用上卻也提供額外的優勢;即多隻小動物同時進行掃描。由上述諸多優勢可知,目前臨床用PET/CT掃描儀於小動物的研究上是可行且具有潛力的。本研究即利用目前臨床用PET/CT掃描儀進行小動物造影之特性評估。掃描模式可分為:單一小鼠高解析度掃描模式(single high-resolution mode, SHR)與多數小鼠高效率掃描模式(multi high-throughput mode, MHT)兩種;評估及量化其影像品質並與microPET的影像作比較。結果發現,雖然microPET在各項影像評估的表現上皆稍優於臨床的正子電腦斷層掃描機,但並非有非常顯著的差異。再者,經由實驗也發現使用x-ray CT做衰減修正後,確實能改善PET影像均勻度並提升藥物量化的精確度。證實目前臨床用PET/CT掃描儀於小動物的研究上是可行且具有潛力的。
The feasibility study to the use of clinical positron emission tomography/computed tomography (PET/CT) scanner on small animal imaging with [F-18]-fluoro-2-deoxy-D-glucose (FDG) was evaluated. In this study, we proposed several CT imaging protocols in the PET/CT system to investigate it capacities and also explore the performance differences between the microPET and the clinical PET/CT. Two CT protocols in the PET/CT system, high-resolution mode protocol specialized for single mouse, and high throughput mode protocol for multiple mice imaging, were employed. Finally, the images were evaluated based on resolution, uniformity.
In this study, we have found that the overall image quality from the microPET scanner is superior to those from PET/CT scanner but the difference is not quite significant. The image quality under high throughput mode protocol was as well as those under high-resolution mode protocol, suggesting that high-throughput studies may be feasible with the reliability we demand. We also observed that the CT-based attenuation correction can improve the image uniformity characteristics either in single-mouse or multi-mice protocol setting.
In summary, although the clinical PET/CT scanner could not compete with the microPET scanner, especially in spatial resolution, the combined CT could bring about several possibilities in research application, such as high-throughput studies. In addition, the precise fusion of PET images with the CT images facilitates anatomic localization of the PET findings, overcoming alignment problems, which is often encountered with techniques that register images obtained from two separate systems. Finally, considering all advantages, the clinical PET/CT imaging might be a potential tool for small animal study.
顯示完整詮釋資料 統計資料

引用此條目

王澤景 (2006). 利用正子電腦斷層掃描於小動物造影之特性評估. https://thesis.lib.nycu.edu.tw/handle/20.500.14371/33233

BibTeX RIS

引用文字依 APA 第 7 版格式自動產生,使用前請再核對。

分享

永久網址:https://thesis.lib.nycu.edu.tw/handle/20.500.14371/33233