摘要

Breathing is one of the essential functions in human life, and thus respiratory monitoring has always been an important topic in the medical and health management fields. Our team developed a wearable respiratory monitoring ultrasound patch that can be attached to the patient’s chest between the ribs. It can identify the patient’s breathing is normal or not by detecting the respiratory movement through the pleura. This device can be used as respiratory monitoring during surgery, painless endoscopy, sleep apnea monitoring. Among other applications in medical monitoring and daily heath management fields.
In the early stages of the research, the wearable ultrasound patch successfully collected the RF-Mode signals of the pleura. After transformation, these signals were converted into M-Mode images, allowing observation of the patients’ actual respiratory condition in real-time. Due to the lack of a successful method to attach the patch to the skin in previous research on ultrasound probes, it is still necessary for the subjects to hold the probe manually during the measurement. Therefore, the breakthroughs of this study include: the successful attachment of the patch-type ultrasound probe to the intercostal skin, the ability to capture clear pleural waveforms, and the capability to identify normal and abnormal respiration patterns through the monitoring system. It is expected that the patch-type ultrasound probe can move towards becoming a professional clinical medical device.
In this study, we improved the ultrasonic patch manufacturing process and changed it from a hand-held type to an attached type, which can be closer to the clinical measurement method. Through the human subject research in school, we aim at each measurement point under different postures, through (1) completion of short-term attachment and finding the pleural signal, (2) obvious brightness of the pleural signal, (3) regularity of the pleural signal wave, etc.
It is observed that when the ultrasonic patch is attached to the chest, it is easier to measure good pleural signals at the measurement point close to the parasternal line in the lying posture. The right side T1-T2 and left side T2- T3 has a 100% measurement success rate. In side lying posture measurement, when the patch is attached to the chest, it is easier to measure a good pleural signal at the measurement point close to the anterior axillary line, and the measurement success rate of this area is 90% on both the left and right sides. Beside the film is attached to the back, the T2-T3 between the paraspinal line and the scapular line has the best measurement results, and this area has an 80% measurement success rate on the left and right sides. In addition, when the patch is placed on the back in a prone position, the best signal results can be obtained at the measurement point parallel to the spine of the scapula T3-T4, and there is a 90% measurement success rate at this measurement point.
呼吸是生命中不可或缺的基本功能之一,因此呼吸監測一直都是醫學及健康管理領域的重要課題。本研究開發了一種穿戴式呼吸監測超音波貼片,將其貼附於患者的胸前肋骨間隙,即可透過肋膜的呼吸變化辨別患者當下的呼吸運動是否正常。呼吸監測裝置可以做為手術中、無痛內視鏡檢查過程的呼吸監測以及監測睡眠呼吸中止等等,廣泛應用於醫療監測和日常健康管理相關領域。
研究初期,超音波貼片可成功紀錄肋膜的Radio Frequency-Mode (RF-Mode)訊號,再經過程式進行後處理而轉換成容易判讀的灰階Motion-Mode (M-Mode)影像,將接收訊號之儀器連結到顯示面板上就可以即時觀測病患肺部的實際呼吸狀況。前期研究之超音波貼片尚未研究出能夠將其順利貼附於皮膚的方式,仍需讓受測者以手持的方式進行量測,因此本研究的技術突破分別為:(1) 超音波貼片能夠良好貼附在肋骨間隙皮膚上、(2) 能夠10秒內方便快速的貼附使用、(3) 能夠監測到清晰的肋膜波形以及能夠透過監測系統辨識出正常與異常呼吸等三大目標,期望超音波貼片能朝向專業臨床醫療器材邁進。
本研究中,我們改善超音波貼片製程,將其由手持式改進為貼附式,能更貼近臨床量測方式。藉由校內人體試驗我們針對不同姿勢下的每個量測點,透過(1)短時間貼附完成並找到肋膜訊號、(2) 肋膜訊號亮度明顯、(3) 肋膜訊號波行規律等條件進行評分量化。
觀察到超音波貼片貼附於胸前時,正躺姿勢量測下,接近胸骨旁線的量測點更容易測得良好的肋膜訊號,該區域的右側T1-T2以及左側T2-T3有100%的量測成功率。在側躺姿勢量測下,貼片貼於胸前時,接近腋前線的量測點更容易測得良好的肋膜訊號,該區域在左、右側都有90%的量測成功率;貼片貼附於背部時,脊柱旁線與肩胛線之間的T2-T3有最好的量測結果,該區域在左、右側都有70%的量測成功率。而貼片貼於背部呈俯臥姿勢能夠在平行於肩胛岡T3-T4的量測點有最好的訊號結果,在該量測點有90%的量測成功率。