Accuray G3 電腦刀之劑量驗證

Dose verification of Accuray G3 Cyberknife treatment system

檔案

日期

2007

類型

語言

中文

摘要

本研究之主要目的是利用蒙地卡羅模擬程式建立電腦刀機頭結構模擬,藉由百分深度劑量曲線和側向劑量曲線之蒙地卡羅模擬結果與實際量測結果進行比對,找出最佳電子初始入射參數,以利今後電腦刀給予劑量的模擬計算。同時,利用照射發光玻璃劑量計、熱發光劑量計以及EDR2膠片劑量計進行相對輸出因子之量測,以驗證電腦刀模擬之正確性,此外,本研究中亦會模擬電腦刀實際治療程序進行擬人形假體劑量量測,評估劑量量測值與電腦治療計畫系統之計算值進行比較。
本研究完成Acuray G3電腦刀蒙地卡羅模擬程式之建立,所使用之入射電子模擬參數為入射電子射束能量6.0MeV、入射電子射束之寬度2.2mm,百分深度劑量之模擬計算值與量測值,除在深度3毫米處之誤差為5.36%以外,其餘各點之誤差值均小於3%,而側向劑量曲線之模擬計算值與量測值於劑量平坦區域之誤差均小於3%。電腦刀相對輸出因子量測方面,除了最小照野孔徑10mm和5mm外,柯達EDR2軟片量測值有高估之現象以外,其餘照野之量測結果相當符合。擬人形假體劑量量測主要利用照射發光玻璃劑量計進行劑量量測,分析研究結果顯示,照射發光玻璃劑量計量測結果與電腦刀治療計畫系統之劑量計算值之誤差值小於5%。
The aims of this study were to simulate the Cyberknife treatment head. The incident electron energy and full width half maximum were determined by comparing the Monte Carlo calculations with the measured data of the percentage depth dose and lateral dose profiles. The relative output factors of Cyberknife were also measured by the radiophotoluminescent glass dosimeters (GDs), thermoluminescent dosimeters (TLDs) and Kodak EDR2 films. Phantom study was also performed to estimate the difference between the experimental measurements and dose calculations by treatment planning system (TPS).
In this study, the Monte Carlo simulation of Cyberknife was established. The Initial energy and full width half maximum of the incident electron beam were 6.0MeV and 2.2mm. The difference between the Monte Carlo calculations and the measured percentage depth doses were less than 3%, except the location of 3mm in depth. In the flat area of the lateral dose profile, the difference between the Monte Carlo calculations and the measured data were less than 3%. The relative output factors measured by Kodak EDR2 films, GDs, TLDs were all in agreement , except for the area with the two smallest collimators. The relative output factors for 7.5 and 5 mm collimator measured by Kodak EDR2 films were overestimated as compared to than GDs and TLDs. In anthropomorphic phantom study, the difference between the experimental measurements and TPS calculations was within 5%.
顯示完整詮釋資料 統計資料

引用此條目

洪照雄 (2007). Accuray G3 電腦刀之劑量驗證. https://thesis.lib.nycu.edu.tw/handle/20.500.14371/35972

BibTeX RIS

引用文字依 APA 第 7 版格式自動產生,使用前請再核對。

分享

永久網址:https://thesis.lib.nycu.edu.tw/handle/20.500.14371/35972