The precision and accuracy of dose measurement by Fricke MRI gel dosimeter

檔案

日期

2008

類型

語言

中文

摘要

目的:近年來放射線治療的技術已有長足的進步,然而如此精確的治療有賴良好的劑量偵測方法來驗證,因此三維空間劑量偵測儀在現今臨床放射線治療就顯得格外重要。本論文探討硫酸亞鐵凝膠(Fricke gel)於輻射劑量量測之精密度與準確度,並研究劑量梯度對輻射劑量量測的影響,進而探討Fricke劑量凝膠於臨床應用之限制。
實驗方法與材料:以一定之流程與成分濃度配置硫酸亞鐵凝膠,放置於自製假體中12公分見方與4公分深之凹槽,以放射線治療計劃系統設計不同之劑量強度與分布之計畫,然後用高能直線加速器依計畫使用六百萬伏特(6MV)能量之光子照射,再以Philip 1.5T磁振造影機於放射線照射之前與之後掃描自製假體擷取影像資料,搭配磁振影像強度法得吸收劑量資料;同時以相同之放射線照射計畫以X光軟片劑量術(Film-based dosimetry)與空氣游離腔(Ionization chamber)偵測吸收劑量,取得標準工具之偵測資料,再分析所有相關資料以探討硫酸亞鐵凝膠(Fricke gel)於輻射劑量量測之精密度與準確度,及劑量梯度對輻射劑量量測的影響。並運用gamma evaluation 比較實際以直線加速器照射之三維空間劑量分布與治療計劃系統設計之劑量分布及鐵離子擴散現象對劑量偵測之影響,進而探討臨床運用之限制。
結果: Fricke劑量凝膠、X光軟片劑量術與空氣游離腔之精密度分別為3.12 % 、1.23 %、0.62 %;Fricke劑量凝膠之準確度為 8%;以Bland-Altman plot 分別比較Fricke劑量凝膠與X光軟片劑量術、空氣游離腔在劑量偵測上差異於2 - 5戈雷以下即有明顯之差異(大於準確度的兩個標準差);Gamma method 以距離3mm及劑量3%為條件下劑量梯度大於2.79Gy/mm、1Gy/mm及0.5Gy/mm分別在照射後45分鐘、90分鐘、120分鐘Gamma 值大於一。
討論:由於Fricke劑量凝膠之精密度( 3.12 %)與準確度( 8 %)相較於X光軟片劑量術(Film-based dosimetry)及空氣游離腔(Ionization chamber)較差,而且在較低之劑量區準確度有明顯的差異,所以Fricke劑量凝膠雖然無法成為標準定量之劑量儀,但可用於定性的偵測實際輻射照射時劑量分布與治療計畫的差異。在臨床上可盡速發現可能有誤差之處,得以進一步的量測。Fricke劑量凝膠於低劑量準確度較差,且高劑量梯度之區域受擴散影響很大,所以一次照射之劑量須大於 5戈雷且於照射後應於 45分鐘內完成磁振造影掃描。
Purpose: Recent advances in radiation treatment technologies, such as intensity modulated radiotherapy (IMRT) and image-guided radiotherapy (IGRT), had provided more delicate and accurate treatment protocol, therefore how to verify the 3-dimensional dose becomes a significant issue. Gel dosimetry is one of the most promising tools to fill the 3D verification void in modern radiation therapy practice. This research is to define the precision and accuracy of dose measurement in Fricke MRI gel dosimeter and evaluate the diffusion of ferrous ion that affects the precision and accuracy of gel dosimeter by gamma evaluation.
Methods & materials:The Fricke-Gel was manufactured and introduced into the phantom. Then the phantom was irradiated with different protocol, planned by radiotherapeutic planning system, and those protocols composed of different dosage. The MRI imaging data of phantom were got before and after irradiation and were related to radiation dose by image-intensity method. In the mean time the dose of irradiation was detected by film-based dosimetry and ionization chamber. These data were investigated and compared for the presicion and accuracy of gel and the effect of dose gradient on diffusion. For diffusion study the images over different time after irradiation were compared by gamma evaluation.
Results: The precision of Fricke gel, film-based dosimeter and ion-chamber is 3.12% ,1.23 % and 0.62% .and the accuracy of Fricke gel is 8 %. As comparison to the data detected from the three methods by Bland-Altman plot there were significant difference under the does of 3Gy . The series images after irradiation were compared by gamma evaluation and the time after irradiation , that gamma value is more than one, with the dose gradient under 0.5 Gy/mm, between 0.5 &1 Gy/mm, and over 2.79Gy/mm is 120minutes, 90minutes, and 45minutes respectively.
Discussion: As the precision of accuracy is much low than ionization chamber, we can use the Fricke-MRI-gel for qualitive evaluation of radiotherapeutic plan. While the dose distribution of radiotherapeutic plan was investigated by Fricke-MRI-gel, the gel should be irradiated with morethan 5 Gy and san by MRI less than 45 minutes after irradiation.