針孔單光子射出電腦斷層於小動物造影之定量研究
Towards Quantitative Small Animal Pinhole SPECT Imaging
摘要
本研究論文之主要目標在於探討小動物分子影像應用中針孔單光子射出斷層造影(pinhole SPECT)之定量成像能力。此系統性研究之重要意義在於能為臨床前(pre-clinical)研究提供深具價值之指導資訊,特別是基於近年來需要使用定量針孔單光子射出斷層造影(quantitative pinhole SPECT)之分子影像研究的數量和種類不斷地增加。目前,已經發表的文獻並無關於此課題之廣泛深入探討,因此,本論文期望填補這一空白,並促進下一階段在此領域之學術研究。
本論文之研究方式主要採用蒙地卡羅模擬(Monte Carlo Simulation, MCS)與實驗方法。首先,我們藉由整合現有之兩種蒙地卡羅模擬工具:SimSET (simulation system for emission tomography)與GATE (Geant4 application for emission tomography) ,開發並驗證一套改良之蒙地卡羅模擬(MCS)工作流程。具體來說,為了保留GATE在模擬複雜的造影儀器構造之能力優勢,同時規避GATE底層Geant4在模擬光子與物質(特別於數位物體模型)作用之相對緩慢的缺點,我們因而於傳統GATE流程中創新插入使用SimSET光子歷史紀錄發生器(photon history generator, PHG)已快速處理光子與物質間之作用。此改良手段不但保留模擬精準度,並提高多達10倍之運算速度。
利用上述改良之電腦模擬工具,我們評估了在進行影像重建修正之前的小動物針孔單光子射出斷層造影(small animal pinhole SPECT)之定量精確度。具體來說,我們選擇會影響影像定量精確度的因素並量化探討其貢獻。此系統性之研究,不但可為目前臨床前實驗中可達到的定量準確度提供深具價值之參考信息,並可用以指導小動物針孔單光子射出斷層造影(small animal pinhole SPECT) 實際流程之設計與執行。
基於上述研究成果,我們確定了會影響影像定量精確度的因素,並將此論文下一步重點擺在探討使用適當影像修正方法,例如發展並評估針對光子衰減,散射,有限造影系統解析度之常用修正方法。基於其改善定量誤差之程度, 我們建議了可行的修正策略以實現有效精確影像定量。
最後,這項研究不僅有助於深入理解定量針孔單光子射出斷層造影(quantitative pinhole SPECT) 之技術,同時也為其研究和使用提供實用指導意見。除了上述創新改良蒙特卡羅模擬(MCS)工作流程,本研究論文還提供其他獨特貢獻,例如,反駁光子散射效應在碘-125 (I-125)針孔單光子射出斷層造影精確定量中可忽略之誤解。
本論文之研究方式主要採用蒙地卡羅模擬(Monte Carlo Simulation, MCS)與實驗方法。首先,我們藉由整合現有之兩種蒙地卡羅模擬工具:SimSET (simulation system for emission tomography)與GATE (Geant4 application for emission tomography) ,開發並驗證一套改良之蒙地卡羅模擬(MCS)工作流程。具體來說,為了保留GATE在模擬複雜的造影儀器構造之能力優勢,同時規避GATE底層Geant4在模擬光子與物質(特別於數位物體模型)作用之相對緩慢的缺點,我們因而於傳統GATE流程中創新插入使用SimSET光子歷史紀錄發生器(photon history generator, PHG)已快速處理光子與物質間之作用。此改良手段不但保留模擬精準度,並提高多達10倍之運算速度。
利用上述改良之電腦模擬工具,我們評估了在進行影像重建修正之前的小動物針孔單光子射出斷層造影(small animal pinhole SPECT)之定量精確度。具體來說,我們選擇會影響影像定量精確度的因素並量化探討其貢獻。此系統性之研究,不但可為目前臨床前實驗中可達到的定量準確度提供深具價值之參考信息,並可用以指導小動物針孔單光子射出斷層造影(small animal pinhole SPECT) 實際流程之設計與執行。
基於上述研究成果,我們確定了會影響影像定量精確度的因素,並將此論文下一步重點擺在探討使用適當影像修正方法,例如發展並評估針對光子衰減,散射,有限造影系統解析度之常用修正方法。基於其改善定量誤差之程度, 我們建議了可行的修正策略以實現有效精確影像定量。
最後,這項研究不僅有助於深入理解定量針孔單光子射出斷層造影(quantitative pinhole SPECT) 之技術,同時也為其研究和使用提供實用指導意見。除了上述創新改良蒙特卡羅模擬(MCS)工作流程,本研究論文還提供其他獨特貢獻,例如,反駁光子散射效應在碘-125 (I-125)針孔單光子射出斷層造影精確定量中可忽略之誤解。
The main goal of this project is to investigate the quantitative imaging capability of pinhole SPECT in the context of small animal molecular imaging. The significance of this endeavor lies in its comprehensiveness and value in providing important guideline for preclinical practice, as the number and variety of molecular imaging studies that require the use of quantitative pinhole SPECT increased dramatically during recent years. To our knowledge, a comprehensive study of this nature has not been published, and the work presented herein could fill this void and facilitate the next stage of quantitative small animal imaging research.
Both Monte-Carlo and experimental methods were used for the study. As an important first step, we developed and validated an improved Monte Carlo simulation (MCS) workflow, which seamlessly integrates two existing MCS tools: simulation system for emission tomography (SimSET) and Geant4 application for emission tomography (GATE). Specifically, we retained the strength of GATE in describing complex collimator/detector configurations to meet the anticipated needs for studying advanced pinhole collimation (e.g., multipinhole) geometry, while inserting the fast SimSET photon history generator (PHG) to circumvent the relatively slow GEANT4 MCS code used by GATE in simulating photon interactions inside voxelized phantoms. While retaining the simulation accuracy, a significant computational speedup (up to ~10-fold) was achieved using the new workflow.
Next, we assessed the quantitation accuracy in small animal pinhole SPECT prior to implementing comprehensive image compensations. Specifically, we identified several image degrading factors and quantified how each of them affects quantitation accuracy by using the aforementioned new SimSET-GATE workflow. This systematic assessment would provide valuable reference information for the achievable quantitation accuracy in practice and guide the design and execution of small animal SPECT imaging protocols.
Subsequently, by quantifying how each individual factor affects quantitation accuracy, we identified the most influential factors and focused on compensating for them to improve the accuracy by adapting, and evaluating the outcome of, appropriate compensation methods, such as those for attenuation, scatter, and degradation caused by limited system resolutions. Practical strategies for achieving accurate quantitation were suggested based on these results.
In summary, this work not only contributed to the better understanding of various aspects in quantitative pinhole SPECT, but also provided a practical workflow for studying and implementing such technique. Other unique contributions were also made during the process, such as the development of the novel MCS method, and the clarification of a misconception regarding the scatter effects in quantitative I-125 small animal SPECT.
Both Monte-Carlo and experimental methods were used for the study. As an important first step, we developed and validated an improved Monte Carlo simulation (MCS) workflow, which seamlessly integrates two existing MCS tools: simulation system for emission tomography (SimSET) and Geant4 application for emission tomography (GATE). Specifically, we retained the strength of GATE in describing complex collimator/detector configurations to meet the anticipated needs for studying advanced pinhole collimation (e.g., multipinhole) geometry, while inserting the fast SimSET photon history generator (PHG) to circumvent the relatively slow GEANT4 MCS code used by GATE in simulating photon interactions inside voxelized phantoms. While retaining the simulation accuracy, a significant computational speedup (up to ~10-fold) was achieved using the new workflow.
Next, we assessed the quantitation accuracy in small animal pinhole SPECT prior to implementing comprehensive image compensations. Specifically, we identified several image degrading factors and quantified how each of them affects quantitation accuracy by using the aforementioned new SimSET-GATE workflow. This systematic assessment would provide valuable reference information for the achievable quantitation accuracy in practice and guide the design and execution of small animal SPECT imaging protocols.
Subsequently, by quantifying how each individual factor affects quantitation accuracy, we identified the most influential factors and focused on compensating for them to improve the accuracy by adapting, and evaluating the outcome of, appropriate compensation methods, such as those for attenuation, scatter, and degradation caused by limited system resolutions. Practical strategies for achieving accurate quantitation were suggested based on these results.
In summary, this work not only contributed to the better understanding of various aspects in quantitative pinhole SPECT, but also provided a practical workflow for studying and implementing such technique. Other unique contributions were also made during the process, such as the development of the novel MCS method, and the clarification of a misconception regarding the scatter effects in quantitative I-125 small animal SPECT.