降低電化學式NH3及H2S氣體感測器受環境因子的影響

Reducing the impact of environmental factors on electrochemical NH3 and H2S gas sensors

檔案

全文將於 2027年8月27日 起公開。

日期

2024

類型

語言

中文

摘要

"Odor" refers to pollutants that can cause disgust or other adverse emotional reactions. It is caused by nitrogen- and sulfur-containing compounds such as ammonia, hydrogen sulfide, methyl mercaptan, and other aldehydes, phenols, fatty acids, and other pollutants. Ammonia and Hydrogen sulfide are common odor pollutants, and they are also harmful pollutants that can cause damage to human health and safety. The emission standards for these pollutants must not exceed 1 ppm and 0.1 ppm, respectively.
Electronic nose systems composed of a variety of low-cost Metal Oxide Semiconductor and Electrochemical gas sensors have been applied to odor sites to identify the location and extent of odor generation. However, the ambient temperature and humidity affect the sensor value easily. This study uses heating and humidity control methods to maintain the relative humidity in the Figaro EC NH3 (FECS44-100) and H2S (FECS50-100) sensor chambers between 40% and 48%, reducing changes in temperature and humidity.
FECS44-100 NH3 sensor heating background value (NH3,L,F,HBG) can be reduced to -0.23 to 0.19 mV through ΔAH (absolute humidity difference) correction, the measured value jump range is 308.6 ppb, LOD is 179.3 ppb. Unheated background value (NH3,F,UHBG) can be reduced to -0.65 to 0.5 mV through MLR-Ta and RHa correction. The measured value jump range is 554.11 ppb, and the LOD is 441.67 ppb. Heating measurement value (NHC3,L,F,H) compared with the unheated measurement value (NHC3,F,UH), the slope increased from 0.72 to 0.86, the intercept decreased from 280.9 ppb to 119.44 ppb, R2 increased from 0.72 to 0.86, RMSE and NRMSE were reduced to 204.53 ppb and 23.94%, and the data quality was improved. The background value of the FECS50-100 H2S sensor is not affected by the ambient temperature and humidity. Compared with the heated measurement value (H2SH) and the unheated measurement value (H2SUH), R2 increased from 0.39 to 0.95, and the intercept dropped from 75.17 ppb to 5.71 ppb, RMSE dropped from 40.25 ppb to 14.83 ppb and NRMSE dropped from 32.42% to 13.7%.
異味指具有氣味,足以引起厭惡或其他不良情緒反應之污染物,由氨氣、硫化氫、甲基硫醇等含氮、含硫化合物和其他醛類、酚類、脂肪酸類等污染物所引起,當中的氨和硫化氫為常見的異味污染物,同時也是對人體的健康安全造成危害的有害污染物,周界的排放標準分別不得超過1 ppm和0.1 ppm。
由多種低成本的MOS(Metal Oxide Semiconductor)和EC(Electrochemical)氣體感測器所組成的電子鼻(electronic noses, e-noses)系統已運用於異味場址,來判別異味生成的位置和有無異味的產生,但感測器測值容易受環境溫溼度影響。本研究使用加熱控溼的方法,維持Figaro EC NH3(FECS44-100)和H2S(FECS50-100)感測器腔體內的相對溼度在40%至48%之間,降低溫溼度的變化。
FECS44-100 NH3感測器加熱背景值(NH3,L,F,HBG)可透過ΔAH(絕對溼度差)校正降低至-0.23至0.19 mV之間,測值跳動範圍為308.6 ppb,LOD為179.3 ppb,未加熱背景值(NH3,F,UHBG)則可透過MLR-Ta、RHa校正降低至-0.65至0.5 mV之間,測值跳動範圍為554.11 ppb,LOD為441.67 ppb。加熱測值(NHC3,L,F,H)與未加熱測值(NHC3,F,UH)相比,斜率由0.72提升為0.86,截距由280.9 ppb降低為119.44 ppb,R2由0.72提升至0.86,RMSE和NRMSE則由397.23 ppb降低至204.53 ppb和39.92%降低至23.94%,數據品質都有所提升。FECS50-100 H2S感測器的背景值不受到環境溫溼度的影響,加熱測值(H2SH)跟未加熱測值(H2SUH)相比 ,R2由0.39提升至0.95,截距由75.17 ppb降至5.71 ppb,RMSE和NRMSE由40.25 ppb和32.42%降至14.83 ppb和13.7%。