動態假體於影像導引放射治療系統重組體積誤差分析
Analysis of Reconstructed Volume Error from Moving Phantom in Image-Guided Radiotherapy System
摘要
摘要
利用錐形射束電腦斷層(cone-beam computed tomography, CBCT)影像做為執行重新計畫設計(re-planning)為前提之目的來評估當器官運動時物體的重組體積之準確性。此實驗評估於不同移動頻率與振幅參數變化下,於醫用直線加速器的CBCT、螺旋刀(TomoTherapyTM)的MV級電腦斷層(megavoltage computed tomography, MVCT)與模擬定位扇狀射束電腦斷層(fan-beam computed tomography, FBCT)影像重組體積間之差異。
分別透過FBCT、Elekta SynergyTM的XVI(X-ray Volumetric Image)系統與MVCT擷取假體影像。靜態與動態實驗中所使用的3個假體直徑分別為5.1、9.9 與12.2 cm之壓克力球。所獲得的CBCT、MVCT與FBCT影像分別傳入治療計畫系統(Pinnacle3 7.6c)中進行影像重組體積比較。透過調整CBCT影像後可讓CBCT之CT值在治療計畫系統中運算更為準確。為了模擬呼吸運動,假體分別以振幅各為7.5、10、12.5、15、17.5與20 mm以及頻率每分鐘各為8、10、12、14、16與20次的條件下進行縱向(superior-inferior direction)運動。
在靜態假體實驗中,對於CBCT與FBCT影像之間的重組體積誤差於3個小、中與大的球假體分別為13.35%、7.47%與5.08%;經過影像調整CBCT後稱modified CBCT(mCBCT)與FBCT之間的重組體積誤差對於小、中、大球假體分別為-6.20%、-2.47%及-2.19%;在MVCT與FBCT之間的重組體積誤差對於小、中、大球假體則分別各為-10.15%、-3.65%及2.77%。不同直徑大小的球假體對於其重組體積的誤差值有明顯的差異,即愈小的球假體其重組體積誤差愈多。
在動態條件的假體實驗中,mCBCT 與MVCT影像也可以發現重組體積損失的現象。隨著振幅值的增加下重組體積誤差愈明顯。但相對地,不同的頻率變化對於重組體積誤差的影響則較不顯著。相同的,對於動態假體的mCBCT 與MVCT影像中愈小的球假體其重組體積誤差愈大。動態假體的mCBCT影像透過挑選(sort)出相同相位的投影(稱mSortCBCT)而進行體積重組,在頻率每分鐘為20次與振幅為7.5mm下所得到大球假體的重組體積與FBCT重組體積之間的誤差為-4.43%。
總結來說,靜態下的假體會因為散射的關係而高估由CBCT影像所得之重組體積,但透過影像調整的處理過程可以修正體積誤差。在動態假體下的CBCT與MVCT會有明顯重組體積誤差情況出現,尤其會隨著振幅值的增加而增加,但頻率影響則較不顯著。並藉由4D(four-dimensional)挑選同相位投影的處理過程可以減低在動態假體中重組體積的誤差。
關鍵字: 扇形射束電腦斷層、錐狀射束電腦斷層、MV級電腦斷層、體積、移動假體
利用錐形射束電腦斷層(cone-beam computed tomography, CBCT)影像做為執行重新計畫設計(re-planning)為前提之目的來評估當器官運動時物體的重組體積之準確性。此實驗評估於不同移動頻率與振幅參數變化下,於醫用直線加速器的CBCT、螺旋刀(TomoTherapyTM)的MV級電腦斷層(megavoltage computed tomography, MVCT)與模擬定位扇狀射束電腦斷層(fan-beam computed tomography, FBCT)影像重組體積間之差異。
分別透過FBCT、Elekta SynergyTM的XVI(X-ray Volumetric Image)系統與MVCT擷取假體影像。靜態與動態實驗中所使用的3個假體直徑分別為5.1、9.9 與12.2 cm之壓克力球。所獲得的CBCT、MVCT與FBCT影像分別傳入治療計畫系統(Pinnacle3 7.6c)中進行影像重組體積比較。透過調整CBCT影像後可讓CBCT之CT值在治療計畫系統中運算更為準確。為了模擬呼吸運動,假體分別以振幅各為7.5、10、12.5、15、17.5與20 mm以及頻率每分鐘各為8、10、12、14、16與20次的條件下進行縱向(superior-inferior direction)運動。
在靜態假體實驗中,對於CBCT與FBCT影像之間的重組體積誤差於3個小、中與大的球假體分別為13.35%、7.47%與5.08%;經過影像調整CBCT後稱modified CBCT(mCBCT)與FBCT之間的重組體積誤差對於小、中、大球假體分別為-6.20%、-2.47%及-2.19%;在MVCT與FBCT之間的重組體積誤差對於小、中、大球假體則分別各為-10.15%、-3.65%及2.77%。不同直徑大小的球假體對於其重組體積的誤差值有明顯的差異,即愈小的球假體其重組體積誤差愈多。
在動態條件的假體實驗中,mCBCT 與MVCT影像也可以發現重組體積損失的現象。隨著振幅值的增加下重組體積誤差愈明顯。但相對地,不同的頻率變化對於重組體積誤差的影響則較不顯著。相同的,對於動態假體的mCBCT 與MVCT影像中愈小的球假體其重組體積誤差愈大。動態假體的mCBCT影像透過挑選(sort)出相同相位的投影(稱mSortCBCT)而進行體積重組,在頻率每分鐘為20次與振幅為7.5mm下所得到大球假體的重組體積與FBCT重組體積之間的誤差為-4.43%。
總結來說,靜態下的假體會因為散射的關係而高估由CBCT影像所得之重組體積,但透過影像調整的處理過程可以修正體積誤差。在動態假體下的CBCT與MVCT會有明顯重組體積誤差情況出現,尤其會隨著振幅值的增加而增加,但頻率影響則較不顯著。並藉由4D(four-dimensional)挑選同相位投影的處理過程可以減低在動態假體中重組體積的誤差。
關鍵字: 扇形射束電腦斷層、錐狀射束電腦斷層、MV級電腦斷層、體積、移動假體
Abstract
The accurate volumetric estimation of structures during organ motion forms the prerequisite basis of using cone-beam computed tomography (CBCT) images for future re-planning purposes. This study evaluated the difference in the reconstructed volume of moving phantoms between CBCT on linear accelerator, megavoltage CT (MVCT), and fan-beam CT (FBCT) from simulator, under different moving frequencies and amplitudes.
Phantom images were separately acquired by FBCT, CBCT and MVCT. Three acrylic ball phantoms with diameters of 5.1 cm, 9.9 cm, and 12.2 cm, respectively, were used for static and moving experiments. CBCT, MVCT and FBCT images were imported to a treatment planning system (Pinnacle3 7.6c) to compare the reconstructed volumes. The imaging adjustment program of CBCT was done to modify CT numbers by transforming the functions to relative electron density for the more accurate calculation. To simulate respiratory movement, the phantoms were moved longitudinally with amplitudes of 7.5, 10, 12.5, 15, 17.5 and 20 mm, and frequencies of 8, 10, 12, 14, 16 and 20 oscillations per minute, respectively.
For static phantoms, the differences in reconstructed volume of 3 phantoms (small, medium, large) between FBCT and CBCT were 13.35%, 7.47%, and 5.08%, respectively. With imaging adjustment program, the differences in reconstructed volume of 3 phantoms (small, medium, large) between FBCT and modified CBCT (mCBCT) were -6.20%, -2.47%, and -2.19%, respectively. Furthermore, the discrepancies in reconstructed volume of 3 phantoms (small, medium, large) between FBCT and MVCT were -10.15%, -3.65%, and 2.77%, respectively. The volume differences of different-sized moving phantoms were significantly different, with the greater volume losses were in small phantom than large phantom.
Significant volume losses were seen on mCBCT and MVCT for the moving phantoms. The differences in reconstructed volume significantly increased with the larger amplitude. In contrast, the differences in reconstructed volume did not differ between different frequencies. More differences of reconstructed volume in the small than large phantoms were again shown. The sorted mCBCT images of moving phantoms (mSortCBCT) were reconstructed with those at the same phases for accurate reconstruction. The differences in reconstructed volume of large phantom between FBCT and mSortCBCT (modified SortCBCT) were -4.43% with the amplitudes of 7.5mm and frequency of 20 per minute.
In conclusion, CBCT over-estimated the static phantom volume from the X-ray scattering effect. Imaging modification (mCBCT) was able to correct the volumetric difference. Significant differences in reconstructed volume were seen on mCBCT and MVCT for the moving phantom, especially with the amplitude than the frequency. Four-dimensional sorting procedure reduced the volume losses for the moving phantoms.
Keywords: Fan-beam computed tomography (FBCT), Cone-beam computed tomography (CBCT), megavoltage computed tomography (MVCT), Volume, Moving phantom
The accurate volumetric estimation of structures during organ motion forms the prerequisite basis of using cone-beam computed tomography (CBCT) images for future re-planning purposes. This study evaluated the difference in the reconstructed volume of moving phantoms between CBCT on linear accelerator, megavoltage CT (MVCT), and fan-beam CT (FBCT) from simulator, under different moving frequencies and amplitudes.
Phantom images were separately acquired by FBCT, CBCT and MVCT. Three acrylic ball phantoms with diameters of 5.1 cm, 9.9 cm, and 12.2 cm, respectively, were used for static and moving experiments. CBCT, MVCT and FBCT images were imported to a treatment planning system (Pinnacle3 7.6c) to compare the reconstructed volumes. The imaging adjustment program of CBCT was done to modify CT numbers by transforming the functions to relative electron density for the more accurate calculation. To simulate respiratory movement, the phantoms were moved longitudinally with amplitudes of 7.5, 10, 12.5, 15, 17.5 and 20 mm, and frequencies of 8, 10, 12, 14, 16 and 20 oscillations per minute, respectively.
For static phantoms, the differences in reconstructed volume of 3 phantoms (small, medium, large) between FBCT and CBCT were 13.35%, 7.47%, and 5.08%, respectively. With imaging adjustment program, the differences in reconstructed volume of 3 phantoms (small, medium, large) between FBCT and modified CBCT (mCBCT) were -6.20%, -2.47%, and -2.19%, respectively. Furthermore, the discrepancies in reconstructed volume of 3 phantoms (small, medium, large) between FBCT and MVCT were -10.15%, -3.65%, and 2.77%, respectively. The volume differences of different-sized moving phantoms were significantly different, with the greater volume losses were in small phantom than large phantom.
Significant volume losses were seen on mCBCT and MVCT for the moving phantoms. The differences in reconstructed volume significantly increased with the larger amplitude. In contrast, the differences in reconstructed volume did not differ between different frequencies. More differences of reconstructed volume in the small than large phantoms were again shown. The sorted mCBCT images of moving phantoms (mSortCBCT) were reconstructed with those at the same phases for accurate reconstruction. The differences in reconstructed volume of large phantom between FBCT and mSortCBCT (modified SortCBCT) were -4.43% with the amplitudes of 7.5mm and frequency of 20 per minute.
In conclusion, CBCT over-estimated the static phantom volume from the X-ray scattering effect. Imaging modification (mCBCT) was able to correct the volumetric difference. Significant differences in reconstructed volume were seen on mCBCT and MVCT for the moving phantom, especially with the amplitude than the frequency. Four-dimensional sorting procedure reduced the volume losses for the moving phantoms.
Keywords: Fan-beam computed tomography (FBCT), Cone-beam computed tomography (CBCT), megavoltage computed tomography (MVCT), Volume, Moving phantom