股骨髓腔幾何形狀、骨密度與股骨柄外型對股骨近端應力與股骨柄穩定度之影響

The effects of femoral canal geometries, bone densities and stem shapes on the stresses of proximal femur and stability of femoral stem

檔案

日期

2008

類型

語言

中文

摘要

全人工髖關節置換術為近代治療嚴重髖關節疾病的有效手術。近年來全人工髖關節置換術十年內的存活率通常可達到九成以上。不過術後的存活率往往會因為病患的年齡、性別與疾病等不同因素而有產生差異。然而導致全人工髖關節股骨柄失敗而必須施行再置換手術的原因,絕大部分為股骨柄無菌性鬆脫所引起。臨床上發現老化所造成的近端股骨髓腔骨質流失,使得髓腔形狀由倒香檳型變為煙囪型是導致骨水泥式股骨柄鬆脫的主因。在亞洲,因為藥物或基因的因素使得年輕人罹患高的股骨頭缺血性壞死症,會導致鬆質骨骨密度下降,這被認為是引起非骨水泥式股骨柄鬆脫的原因。目前市面上有許多外型不同的股骨柄,但是尚未有一種人工股骨柄能夠完全克服上述鬆脫的問題。因此,本研究將利用有限元素法模擬不同的股骨髓腔幾何形狀、骨密度與股骨柄外型對股骨近端應力與股骨柄穩定度之影響。本研究結果也將提供醫師在針對罹患股骨頭缺血性壞死的年輕患者與退化性關節炎的老年患者進行全人工髖關節手術前挑選股骨柄的參考。
本研究在骨水泥式模型部分,建立2組正常髓腔形狀之股骨模型並且分別植入Omnifit (high flare)與Charnley (low flare)骨水泥式股骨柄以及2組煙囪型髓腔之股骨模型分別植入Omnifit與Charnley骨水泥式股骨柄,並且以骨水泥所受之最大主應力值作為評估股骨柄是否鬆脫的標準。在非骨水泥式模型部分,建立2組非骨水泥式股柄置換模型,取正常髓腔股骨模型分別植入VerSys(直柄)與ABG(解剖柄)非骨水泥式股骨柄,並且以股骨柄與骨間的微動量值作為評估股骨柄是否鬆脫的標準。上述模型均利用有限元素分析軟體ANSYS模擬人步態時髖關節的受力狀態,並且改變股骨楊氏模數以模擬不同骨密度情況,以瞭解股骨柄的穩定度與股骨應力分佈狀態。
在骨水泥式模型部分之有限元素分析結果發現股骨髓內腔為煙囪型並植入Charnley骨水泥式股骨柄的模型,其骨水泥所受的最大主應力(73.5 MPa)與達到足以導致骨水泥破壞的體積百分比(8.2%)均較股骨髓內腔為正常型的模型高(23.3 MPa與1.7%)。若改為植入Omnifit骨水泥式股骨柄,則股骨髓內腔為煙囪型的模型其骨水泥所受的最大主應力(32.8 MPa)與骨水泥破壞體積百分比(4.9%)均會比植入Charnley股骨柄低;股骨髓內腔為正常型的模型若改為植入Omnifit骨水泥式股骨柄,也會比植入Charnley股骨柄低(15 MPa與0.26%)。當骨密度下降時(楊氏模數下降),所有模型的最大主應力與骨水泥破壞體積百分比也隨之增加。其中以股骨髓內腔為煙囪型的模型增加幅度最大,分別由2.8%(Omnifit股骨柄)與8.3%(Charnley股骨柄)增加至9.8%與16%。
在非骨水泥式模型之股骨柄穩定度有限元素分析結果發現,股骨植入直柄式股骨柄之模型,其骨與股骨柄間微動量最大值為135.7 μm,超過150 μm的接觸面積為0%;植入解剖式股骨柄其微動量最大值為199.25 μm,超過150 μm的接觸面積為2.2%。當股骨鬆質骨楊氏模數由100%下降至50%,植入直柄式股骨柄模型其微動量最大值上升為165.94 μm,接觸面積微動量超過150 μm增至0.2%;植入解剖式股骨柄之模型其接觸面微動量最大值增加為225.68μm,接觸面積微動量超過150 μm增至4.27%。股骨內側硬質骨表面von Mises應力顯示:植入直柄與解剖柄之股骨近端的von Mises應力分佈相似。但是,於距離近端股骨20 mm處之應力,解剖柄模型高於直柄模型10 MPa;於距離近端股骨60 mm處,直柄模型又高於解剖模型12 MPa。
本研究建議在施行骨水泥式股骨柄時,應該考慮病患股骨髓腔形狀與骨質,若屬煙囪型或骨質不佳應採用high flare外型的股骨柄。因為煙囪型股骨在股骨柄植入後,會造成股骨柄遠端骨水泥厚度比股骨柄近端的骨水泥厚,使得股骨柄遠端的骨水泥與硬質骨整體勁度下降。當股骨受載重時,該處所受的應力值會上升,增加骨水泥破壞的機率。採用high flare外型的股骨柄能降低骨水泥遠端應力。在施行非骨水泥式股骨柄部分,由於解剖型股骨柄是利用與股骨近端鬆質骨緊配為固定方式,而直柄型為緊配骨峽處硬質骨。因此,非創傷性股骨頭缺血性壞死病患因為鬆質骨的骨質不佳,若植入解剖型股骨柄容易導致股骨近端鬆質骨無法承受載重,引起股骨柄與骨間交界面的微動量超過骨生長臨界值。建議應採用直柄型非骨水泥式股骨柄,而避免使用解剖型非骨水泥式股骨柄。
Total hip arthroplasty is a successful surgery that fails at a rate of approximately 10% at ten years from surgery. Revision reasons varied according to the age, gender and diseases etc. Aseptic loosening of the femoral component is the main reason leading to femoral stem revised. Stovepipe canal shape which caused by aging is the main factor leading to aseptic loosening of cemented femoral component. Avascular necrosis of the femoral head is the main indication for total hip replacement in Asia. It decreased the density of cancellous bone and is considered as the main factor leading to aseptic loosening of cementless femoral component. There were many designs of femoral stem but can not avoid the aseptic loosening. The purpose of this study was, therefore, to find out the effects of femoral canal geometries, bone densities and stem shapes on the stresses of proximal femur and stability of femoral stem using finite element method. We expected that the results could provide suggestions for surgeons to select a femoral stem in patients with avascular necrosis of the femoral head or stovepipe canal before total hip arthroplasty.
Four finite element models of proximal femurs including normal and stovepipe femoral canal shapes were used to predict the failure of cement mantles for different stem shapes (Omnifit and Charnley cemented stem) and Young’s modulus of bone. Furthermore, two finite element models of proximal femur (normal canal shape) were used to evaluate the stability and stresses distribution of cementless femoral stem for different designs (VerSys and ABG) and Young’s modulus of cancellous bone. All models were simulated the loading of hip joint during heel strike of the gait cycle using the finite element analysis software (ANSYS 7.0).
The results on cemented stem analysis showed that high failure volume of cement mantle (8.2%) and tensile stress (73.5 MPa) occurred in the cement mantle of the stovepipe canal model with a low flare sem (Charnley) as compared with normal canal model (1.7% and 23.3 MPa). A high flare stem (Omnifit) reduced the peak stress and failure volume of the cement mantle not only in the stovepipe canal model but also in the normal canal model. Moreover, the peak stress and probability-of-failure of the cement mantle were sensitively with the decrease of bone density (Young’s modulus) in the stovepipe canal models. The results on cementless stem analysis showed that in the model with VerSys stem (Straight type), the peak micromotion between bone and stem was 135.7μm, and 0% of contact area at the bone/stem interface exceeded 150μm. In the model with ABG stem (Anatomic type), the peak micromotion between bone and stem was 199.25μm, and 2.2% of contact area at the bone/stem interface exceeded 150μm. The peak micromotion at the bone/stem interface was increased when the cancellous bone quality (Young’s modulus) was decreased whether implant with straight or anatomic stem. The peak micromotion was lower in the model with straight stem when decreased the modulus of cancellous bone as compared with the model with anatomic stem.
We concluded that the femoral canal shape and bone quality of patients should be considered before hip replacement surgery. A high flare, cemented femoral stem decreased the stress in the cement mantle around the stem tip. It should be considered as a better type of femoral prosthesis for patients with stovepipe canal shape or inferior bone quality. Because the cement mantle around the stem tip is thicker than proximal due to the wide isthmus of stovepipe canal after implanted a regular cemented stem. It decreased the stiffness of femur model at the region of stem tip and then increased the stress in distal cement mantle when loading occured. For young, non-trauma avascular necrosis of the femoral head patients, a cementless, straight stem which fill the isthmus is better than anatomic stem which fill the proximal part of the canal only. Because the weak cancellous bone can not resist the loading from the anatomic stem and increased the micromotion at stem/bone interface.