基於三週期最小曲面的彈性下顎彌補物:開發和評估
Triply Periodic Minimal Surfaces Based Elastic Mandibular Prosthesis: Development and Evaluation
摘要
口腔癌在台灣男性癌症中排名第四,對國人健康構成重大威脅。晚期口腔癌患者主要接受手術治療,導致下顎骨節段性缺損,影響臉部功能和外觀。為恢復下顎骨結構和功能,腓骨游離皮瓣 (Fibula free flap, FFF) 移植被視為黃金標準,但其尺寸限制和術後併發症仍需解決。常見併發症源於骨與植入物間隙造成的骨癒合不良,導致硬體因應力集中而損壞。為解決此問題,本研究旨在開發並評估基於三週期最小曲面 (Triply periodic minimal surfaces, TPMS) 的彈性下顎彌補物,透過特殊設計減少骨與植入物間隙,優化下顎重建系統的力學表現。
下顎彌補物針對患者下顎輪廓進行客製化設計,支架具適合的晶格參數,有助於骨組織內生長。TPMS支架用於支撐咬合力造成的負載,正弦絲網絡 (Sinusoidal filament networks, SFN) 支架作為彈性結構,能向外撐開減少間隙。客製化重建板設計能緊密貼合彌補物支架的幾何輪廓,提高整體結構穩定性。本研究在開發時嚴格遵循選擇性雷射燒結 (Selective laser sintering, SLS) 技術的限制,以確保支架製造可行性。
為驗證下顎彌補物的成效,本研究使用有限元素分析 (Finite element analysis, FEA) 評估支架和下顎重建系統的性能。建立支架骨內生長模型模擬骨組織長入支架孔洞過程,透過壓縮測試模擬獲得支架骨內生長模型應力、應變關係。基於材料力學建立均質化的等效實心模型用於降低多孔結構模擬運算資源。最後,通過咬合模擬探討咬合力對下顎重建系統的影響。
結果顯示,穩定的骨內生長能提升支架3.5倍的強度和33.9倍的剛性,減少系統內部85%的能量累積,分擔更多來自硬體的機械負載,降低損壞風險,減少系統內90%的最大形變。重建板的特殊設計能提升系統穩定性和安全性。隨著骨組織增加,下顎彌補物能應付日常咀嚼,但對抗最大咬合力仍有疑慮。
下顎彌補物針對患者下顎輪廓進行客製化設計,支架具適合的晶格參數,有助於骨組織內生長。TPMS支架用於支撐咬合力造成的負載,正弦絲網絡 (Sinusoidal filament networks, SFN) 支架作為彈性結構,能向外撐開減少間隙。客製化重建板設計能緊密貼合彌補物支架的幾何輪廓,提高整體結構穩定性。本研究在開發時嚴格遵循選擇性雷射燒結 (Selective laser sintering, SLS) 技術的限制,以確保支架製造可行性。
為驗證下顎彌補物的成效,本研究使用有限元素分析 (Finite element analysis, FEA) 評估支架和下顎重建系統的性能。建立支架骨內生長模型模擬骨組織長入支架孔洞過程,透過壓縮測試模擬獲得支架骨內生長模型應力、應變關係。基於材料力學建立均質化的等效實心模型用於降低多孔結構模擬運算資源。最後,通過咬合模擬探討咬合力對下顎重建系統的影響。
結果顯示,穩定的骨內生長能提升支架3.5倍的強度和33.9倍的剛性,減少系統內部85%的能量累積,分擔更多來自硬體的機械負載,降低損壞風險,減少系統內90%的最大形變。重建板的特殊設計能提升系統穩定性和安全性。隨著骨組織增加,下顎彌補物能應付日常咀嚼,但對抗最大咬合力仍有疑慮。
Oral cancer ranks fourth among cancers affecting Taiwanese males, posing a significant health threat to the population. Surgery remains the primary treatment for advanced stages of oral cancer but often results in segmental defects of the mandible, impacting facial function and appearance. Fibular-free flap (FFF) transplantation is considered the gold standard for mandibular reconstruction to restore the mandible's structure and function. However, challenges such as natural size limitations and postoperative complications need addressing. Common complications arise from poor bone integration due to gaps between bone and implants, leading to hardware failure. To address these issues, this study aims to develop and evaluate an elastic mandibular prosthesis based on triply periodic minimal surfaces (TPMS) designed to minimize gaps between bone and implant through a unique structural design, optimizing the mechanical performance of the mandibular reconstruction system.
The development of the mandibular prosthesis focuses on customized designs tailored to the patient's mandibular contour. The prosthesis scaffold supports proper lattice parameters to facilitate osseointegration, with an internal TPMS scaffold to withstand occlusal loads and a sinusoidal filament network (SFN) scaffold providing flexibility to minimize gaps. Customized reconstruction plate designs ensure close adherence to the prosthesis scaffold's geometric profile, enhancing overall structural stability. Strict adherence to selective laser sintering (SLS) technology constraints ensures the manufacturability of the prosthesis scaffold.
Finite element analysis (FEA) was employed to assess the performance of the scaffold and mandibular reconstruction system and validate the efficacy of the mandibular prosthesis. Scaffold bone ingrowth models were established to simulate bone ingrowth into the scaffold pores, stress-strain relationships in the scaffold bone ingrowth model via simulated compression testing and reduced computational resources during porous structure simulation using a homogenized equivalent entity model based on material mechanics. Occlusion force simulation was conducted to investigate the impact of occlusal forces on the mandibular reconstruction system.
Results indicate that stable bone ingrowth enhances scaffold strength by 3.5 times and stiffness by 33.9 times, reduces 85% of the internal energy accumulation within the system, enabling it to withstand higher mechanical loads from the hardware, and reducing the risk of damage while reducing 90% of maximum deformation in the system. The specialized design of the reconstruction plate improves system stability and safety. As bone tissue increases, the mandibular prosthesis can accommodate daily masticatory conditions, though concerns remain regarding resistance to maximum bite force.
The development of the mandibular prosthesis focuses on customized designs tailored to the patient's mandibular contour. The prosthesis scaffold supports proper lattice parameters to facilitate osseointegration, with an internal TPMS scaffold to withstand occlusal loads and a sinusoidal filament network (SFN) scaffold providing flexibility to minimize gaps. Customized reconstruction plate designs ensure close adherence to the prosthesis scaffold's geometric profile, enhancing overall structural stability. Strict adherence to selective laser sintering (SLS) technology constraints ensures the manufacturability of the prosthesis scaffold.
Finite element analysis (FEA) was employed to assess the performance of the scaffold and mandibular reconstruction system and validate the efficacy of the mandibular prosthesis. Scaffold bone ingrowth models were established to simulate bone ingrowth into the scaffold pores, stress-strain relationships in the scaffold bone ingrowth model via simulated compression testing and reduced computational resources during porous structure simulation using a homogenized equivalent entity model based on material mechanics. Occlusion force simulation was conducted to investigate the impact of occlusal forces on the mandibular reconstruction system.
Results indicate that stable bone ingrowth enhances scaffold strength by 3.5 times and stiffness by 33.9 times, reduces 85% of the internal energy accumulation within the system, enabling it to withstand higher mechanical loads from the hardware, and reducing the risk of damage while reducing 90% of maximum deformation in the system. The specialized design of the reconstruction plate improves system stability and safety. As bone tissue increases, the mandibular prosthesis can accommodate daily masticatory conditions, though concerns remain regarding resistance to maximum bite force.