以光纖輔助硬脊膜外腔定位之技術
A Fiberoptic Guided Technique to Assist Epidural Needle Placement
摘要
本研究係利用已廣泛使用之光纖技術(fiber optics)將光纖束(optical fiber bundle) 放置於施打硬脊膜外導管的epidural needle內,藉由不同組織間反射光譜不同的特徵來達到偵測硬脊膜腔之目的,以輔助臨床麻醉醫師能正確地放置硬脊膜導管。
硬脊膜外腔麻醉法(epidural aneshtsia)是屬於神經軸(neural axial)麻醉技術的一種,它可以提供一個固定範圍的麻醉而不像類似的脊椎麻醉法(spinal anesthesia),是一種全有全無(all or none)的麻醉,所以有較大的彈性。這種硬脊膜外腔的麻醉法,廣泛的應用在許多方面,包括:手術中麻醉、手術後疼痛控制、婦產科的無痛分娩(painless labor)、與慢性疼痛控制的各個領域。
解剖學上,硬脊膜外腔(epidural space)是環繞在硬腦膜(dura mata)外,從前、 外、後三側,三方面包圍著硬腦膜,神經根(nerve root)從這個空間通過,然後從外側的孔洞離開而形成了周邊的神經(peripheral nerve)。 在硬脊膜外腔中,除了神經根之外還包含了許多組織,像脂肪結締組織(fatty connective tissue)、淋巴和豐富的靜脈叢(venous plexus)外, 也有一些結締組織的分隔(septa)存在其中。
硬脊膜腔導管的置入可以讓我們間段的或連續性的對這個地方來加入藥物,達到我們要的麻醉效果。目前的傳統置放方法又稱阻力消失法(loss of resistence),是利用少許空氣或水在空針中,當穿刺針由皮膚逐漸接近硬脊膜外腔時,施打者不斷的測試其阻力。一般來說,在到達真正的硬脊膜外腔前的阻力均較大,特別是當最後來到黃韌帶時,會有特殊的手感,一旦穿過黃韌帶,阻力將瞬間消失,有經驗的麻醉醫師將能輕易的判定已經進入該空腔;但是因為個體的差異,限制了這種方法的成功率。當有一些特殊的病人其解剖結構特異時,也可能誤判而造成導管進入錯誤的空間(pseudo space)而不是真正的硬脊膜外腔。另一方面,很可能因為阻力消失的不夠明顯,讓施打者誤判而將穿刺針繼續前以致造成硬腦膜穿刺(dura puncture),引起病人頭痛。
自從上世紀中業開始, 硬脊膜外腔麻醉法首先被應用在臨床上後, 已經經過半世紀之久。 雖然大家不斷努力創新,但是仍然缺乏一個客觀實用又易學易用的方法來進行硬脊膜外腔導管的置放。近期加拿大學者 Tsui 發表了一系列的論文探討關於利用電刺激的方法來確認導管位置的方法已經成為一個熱門的話題, 但是這種方法仍然是事後驗證的工具, 無法提供麻醉醫師一個有效在置放導管前的一個幫助。超音波輔助的硬脊膜外腔偵測,近年來因超音波探頭技術的精進,已經廣為人們注意,而且在近三年來有許多論文出現。但是目前的應用只能在新生兒與幼童身上使用,在成人身上仍然因為身體結構上骨頭的阻擾,所以只能當作輔助的工具,無法真正的在成人身上作精確定位。因此我們這個研究想要積極尋找一種可以正確定位硬脊膜外腔的方法,來幫助麻醉醫師甚至一般醫師都可以輕易的找到該空腔,並成功置放導管、給藥。 而不用擔心硬腦膜穿刺和置放於錯誤空腔的問題。
利用光纖的幫助我們可以將一些特定波長的光源導入穿刺針中,在穿刺針前進時,不斷有反射光傳回來,透過後端儀器的分析來及時反應目前針頭所在的位置;在利用不同組織有不同反射特性的原理來達到黃韌帶,硬脊膜外腔的定位, 進而整合出一套好用的醫療儀器。
實驗方法:
本研究經過台北榮總動物實驗委員會認可後實施。其中共包含兩階段的實驗;第一階段是體外組織前導實驗,係利用豬的組織找出最能分辨出黃韌帶與硬腦膜的可見光波長來應用在下一階段。第二階段是活體豬實驗,利用前一階段得出之波長以5隻18-24 公斤重的豬接受全身麻醉誘導後,使用17-gauge 之特殊設計,內含六根光纖之epidural stylet,這種stylet是專門設計來導引我們交錯傳遞的雷射光並收集從皮膚至硬脊膜腔各層組織間不同的反射光用的。 反射光由此收集後噵入光電倍增管並加以處理轉為電子訊號,經放大後,顯示在示波器並儲存在電腦中。然後這些結果再做進一步的統計分析。
實驗結果
體外實驗: 藉由離體組織的反射光譜,我們發現使用650 nm 和532 nm 作為光源時,計算黃韌帶和硬脊膜的反射光比例分別是 2.8191+/- 0.2711 和 1.5882+/- 0.1403 (n=5),似乎表示可以使用這兩個波長來分辨出黃韌帶和硬脊膜,進而定位出硬脊膜外腔,所以我們就選擇這兩種市面上常見的雷射光源作為下一階段動物實驗的光源。
活體實驗: 總共有50次實驗穿刺被記錄下來。Paired t-test 顯示在黃韌帶和硬脊膜外腔間有著顯著的統計差異;其中觀察單獨使用650 nm時,其 p<0.001,而使用532 nm時,p=0.014。對硬脊膜腔使用 650 nm 和 532 nm 雷射光源時它們反射光的強度以及兩強度間的比值分別是3.565+/-0.194, 2.542+/-0.145, 0.958+/-0.172 而同樣方法在黃韌帶得到的結果依序是 3.842+/-0.191, 2.563+/-0.131, 1.228+/-0.244 。
同時我們也發現在不同節的胸椎與腰椎中,他們黃韌帶和硬脊膜外腔的光學性質並沒有統計上的差異 (Linear mixed model: p =0.9 for 650nm and p=0.35 for 532nm) 。
結論:
這是第一個引用全新光學方法來定位硬脊膜外腔的動物實驗。在活體 實驗中, 我們發現若使用單一波長650nm的雷射光時, 從黃韌帶和硬脊膜外腔來的反射光也可以很輕易的分辨兩層的位置,表示這個技術可以成功的定位出硬脊膜外腔,進而幫助硬脊膜外腔導管的置放。
硬脊膜外腔麻醉法(epidural aneshtsia)是屬於神經軸(neural axial)麻醉技術的一種,它可以提供一個固定範圍的麻醉而不像類似的脊椎麻醉法(spinal anesthesia),是一種全有全無(all or none)的麻醉,所以有較大的彈性。這種硬脊膜外腔的麻醉法,廣泛的應用在許多方面,包括:手術中麻醉、手術後疼痛控制、婦產科的無痛分娩(painless labor)、與慢性疼痛控制的各個領域。
解剖學上,硬脊膜外腔(epidural space)是環繞在硬腦膜(dura mata)外,從前、 外、後三側,三方面包圍著硬腦膜,神經根(nerve root)從這個空間通過,然後從外側的孔洞離開而形成了周邊的神經(peripheral nerve)。 在硬脊膜外腔中,除了神經根之外還包含了許多組織,像脂肪結締組織(fatty connective tissue)、淋巴和豐富的靜脈叢(venous plexus)外, 也有一些結締組織的分隔(septa)存在其中。
硬脊膜腔導管的置入可以讓我們間段的或連續性的對這個地方來加入藥物,達到我們要的麻醉效果。目前的傳統置放方法又稱阻力消失法(loss of resistence),是利用少許空氣或水在空針中,當穿刺針由皮膚逐漸接近硬脊膜外腔時,施打者不斷的測試其阻力。一般來說,在到達真正的硬脊膜外腔前的阻力均較大,特別是當最後來到黃韌帶時,會有特殊的手感,一旦穿過黃韌帶,阻力將瞬間消失,有經驗的麻醉醫師將能輕易的判定已經進入該空腔;但是因為個體的差異,限制了這種方法的成功率。當有一些特殊的病人其解剖結構特異時,也可能誤判而造成導管進入錯誤的空間(pseudo space)而不是真正的硬脊膜外腔。另一方面,很可能因為阻力消失的不夠明顯,讓施打者誤判而將穿刺針繼續前以致造成硬腦膜穿刺(dura puncture),引起病人頭痛。
自從上世紀中業開始, 硬脊膜外腔麻醉法首先被應用在臨床上後, 已經經過半世紀之久。 雖然大家不斷努力創新,但是仍然缺乏一個客觀實用又易學易用的方法來進行硬脊膜外腔導管的置放。近期加拿大學者 Tsui 發表了一系列的論文探討關於利用電刺激的方法來確認導管位置的方法已經成為一個熱門的話題, 但是這種方法仍然是事後驗證的工具, 無法提供麻醉醫師一個有效在置放導管前的一個幫助。超音波輔助的硬脊膜外腔偵測,近年來因超音波探頭技術的精進,已經廣為人們注意,而且在近三年來有許多論文出現。但是目前的應用只能在新生兒與幼童身上使用,在成人身上仍然因為身體結構上骨頭的阻擾,所以只能當作輔助的工具,無法真正的在成人身上作精確定位。因此我們這個研究想要積極尋找一種可以正確定位硬脊膜外腔的方法,來幫助麻醉醫師甚至一般醫師都可以輕易的找到該空腔,並成功置放導管、給藥。 而不用擔心硬腦膜穿刺和置放於錯誤空腔的問題。
利用光纖的幫助我們可以將一些特定波長的光源導入穿刺針中,在穿刺針前進時,不斷有反射光傳回來,透過後端儀器的分析來及時反應目前針頭所在的位置;在利用不同組織有不同反射特性的原理來達到黃韌帶,硬脊膜外腔的定位, 進而整合出一套好用的醫療儀器。
實驗方法:
本研究經過台北榮總動物實驗委員會認可後實施。其中共包含兩階段的實驗;第一階段是體外組織前導實驗,係利用豬的組織找出最能分辨出黃韌帶與硬腦膜的可見光波長來應用在下一階段。第二階段是活體豬實驗,利用前一階段得出之波長以5隻18-24 公斤重的豬接受全身麻醉誘導後,使用17-gauge 之特殊設計,內含六根光纖之epidural stylet,這種stylet是專門設計來導引我們交錯傳遞的雷射光並收集從皮膚至硬脊膜腔各層組織間不同的反射光用的。 反射光由此收集後噵入光電倍增管並加以處理轉為電子訊號,經放大後,顯示在示波器並儲存在電腦中。然後這些結果再做進一步的統計分析。
實驗結果
體外實驗: 藉由離體組織的反射光譜,我們發現使用650 nm 和532 nm 作為光源時,計算黃韌帶和硬脊膜的反射光比例分別是 2.8191+/- 0.2711 和 1.5882+/- 0.1403 (n=5),似乎表示可以使用這兩個波長來分辨出黃韌帶和硬脊膜,進而定位出硬脊膜外腔,所以我們就選擇這兩種市面上常見的雷射光源作為下一階段動物實驗的光源。
活體實驗: 總共有50次實驗穿刺被記錄下來。Paired t-test 顯示在黃韌帶和硬脊膜外腔間有著顯著的統計差異;其中觀察單獨使用650 nm時,其 p<0.001,而使用532 nm時,p=0.014。對硬脊膜腔使用 650 nm 和 532 nm 雷射光源時它們反射光的強度以及兩強度間的比值分別是3.565+/-0.194, 2.542+/-0.145, 0.958+/-0.172 而同樣方法在黃韌帶得到的結果依序是 3.842+/-0.191, 2.563+/-0.131, 1.228+/-0.244 。
同時我們也發現在不同節的胸椎與腰椎中,他們黃韌帶和硬脊膜外腔的光學性質並沒有統計上的差異 (Linear mixed model: p =0.9 for 650nm and p=0.35 for 532nm) 。
結論:
這是第一個引用全新光學方法來定位硬脊膜外腔的動物實驗。在活體 實驗中, 我們發現若使用單一波長650nm的雷射光時, 從黃韌帶和硬脊膜外腔來的反射光也可以很輕易的分辨兩層的位置,表示這個技術可以成功的定位出硬脊膜外腔,進而幫助硬脊膜外腔導管的置放。
Epidural block is widely used in many fields of anesthesia such as epidural anesthesia, postoperative analgesia, and painless labor. Up to 10% of epidurals fail due to incorrect catheter placement. Unfortunately, there is still no reliable, objective, “on the spot” way of determining correct catheter placement in the epidural space. Technique of loss-of-resistance (LOR) for epidural block is commonly used in humans with approximately 90% successful rate. However the learning curve is pretty long for a new operator. So an efficient, convenient, real-time, and reliable tool for epidural catheter placement is truly needed.
Fiber optics are used widely especially in medical application. A single optical fiber is so slim and it could be fit into any needle to help us to exactly locate the position we were between the tissues.
We describe a novel optical method to assist epidural catheter insertion in a porcine model. A two-wavelength (650 nm and 532 nm) fiber-optical method has been developed according to the characteristic reflectance spectra of the tissues during the needle insertion to localize the epidural space (ES).
Methods: The study was approved by the institutional animal care and use committee of the Taipei Veterans General Hospital. Two experiments were carried out in this study. The first was a pilot in-vitro study to find the feature of tissue in optics which could be used to distinguish the dura and ligamentum flavum. In the second experiments, a 17 Ga epidural needle with a special designed hollow stylet which contained 6 optic fibers was used. The stylet was designed to deliver two counter-phased laser lights with wavelengths of 650 and 532 nm for illuminating the tissues. The reflective light was also collect from the stylet and then converted to electrical signal. The electrical signal was then amplified again and displayed on an oscilloscope so that the magnitudes of the reflective lights of 650 and 532 nm were able to be recorded and analysis. Data were expressed as mean+/-SD.
Results:
Ex-vivo study: While calculated the ratios of the reflective lights from LF and Dura for 650 nm to 532 nm, the results were 2.8191+/- 0.2711 and 1.5882+/- 0.1403 (n=5), respectively. It seems that the ratios between dura and ligamentum flavum have great difference and can be used to locate the epidural space. The ratios of the reflective lights for 650 nm to 532 nm were for dura mater and LF, respectively.
In vivo study: A total of 50 punctures were done in four laboratory pigs. Paired t-test show that significant optical differences between the epidural space and ligamentum flavum at both 650 nm (p<0.001), and 532 nm (p=0.014). Mean magnitudes for 650 nm and 532 nm and their ratio at epidural space and ligamentum flavum were 3.565+/-0.194, 2.542+/-0.145, 0.958+/-0.172 and 3.842+/-0.191, 2.563+/-0.131, 1.228+/-0.244 respectively.
There were no differences in optical characteristics of the ligamentum flavum and epidural space at different levels in the lumbar and thoracic region (Linear mixed model: p =0.9 for 650nm and p=0.35 for 532nm).
Conclusions: This is the first study to introduce a new optical method to localize the epidural space in a porcine model. The results of in vivo study in porcine showed that the reflected lights from ES and ligamentum flavum (FL) is highly distinguishable. This indicates that this technique can localize the ES on the spot without the help of additional guiding assistance. The epidural space could be identified by changes in the reflective pattern of light emitted at 650 nm that were specific for the ligamentum flavum and dural tissue. Real-time optical information successfully guided a modified Tuohy needle into the epidural space.
Fiber optics are used widely especially in medical application. A single optical fiber is so slim and it could be fit into any needle to help us to exactly locate the position we were between the tissues.
We describe a novel optical method to assist epidural catheter insertion in a porcine model. A two-wavelength (650 nm and 532 nm) fiber-optical method has been developed according to the characteristic reflectance spectra of the tissues during the needle insertion to localize the epidural space (ES).
Methods: The study was approved by the institutional animal care and use committee of the Taipei Veterans General Hospital. Two experiments were carried out in this study. The first was a pilot in-vitro study to find the feature of tissue in optics which could be used to distinguish the dura and ligamentum flavum. In the second experiments, a 17 Ga epidural needle with a special designed hollow stylet which contained 6 optic fibers was used. The stylet was designed to deliver two counter-phased laser lights with wavelengths of 650 and 532 nm for illuminating the tissues. The reflective light was also collect from the stylet and then converted to electrical signal. The electrical signal was then amplified again and displayed on an oscilloscope so that the magnitudes of the reflective lights of 650 and 532 nm were able to be recorded and analysis. Data were expressed as mean+/-SD.
Results:
Ex-vivo study: While calculated the ratios of the reflective lights from LF and Dura for 650 nm to 532 nm, the results were 2.8191+/- 0.2711 and 1.5882+/- 0.1403 (n=5), respectively. It seems that the ratios between dura and ligamentum flavum have great difference and can be used to locate the epidural space. The ratios of the reflective lights for 650 nm to 532 nm were for dura mater and LF, respectively.
In vivo study: A total of 50 punctures were done in four laboratory pigs. Paired t-test show that significant optical differences between the epidural space and ligamentum flavum at both 650 nm (p<0.001), and 532 nm (p=0.014). Mean magnitudes for 650 nm and 532 nm and their ratio at epidural space and ligamentum flavum were 3.565+/-0.194, 2.542+/-0.145, 0.958+/-0.172 and 3.842+/-0.191, 2.563+/-0.131, 1.228+/-0.244 respectively.
There were no differences in optical characteristics of the ligamentum flavum and epidural space at different levels in the lumbar and thoracic region (Linear mixed model: p =0.9 for 650nm and p=0.35 for 532nm).
Conclusions: This is the first study to introduce a new optical method to localize the epidural space in a porcine model. The results of in vivo study in porcine showed that the reflected lights from ES and ligamentum flavum (FL) is highly distinguishable. This indicates that this technique can localize the ES on the spot without the help of additional guiding assistance. The epidural space could be identified by changes in the reflective pattern of light emitted at 650 nm that were specific for the ligamentum flavum and dural tissue. Real-time optical information successfully guided a modified Tuohy needle into the epidural space.