新型動態髖骨螺釘之 有限元素分析及生物力學測試
Finite element analysis and biomechanical testing of a modified dynamic hip screw
摘要
動態髖骨螺釘(Dynamic hip screw, DHS)為應用於治療轉子間骨折常見的內固定器,特別是骨質疏鬆症患者伴隨不穩定型骨折,臨床上已有不錯的成效。針對髖螺釘(lag screw)的部分,儘管市面上已存在多種設計,目前仍以傳統型及螺旋葉片型螺牙為主,其最主要失敗模式為股骨球頭內翻崩陷,進而使得髖螺釘相對於股骨球頭向上穿出,或稱為骨切效應(Superior cut-out),因此,本研究擬設計出一款新型髖螺釘,利用傳統型螺釘設計搭配兩側刃片,增加植入物與鬆質骨接觸面積,提供防穿出、抗旋轉功能,並提升髖螺釘在球頭內的穩定性。
本研究於新型髖螺釘設計階段,利用有限元素分析比較三種植入物優劣(傳統型螺牙設計、螺旋形葉片及新型髖螺釘),以多孔聚氨酯硬性發泡材料模擬人體的鬆質骨作為植入基材,並給予軸向負載模擬人體正常步態下髖關節最大受力情形。而在新型產品實體打樣完成後,將其實際鎖入上述多孔聚氨酯硬化發泡材料模擬之人工假骨球頭,進行週期性測試實驗,最後測試結果搭配X-ray攝影進行觀察,探討三者抗穿出情形。
研究結果顯示,傳統型螺牙設計傾向於向上及內側偏移,螺旋葉片型設計則存在著人造骨破裂風險,新型髖螺釘設計相較於其他兩款產品,顯示有較佳抵抗髖螺釘位移的能力(DHS Screw: 1.64±0.32mm、DHS Blade: 1.36±0.17mm、ODRC DHS: 1.14±0.22mm, p=0.000)及抗旋轉功能(DHS Screw:14.48°±12.96°、DHS Blade: 26.35°±18.54°、ODRC DHS: 3.17°±1.43°, p=0.054),增進髖螺釘的錨定強度。
本研究於新型髖螺釘設計階段,利用有限元素分析比較三種植入物優劣(傳統型螺牙設計、螺旋形葉片及新型髖螺釘),以多孔聚氨酯硬性發泡材料模擬人體的鬆質骨作為植入基材,並給予軸向負載模擬人體正常步態下髖關節最大受力情形。而在新型產品實體打樣完成後,將其實際鎖入上述多孔聚氨酯硬化發泡材料模擬之人工假骨球頭,進行週期性測試實驗,最後測試結果搭配X-ray攝影進行觀察,探討三者抗穿出情形。
研究結果顯示,傳統型螺牙設計傾向於向上及內側偏移,螺旋葉片型設計則存在著人造骨破裂風險,新型髖螺釘設計相較於其他兩款產品,顯示有較佳抵抗髖螺釘位移的能力(DHS Screw: 1.64±0.32mm、DHS Blade: 1.36±0.17mm、ODRC DHS: 1.14±0.22mm, p=0.000)及抗旋轉功能(DHS Screw:14.48°±12.96°、DHS Blade: 26.35°±18.54°、ODRC DHS: 3.17°±1.43°, p=0.054),增進髖螺釘的錨定強度。
Dynamic hip screw (DHS) is the commonly utilized treatment to femoral introterchanteric fracture, especially in unstable osteoporotic fracture. High postoperative satisfactory has been achieved. Conventional thread and helical blade designs of lag screws are mainly applied in DHS system. The most common failure mechanism is the varus collapse of femoral head, and subsequently, followed by superior migration (Superior cut-out) of the lag screw through the cancellous bone of the femoral head. The purpose of this study is to design a lag screw combined with conventional screw and two supplemental blades. The structure of the implant is designed to provide larger contact surface in order to prevent superior cut-out and rotation of the femoral head, thus the stability of internal fixation shall be improved.
During the design stage, finite element method is applied to com-pare superior cut-outresistance with those three implants (DHS Screw, DHS Blade, and ODRC DHS). Implants were virtually fixed into cellular polyurethane rigid foam which is designated to simulate the osteoporotic cancellous bone. Axial loads were applied as the representative hip loading during normal gait. After the prototype of new design was manufactured, the actual implantation of three implants into cellular polyurethane rigid foam as surrogate femoral head, were performed for mechanical tests under cyclic loading. The final test results were compared with radiographs.
The results showed that the conventional screw was prone to superior and medial migration, while the blade-type lag screw could lead to micro-fracture and breakage of the surrogate femoral head. Compared to those two commercial products, significant improvement in superior cut-out resistance (DHS Screw: 1.64±0.32mm、DHS Blade: 1.36±0.17mm、ODRC DHS: 1.14±0.22mm, p=0.000) and anti-rotation (DHS Screw:14.48°±12.96°、DHS Blade: 26.35°±18.54°、ODRC DHS: 3.17°±1.43°, p=0.054) have been revealed in ODRC DHS.
During the design stage, finite element method is applied to com-pare superior cut-outresistance with those three implants (DHS Screw, DHS Blade, and ODRC DHS). Implants were virtually fixed into cellular polyurethane rigid foam which is designated to simulate the osteoporotic cancellous bone. Axial loads were applied as the representative hip loading during normal gait. After the prototype of new design was manufactured, the actual implantation of three implants into cellular polyurethane rigid foam as surrogate femoral head, were performed for mechanical tests under cyclic loading. The final test results were compared with radiographs.
The results showed that the conventional screw was prone to superior and medial migration, while the blade-type lag screw could lead to micro-fracture and breakage of the surrogate femoral head. Compared to those two commercial products, significant improvement in superior cut-out resistance (DHS Screw: 1.64±0.32mm、DHS Blade: 1.36±0.17mm、ODRC DHS: 1.14±0.22mm, p=0.000) and anti-rotation (DHS Screw:14.48°±12.96°、DHS Blade: 26.35°±18.54°、ODRC DHS: 3.17°±1.43°, p=0.054) have been revealed in ODRC DHS.