低劑量內插平均CT降低PETCT呼吸假影可行性研究

Feasibility studies of low dose interpolated average CT in PET imaging

檔案

日期

2012

類型

語言

中文

摘要

為了解決電腦斷層掃描(computed tomography, CT)和正子發射斷層掃描(positron emission tomography, PET)由於時間解析度不同所導致的影像錯位(image misalignment)、以及錯估SUV (standard uptake value)值,Cine average CT (CACT)掃描模式,可用來取代臨床預設的螺旋CT (helical CT, HCT)對PET影像進行衰減校正(attenuation correction),但是,CACT掃描帶來的輻射劑量(radiation dose)比臨床預設的CT掃描高出許多。因此,本研究的目的為建立低劑量內插平均CT(interpolated average CT, IACT)對PET影像進行衰減校正以解決呼吸假影的問題。
首先,從CACT所獲得的13個時期影像中,挑出8組、6組、4組和2組時期的原始CACT影像(original images),藉由光流演算法(optical flow)計算呼吸運動中每個像素的位移量,並依序內插出其他5組、7組、9組和11組相位影像(interpolated images),最後將此13組相位的CT影像進行平均,以獲得IACT影像,稱之為IACT8o5i 、IACT6o7i、IACT4o9i和IACT2o11i,此目的為證明使用兩組原始影像的IACT2o11i即可獲得和CACT同樣品質的影像。再進一步使用兩次低劑量的HCT,其中各別在吸氣(inhalation)和吐氣(exhalation)兩種狀態下掃描,並使用光流演算法內插11組未掃描的呼吸週期之CT影像,最後將此13組相位的CT影像平均,以獲得IACTin-ex影像。
研究結果發現,使用最多組真實影像所平均的IACT可獲得最佳的影像資訊,IACT2o11i影像和CACT影像的SUV值差異度最大值為3%,然而,HCT影像和CACT影像的SUV值差異度最大高達35.5%,相較之下,使用IACT2o11i影像已可獲得足夠的診斷資訊,也並沒有產生呼吸假影。進而發現IACTin-ex影像和HCT、CACT相較後證實,可獲得與CACT相同的診斷資訊,並且僅帶來0.38 mSv的輻射劑量,與HCT和CACT技術相比大幅下降92 %和84%的輻射劑量。
To reduce misalignments and standardized uptake value (SUV) errors resulted from the difference of temporal resolution between positron emission tomography (PET) and computed tomography(CT) scans in PET/CT imaging, cine average CT (CACT) has been proposed for attenuation correction (AC). In terms of radiation dose delivered to patients CACT technology raises the concern. Therefore, the purpose of this study is to explore the feasibility of applying low dose interpolated average CT (IACT) for AC of the PET in PET/CT.
First, the 8, 6, 4 and 2 phases from CACT were used as the original phases. Optical flow method with motion estimation from two phases helps to generate interpolated phases. Those 13 phases were averaged to generate the IACT8o5i, IACT6o7i, IACT4o9i and IACT2o11i. This purpose was to prove image quality of IACT2o11i images was the same with CACT. Next, IACTin-ex was generated from all the 13 phase CT images, including original inhale/exhale breath-hold HCT images and 11 interpolated CT sets.
The results showed that as the number of original phases in IACT increases, IACT and CACT become more relevant. With IACT2o11i where the difference is the largest among all IACT configurations, 3% error in comparison of CACT was acquired. Maximum difference between the HCT and CACT is 35.5%. The radiation dose using IACTin-ex could be reduced by 92% and 84% compared to HCT and CACT, respectively. The application of IACT for PET AC reduces misalignment artifacts and SUV quantification errors for thoracic tumors as compared to that of HCT.