比較使用平板偵檢器電腦斷層與256切電腦斷層施作腦血流灌注檢查之輻射劑量與致癌風險
Flat Detector CT versus 256-slice CT to Assess Cerebral Perfusion – Comparison of Radiation Dose and Related Cancer Risk
摘要
背景:
目前在臨床的放射攝影和透視攝影中,平板偵檢器(flat-panel detector, FD)大都已經取代了標準X光底片、底片結合增感屏和影像增強管;由於該偵檢器具有高偵測量子效率(detective quantum efficiency, DQE)、高取像速率(frame rate)、高空間解析度(spatial resolution)及寬廣的動態範圍(dynamic range)等優勢,因此若能應用在血管攝影機組(angiographic suites)上,將使一站式造影服務(one-stop imaging service, OSIS)在處理腦神經血管等相關問題成為可行。在一站式造影服務中,腦血流動力學(即腦循環時間和腦實質的血流容積)可從一次造影中獲得,然而在廣泛應用該造影策略之前,其輻射劑量和相關的輻射危害仍是目前臨床主要關注的焦點。
目的:
本研究之目的在於評估使用平板偵檢器電腦斷層(flat detector computed tomography, FDCT)和256切電腦斷層(256-slice CT)施作腦血流灌注檢查,所累積之輻射劑量與致癌風險。
材料與方法:
本研究使用平板偵檢器電腦斷層(FDCT)和256切電腦斷層造影設備,搭配擬人假體和熱發光劑量計(thermo-luminescence dosimeter, TLD),進行輻射劑量的量測。在8公分和16公分腦實質血流容積(parenchymal blood volume, PBV)量測方面,使用數位減贅血管攝影(digital subtraction angiography, DSA)方式得之,其造影序列先以每秒60張的取像速率旋轉8秒,取得面罩影像,接續再以相同的取像速率旋轉8秒,取得充盈影像。在8公分影像的掃描參數如下:管電壓73 kV、管電流486 mA、曝射時間4.0 ms、照野48cm × 48cm;而16公分影像的掃描參數則為:管電壓72 kV、管電流540 mA、曝射時間4.1 ms、照野48cm × 48cm。8公分和16公分的掃描範圍分別為從蝶鞍到顱頂以及整個腦部。80顆熱發光劑量計(TLD)分別塞入擬人假體之頭部到骨盆等20個器官部位中。使用256切電腦斷層進行電腦斷層灌注造影(computed tomography perfusion, CTP)檢查,其掃描範圍和8公分腦實質血流容積(PBV)檢查相同,其掃描參數如下:切片厚度為5 mm、準直儀所開的範圍為64 × 1.25mm、照野大小為20cm × 20cm、管電壓為80 kV、管電流為250 mA、管電流時間乘積為100 mAs、管球的旋轉和曝射時間均為0.4 sec、掃描區間為1.5 sec然後持續1 min。本研究採用國際放射防護委員會(International Commission on Radiological Protection, ICRP)第103號報告中所建議之器官加權因子,進行有效劑量(effective dose)之計算;另採用游離輻射生物效應(Biological Effects of Ionizing Radiation, BEIR)第七份報告,做為致癌風險評估之依據。
結果:
本研究在8公分和16公分腦實質血流容積(PBV)檢查下,其有效劑量分別為0.87 ± 0.55 mSv和3.91 ± 0.78 mSv;而在8公分電腦斷層灌注造影(CTP)檢查時,其有效劑量則為2.77 ± 1.59 mSv。在主要射束照射範圍內的曝露器官(例如大腦),其致癌風險高於主要射束照射範圍外之器官(例如甲狀腺);不過8公分腦實質血流容積(PBV)檢查的致癌風險,則低於傳統的頭部電腦斷層掃描。
結論:
使用平板偵檢器電腦斷層(FDCT)施作腦實質血流容積(PBV)量測,不僅能節省劑量,亦具高再現性與高可靠性,且其吸收劑量亦低於使用256切電腦斷層施作腦血流灌注造影(CTP)檢查。此外,使用平板偵檢器電腦斷層做一站式造影服務(OSIS),還能有效減少病人換床次數,並可縮短完成手術的時間,因此將其應用在腦神經血管的診斷上是可推薦的。
關鍵字:
平板偵檢器電腦斷層、256切電腦斷層、腦實質血流容積、電腦斷層灌注造影、有效劑量、風險評估
目前在臨床的放射攝影和透視攝影中,平板偵檢器(flat-panel detector, FD)大都已經取代了標準X光底片、底片結合增感屏和影像增強管;由於該偵檢器具有高偵測量子效率(detective quantum efficiency, DQE)、高取像速率(frame rate)、高空間解析度(spatial resolution)及寬廣的動態範圍(dynamic range)等優勢,因此若能應用在血管攝影機組(angiographic suites)上,將使一站式造影服務(one-stop imaging service, OSIS)在處理腦神經血管等相關問題成為可行。在一站式造影服務中,腦血流動力學(即腦循環時間和腦實質的血流容積)可從一次造影中獲得,然而在廣泛應用該造影策略之前,其輻射劑量和相關的輻射危害仍是目前臨床主要關注的焦點。
目的:
本研究之目的在於評估使用平板偵檢器電腦斷層(flat detector computed tomography, FDCT)和256切電腦斷層(256-slice CT)施作腦血流灌注檢查,所累積之輻射劑量與致癌風險。
材料與方法:
本研究使用平板偵檢器電腦斷層(FDCT)和256切電腦斷層造影設備,搭配擬人假體和熱發光劑量計(thermo-luminescence dosimeter, TLD),進行輻射劑量的量測。在8公分和16公分腦實質血流容積(parenchymal blood volume, PBV)量測方面,使用數位減贅血管攝影(digital subtraction angiography, DSA)方式得之,其造影序列先以每秒60張的取像速率旋轉8秒,取得面罩影像,接續再以相同的取像速率旋轉8秒,取得充盈影像。在8公分影像的掃描參數如下:管電壓73 kV、管電流486 mA、曝射時間4.0 ms、照野48cm × 48cm;而16公分影像的掃描參數則為:管電壓72 kV、管電流540 mA、曝射時間4.1 ms、照野48cm × 48cm。8公分和16公分的掃描範圍分別為從蝶鞍到顱頂以及整個腦部。80顆熱發光劑量計(TLD)分別塞入擬人假體之頭部到骨盆等20個器官部位中。使用256切電腦斷層進行電腦斷層灌注造影(computed tomography perfusion, CTP)檢查,其掃描範圍和8公分腦實質血流容積(PBV)檢查相同,其掃描參數如下:切片厚度為5 mm、準直儀所開的範圍為64 × 1.25mm、照野大小為20cm × 20cm、管電壓為80 kV、管電流為250 mA、管電流時間乘積為100 mAs、管球的旋轉和曝射時間均為0.4 sec、掃描區間為1.5 sec然後持續1 min。本研究採用國際放射防護委員會(International Commission on Radiological Protection, ICRP)第103號報告中所建議之器官加權因子,進行有效劑量(effective dose)之計算;另採用游離輻射生物效應(Biological Effects of Ionizing Radiation, BEIR)第七份報告,做為致癌風險評估之依據。
結果:
本研究在8公分和16公分腦實質血流容積(PBV)檢查下,其有效劑量分別為0.87 ± 0.55 mSv和3.91 ± 0.78 mSv;而在8公分電腦斷層灌注造影(CTP)檢查時,其有效劑量則為2.77 ± 1.59 mSv。在主要射束照射範圍內的曝露器官(例如大腦),其致癌風險高於主要射束照射範圍外之器官(例如甲狀腺);不過8公分腦實質血流容積(PBV)檢查的致癌風險,則低於傳統的頭部電腦斷層掃描。
結論:
使用平板偵檢器電腦斷層(FDCT)施作腦實質血流容積(PBV)量測,不僅能節省劑量,亦具高再現性與高可靠性,且其吸收劑量亦低於使用256切電腦斷層施作腦血流灌注造影(CTP)檢查。此外,使用平板偵檢器電腦斷層做一站式造影服務(OSIS),還能有效減少病人換床次數,並可縮短完成手術的時間,因此將其應用在腦神經血管的診斷上是可推薦的。
關鍵字:
平板偵檢器電腦斷層、256切電腦斷層、腦實質血流容積、電腦斷層灌注造影、有效劑量、風險評估
Background:
The flat-panel detector (FD), for its unique advantages of high detective quantum efficiency (DQE), high frame rates and spatial resolution, and wider dynamic range, has replaced traditional X-ray film in most clinical imaging procedures. Angiographic suites equipped with the FD providing one-stop imaging service (OSIS) is becoming potentially feasible in management of cerebral neurovascular diagonosis. In OSIS procedure, cerebral hemodynamic information (i.e. cerebral circulation time and parenchymal blood volume, PBV) is obtained in one imaging session, but radiation dose and hazard relevant to the OSIS remains one of the major clinical concerns before it can be adopted as a standard imaging strategy.
Purpose:
The aim of our study was to estimate the effective organ doses and their related cancer risks involved in assessing cerebral perfusion examinations between using flat detector computed tomography (FDCT) and 256-slice CT.
Materials and Methods:
In our study, the absorbed organ dose measurements using FDCT and 256-slice CT with an anthropomorphic phantom and thermo-luminescence dosimeters (TLDs) were performed. In 8cm and 16cm wide cerebral parenchymal blood volume (PBV) measurements, we used one standard digital subtraction angiography (DSA) imaging protocol (one 8-second rotational mask, run with 60 frames/second followed by another 8-second rotational contrast medium filled, run with same frame rate) was used. The scanning parameters in the 8cm PBV exam were: 73 kV, 486 mA, 4.0 ms, 48cm×48cm field-of-view (FOV), and in 16cm PBV exam, they were: 72 kV, 540 mA, 4.1 ms, 48cm×48cm FOV. The scanning ranges in 8cm and 16cm exams were from sella turcica to convexity and whole brain coverage, respectively. Eighty TLDs divided into 20 sets were subsequently placed into different organs of the Rando phantom, from the head to the pelvis. In the use of 256-slice CT to measure CT perfusion (CTP), the scanning coverage was identical to that of the 8cm PBV exam, and its scanning parameters were: 5 mm slice thickness, 64×1.25mm collimation, 20cm×20cm FOV, 80 kV, 250 mA, 100 mAs, 0.4 second tube rotation and exposure time, scan interval of 1.5 seconds and lasted for 1 minute. In effective organ doses calculation, we employed the report no. 103 guidelines of the International Commission on Radiological Protection (ICRP-103), and adopted the guidelines of Biological Effects of Ionizing Radiation (BEIR) report VII for related cancer risk assessments.
Results:
In our study, we found that the effective doses were 0.87 ± 0.55 and 3.91 ± 0.78 mSv, respectively for PBV_8cm and PBV_16cm protocols in FDCT exams, and the effective dose in CTP_8cm measured was 2.77 ± 1.59 mSv using the 256-slice CT. Potential cancer risks for organs within the primary radiation beam (e.g., brain) were higher than those outside the primary beam (e.g., thyroid). Nevertheless, the potential cancer risk of PBV_8cm in the FDCT was relatively less, comparing with a head scan of a conventional 256-slice CT.
Conclusions:
Our study concluded that cerebral PBV measurement using the FDCT was dose-saving and with advantages of high reproducibility and reliability. The effective organ doses were lower than those involved in CT perfusion study using 256-slice CT. Moreover, the OSIS procedure can save time on patient transportation and shorten scanning time in PBV measurement, making the FDCT be a potential recommended for cerebral neurovascular diagnosis.
Keywords:
flat detector computed tomography (FDCT), 256-slice CT, parenchymal blood volume (PBV), computed tomography perfusion (CTP), effective dose, risk assessment
The flat-panel detector (FD), for its unique advantages of high detective quantum efficiency (DQE), high frame rates and spatial resolution, and wider dynamic range, has replaced traditional X-ray film in most clinical imaging procedures. Angiographic suites equipped with the FD providing one-stop imaging service (OSIS) is becoming potentially feasible in management of cerebral neurovascular diagonosis. In OSIS procedure, cerebral hemodynamic information (i.e. cerebral circulation time and parenchymal blood volume, PBV) is obtained in one imaging session, but radiation dose and hazard relevant to the OSIS remains one of the major clinical concerns before it can be adopted as a standard imaging strategy.
Purpose:
The aim of our study was to estimate the effective organ doses and their related cancer risks involved in assessing cerebral perfusion examinations between using flat detector computed tomography (FDCT) and 256-slice CT.
Materials and Methods:
In our study, the absorbed organ dose measurements using FDCT and 256-slice CT with an anthropomorphic phantom and thermo-luminescence dosimeters (TLDs) were performed. In 8cm and 16cm wide cerebral parenchymal blood volume (PBV) measurements, we used one standard digital subtraction angiography (DSA) imaging protocol (one 8-second rotational mask, run with 60 frames/second followed by another 8-second rotational contrast medium filled, run with same frame rate) was used. The scanning parameters in the 8cm PBV exam were: 73 kV, 486 mA, 4.0 ms, 48cm×48cm field-of-view (FOV), and in 16cm PBV exam, they were: 72 kV, 540 mA, 4.1 ms, 48cm×48cm FOV. The scanning ranges in 8cm and 16cm exams were from sella turcica to convexity and whole brain coverage, respectively. Eighty TLDs divided into 20 sets were subsequently placed into different organs of the Rando phantom, from the head to the pelvis. In the use of 256-slice CT to measure CT perfusion (CTP), the scanning coverage was identical to that of the 8cm PBV exam, and its scanning parameters were: 5 mm slice thickness, 64×1.25mm collimation, 20cm×20cm FOV, 80 kV, 250 mA, 100 mAs, 0.4 second tube rotation and exposure time, scan interval of 1.5 seconds and lasted for 1 minute. In effective organ doses calculation, we employed the report no. 103 guidelines of the International Commission on Radiological Protection (ICRP-103), and adopted the guidelines of Biological Effects of Ionizing Radiation (BEIR) report VII for related cancer risk assessments.
Results:
In our study, we found that the effective doses were 0.87 ± 0.55 and 3.91 ± 0.78 mSv, respectively for PBV_8cm and PBV_16cm protocols in FDCT exams, and the effective dose in CTP_8cm measured was 2.77 ± 1.59 mSv using the 256-slice CT. Potential cancer risks for organs within the primary radiation beam (e.g., brain) were higher than those outside the primary beam (e.g., thyroid). Nevertheless, the potential cancer risk of PBV_8cm in the FDCT was relatively less, comparing with a head scan of a conventional 256-slice CT.
Conclusions:
Our study concluded that cerebral PBV measurement using the FDCT was dose-saving and with advantages of high reproducibility and reliability. The effective organ doses were lower than those involved in CT perfusion study using 256-slice CT. Moreover, the OSIS procedure can save time on patient transportation and shorten scanning time in PBV measurement, making the FDCT be a potential recommended for cerebral neurovascular diagnosis.
Keywords:
flat detector computed tomography (FDCT), 256-slice CT, parenchymal blood volume (PBV), computed tomography perfusion (CTP), effective dose, risk assessment