動態脊椎內固定器混合固定對於腰椎鄰近椎節之生物力學效應─有限元素分析
Biomechanical Effect of Hybrid Stabilization with Dynamic Spinal Fixator on Adjacent Segment─Finite Element Analysis
摘要
動態脊椎內固定器 (dynamic spinal fixator, DSF)混合固定方式近年來被使用於臨床防治鄰近椎節退化問題,其方式是在退化椎節植入金屬內固定器進行融合,而其上一鄰近椎節以DSF保護。過去研究中發現,此混合固定方式不但能達到減輕疼痛還能改善失能狀況,但在體外測試則發現,混合固定方式比起單節植入金屬內固定器在鄰近椎節活動度上升的較多,可見混合固定方式對鄰近椎節之影響仍有爭議性。因此,本研究將探討動態內固定器混合固定方式相對單節植入金屬內固定器之生物力學效應。
本研究以有限元素軟體ANSYS建立完整腰椎模型L1-5,接著模擬臨床常用的經椎孔腰椎椎體間融合術,並建立三種不同組合植入物模型,包括:(1) L4-5植入金屬內固定器作融合 (rigid fixator, RIG)、(2) L4-5植入金屬內固定器加上L3-4植入鈦合金DSF (Awesome rod system, AWE)、(3) L4-5植入金屬內固定器加上L3-4植入聚醚醚酮 (Polyether-ether-ketone, PEEK) DSF (modified-Awesome rod system, M-AWE)。接著採用混合控制,在L1上方施加150牛頓預壓力和純力矩直到總活動度達前彎16°、後彎5°、側彎5°、扭轉5°。
AWE、M-AWE整體腰椎勁度較RIG高,前彎和側彎時,椎間盤最大應力也較RIG高出至少20%,而後彎時,小面關節接觸力則高出RIG至少65%。本研究發現,只要植入物植入之椎節數量增加,整體的勁度就會上升,儘管穩定了兩個椎節,反而使新產生的鄰近椎節活動度、椎間盤應力、小面關節接觸力增加幅度更大。
本研究以有限元素軟體ANSYS建立完整腰椎模型L1-5,接著模擬臨床常用的經椎孔腰椎椎體間融合術,並建立三種不同組合植入物模型,包括:(1) L4-5植入金屬內固定器作融合 (rigid fixator, RIG)、(2) L4-5植入金屬內固定器加上L3-4植入鈦合金DSF (Awesome rod system, AWE)、(3) L4-5植入金屬內固定器加上L3-4植入聚醚醚酮 (Polyether-ether-ketone, PEEK) DSF (modified-Awesome rod system, M-AWE)。接著採用混合控制,在L1上方施加150牛頓預壓力和純力矩直到總活動度達前彎16°、後彎5°、側彎5°、扭轉5°。
AWE、M-AWE整體腰椎勁度較RIG高,前彎和側彎時,椎間盤最大應力也較RIG高出至少20%,而後彎時,小面關節接觸力則高出RIG至少65%。本研究發現,只要植入物植入之椎節數量增加,整體的勁度就會上升,儘管穩定了兩個椎節,反而使新產生的鄰近椎節活動度、椎間盤應力、小面關節接觸力增加幅度更大。
Hybrid stabilization with dynamic spinal fixator (DSF) has recently been used to protect adjacent disc. The approach is degenerated disc fused by rigid fixation and upper segment stabilized by DSF. Past studies reported that hybrid stabilization with DSF can not only decrease pain but improve spinal instability. However, the in vitro test addressed iatrogenic effects with significantly increase of range of motion (ROM) in adjacent segments. The biomechanical effect of hybrid stabilization on adjacent segments remains controversy. This study aimed to find out the biomechanical effect of hybrid stabilization compared with rigid fixation.
Intact lumbar spine (L1-5) model was created by finite element software ANSYS, then the transforaminal lumbar interbody fusion model was established. Three models with different combination of spinal fixator were constructed, with rigid fixator at L4-5 (rigid fixator, RIG), rigid fixator at L4-5 and Ti-Alloy DSF at L3-4 (Awesome rod system, AWE), and with rigid fixator at L4-5 and PEEK (Polyether-ether-ketone) DSF at L3-4 (modified-Awesome rod system, M-AWE). A hybrid control, with 150N preload, until ROM reach flexion 16°, extension 5°, lateral bending 5°, and rotation 5°, were applied to the models, respectively.
The lumbar stiffness of AWE and M-AWE were found higher than RIG in all direction of movements. The AWE and M-AWE had higher maximum disc stress in adjacent disc than the RIG in flexion and lateral bending for at least 20%. In extension, the facet contact force was 65% higher in AWE and M-AWE than in RIG.
The study found that the more segments stabilized would raise stiffness of lumbar spine. Even the stabilized segments become more stable, the adjacent segment would have higher ROM, disc stress, and facet contact force.
Intact lumbar spine (L1-5) model was created by finite element software ANSYS, then the transforaminal lumbar interbody fusion model was established. Three models with different combination of spinal fixator were constructed, with rigid fixator at L4-5 (rigid fixator, RIG), rigid fixator at L4-5 and Ti-Alloy DSF at L3-4 (Awesome rod system, AWE), and with rigid fixator at L4-5 and PEEK (Polyether-ether-ketone) DSF at L3-4 (modified-Awesome rod system, M-AWE). A hybrid control, with 150N preload, until ROM reach flexion 16°, extension 5°, lateral bending 5°, and rotation 5°, were applied to the models, respectively.
The lumbar stiffness of AWE and M-AWE were found higher than RIG in all direction of movements. The AWE and M-AWE had higher maximum disc stress in adjacent disc than the RIG in flexion and lateral bending for at least 20%. In extension, the facet contact force was 65% higher in AWE and M-AWE than in RIG.
The study found that the more segments stabilized would raise stiffness of lumbar spine. Even the stabilized segments become more stable, the adjacent segment would have higher ROM, disc stress, and facet contact force.