手術後病人小量噴霧器的更換頻率對於噴霧器汙染和口咽菌種移生之影響

The Effect of Replacing Small-Volume Nebulizers on Contamination and Oral Colonization after Patient Surgery

檔案

全文未公開。

日期

2015

類型

語言

中文

摘要

臨床上常使用小量噴霧器做為呼吸道疾病的給藥工具,其可霧化藥物至0.5~10μm,透過吸入方式進入下呼吸道達到祛痰或治療呼吸道炎症反應之療效。然而,噴霧器照護不當可能因外源性途徑造成噴霧器汙染,增加病人呼吸道菌種移生或呼吸道感染之風險。目前文獻對於噴霧器的更換頻率建議不一,各家醫院的噴霧器照護做法也不一致,因此本研究目的在探討噴霧器每日、每兩日及每三日更換之噴霧器汙染和個案口咽菌種移生差異。
本研究採隨機對照試驗(Randomized controlled trial)研究法,自2014年7月9日至2015年2月2日止,於北部某醫學中心之綜合外科病房進行研究,納入20歲以上且使用小量噴霧治療之病人為研究對象,經個案同意後,隨機分配至每日、每兩日及每三日更換噴霧器,也在噴霧治療前24小時採集個案第一套口咽檢體,噴霧治療第四天後再採集第二套口咽檢體,並觀察個案於研究期間是否出現呼吸道感染症狀或徵象。
研究結果:收案三組各70人,採驗個案之噴霧器檢體共525個。每日更換噴霧器組汙染率為14.1%,顯著低於每兩日更換組(24.3%)和每三日更換組(25.6%) (p=0.008),個案之口咽移生率以每三日更換組為9.42‰最低,其次為每日更換組(10.10‰)和每兩日更換組(14.40‰),但三組未達統計上顯著差異(p=0.721)。此外,所有研究個案均無肺炎發生,在呼吸道感染症狀和噴霧治療後胸部X光有新的漸進性浸潤方面,三組個案也無統計上顯著差異(p>0.05)。多變項分析發現噴霧治療每增加一天,口咽移生風險會增加11% (95%信賴區間1.02-1.22,p=0.046),以及噴霧器嚴重汙染個案(菌量>100 CFU/ML)的口咽移生風險是乾淨噴霧器(菌量<10 CFU/ML)個案的5.33倍(95%信賴區間1.06-26.83,p=0.042)。
結論:二天以上更換小量噴霧器的汙染率會顯著增加,但口咽移生率並無統計上顯著差異。此外,嚴重汙染的噴霧器將會增加口咽移生風險。
Nebulizers are equipments widely used in clinical for treating the disease of respiratory tract, which can aerosolize medicines to 0.5-10μm to reaching the lower respiratory tract, so as to induce expectorant action or treat the inflammatory of respiratory tract. However, nebulizers can be comtaminated by the exogenous route owing to improper care, so that increase risks of oral colonization or respiratory tract infection. Currently, suggestions of papers about the frequency of replacing nebulizers are vary wide, and guidelines of hospitals on nebulizers practices are also inconsistent.Therefore,this study aim at investigating the effect of replacing nebulizers ( at 24,48 or 72 hours ) on contamination and oral colonization.
A randomized controlled trial study is conducted from July 9, 2014 to February 2, 2015 in a surgical ward of a medical center in Taipei. During the study period, include above 20 years old people with steam inhalation treatment. After participants agree to join the trail, randomise cases into 24, 48 or 72 hours group. In the protocol, researcher collect first cultures of oropharyngeal at 24 hours before steam inhalation, and collect second cultures on the forth day. Futhermore, follow cases whether happen respiratory tract infection synptoms or signs during the study period.
Results show that 70 people in each group, total 525 specimens of nebulizes are collected. The rate of contaminated nebulizers at 24 hours is 14.1%,
which is significantly higher (p = 0.008) than 48hours (24.3%) and 72 hours
(25.6%). The rate of oral colonization at 72 hours is 9.42‰, which is the lowest
of three groups( 24 hours is 10.10‰, 48 hours is 14.40‰), but not reach
significant difference in statistics (p = 0.721). Besides, no patient occur pneumonia in the study period, as well as from symptoms of respiratory tract infections and progressive infiltration on X-ray after steam inhalation sides, that not reach significant difference in statistics(p>0.05). In multivariate statistical analysis show that steam inhalation increase by one day with oral colonization risks increased by 11%. Risks of oral colonization in severe contaminated nebulizers ( the counting of bacterial colony are over 100 CFU/ML ) are 5.33 times (95% CI: 1.06-26.83, p = 0.042) as much as clean nebulizers ( the counting of bacterial colony are less than 10 CFU/ML ).
Conclusions show that replacing nebulizers more than two days will increase contaminated rates significantly, but rates of oral colonization are not different
significantly in statistics. In addition, rates of oral colonization will increase if nebulizers are contaminated severely.