可偵測多根骨釘鬆緊度之無線被動感測技術開發

Development of passive wireless sensing technology for tightness monitoring on multiple bone screw

檔案

日期

2015

類型

語言

中文

摘要

椎弓根骨螺釘固定系統在骨科中的脊椎融合術中,為最常用的一種固定技術。脊椎融合術目的即為幫助病人以固定椎骨的方式減輕疼痛,目前醫療上所使用於監控椎弓根骨螺釘固定狀況僅有X光的方法,但病人在開刀後六周才再次做X光照射,短期間內無法提早得知骨釘螺絲鬆緊度訊息。手術時,從背部將椎弓根螺絲鎖在每對椎骨裡,再由連桿將所有的螺絲頭鎖在一起以維持椎骨位置,一般手術所使用的數量是一次6~8根,所以能做到多根螺絲的檢測為必要。
本研究配合一無線被動螺絲鬆緊度之感測裝置,以電感電容(LC)迴路原理,讀取共振頻率值,再透過LC設計達到六個不同頻段以監測不同根螺絲之鬆動狀況,目前已設計出六個頻段且彼此不重疊。
本研究也建立一多層矩形電感和寄生電容之模擬技術,透過模擬達到設計不同頻段的要求,在驗證寄生電容的準確性一開始只考慮整體線圈電容值卻沒有考慮兩層線圈之間存在之電容值,經過幾次的修正初始模擬公式配合實驗結果的量測已經大大降低模擬公式的誤差性,寄生電容之誤差從原本的11.1~78%降低到1.5~13.2%,寄生電容誤差隨著垂直圈數增大的原因在於尚未考慮到單層線圈對較遠層數線圈間存在之電容值,若對此加以修正相信準確性會更高。
無線量測方面共振頻率之量測誤差值會從0.1~2.12%,共振頻率變化量則是1.2~5.15%,在感應距離方面之測試,透過不同天線端以及電感的配合,量測距離也可以逐漸加大,目前設計之矩形電感,以長寬35mm*23mm、平行圈數3圈、垂直圈數2圈已能讀取到4.5cm之距離。未來配合天線設計與電感配合希望能將感測距離拉到8cm以上。
Pedicle screw is the most common fixation technique in orthopedic spine fusion surgery. The purpose of spine fusion surgery is to help patients to relieve their pain by spine fixation. Currently, X-ray is the only way to monitor fixing condition of pedicle screws, but the patients are allowing to take X-ray after six weeks of their surgeries, so we can’t get the information of the screw fixing during the short period. Generally, we use 6-8 pedicle screws in one surgery, so to monitor the condition of multiple screws is necessary.
By using LC circuit principle with a wireless passive screw tightness sensing device, we read the resonance frequency and then we reach the goal of 6 frequency bands through different LC designs in order to observe the loosening situation of different screws, 6 frequency bands have been designed and do not overlap each other.
In this study we established a multi-layered rectangular inductor and parasitic capacitance simulation model, we can design 6 different frequency bands through simulation model.
In the accuracy verification of the parasitic capacitance model, we consider only the capacitance of the whole coils but without the capacitance between 2 different layers in the beginning. Through several times modifications, we lower down the error from the original 11.1%~78% to 1.5%~13.2%. The reason why error of parasitic capacitance increases with the vertical turns is because we don’t consider the capacitance between single layers to rather far layers, if we correct this situation the accuracy of the whole model will be even higher.
Through wireless measurement, the error range of the resonance frequency is 0.1~2.12%, and 1.2~5.15% for the resonance frequency change. As for the sensing distance, we can higher the distance by the coordination of different antennas and coils. So far, we design a multi-layered rectangular inductor with its length and width 35*23mm, horizontal turns 3 and vertical turns 2, and the sensing distance is 4.5cm. We hope to reach the goal of 8cm in the future though different antenna and coil design.

所屬典藏

機械工程學系
顯示完整詮釋資料 統計資料

引用此條目

黃得裕 (2015). 可偵測多根骨釘鬆緊度之無線被動感測技術開發. https://thesis.lib.nycu.edu.tw/handle/20.500.14371/61606

BibTeX RIS

引用文字依 APA 第 7 版格式自動產生,使用前請再核對。

分享

永久網址:https://thesis.lib.nycu.edu.tw/handle/20.500.14371/61606