穿透式與反射式X光管劑量分佈差異之研究
Comparison of dose distribution between transmission and reflection X-ray tube
摘要
穿透式X光管由於其薄靶的特殊設計,其輻射輸出特性有別於傳統反射式X光管,本研究之目的為比較反射式及穿透式X光管於相同射質及實驗條件下,其輻射輸出劑量、能譜,以及相對劑量分佈差異。於輻射輸出劑量量測中,反射式及穿透式X光管其劑量輸出隨管電流×曝露時間變化呈現良好線性關係(R2=1),與電壓變化亦具正相關性(R2≧0.999),然而在相同實驗條件及照射參數下,穿透式X光管的劑量輸出為反射式的2.64 - 3.21倍,故穿透式X光管具有較高之輻射輸出產率。於能譜量測中,穿透式X光管之K特性輻射約佔總能譜22.96 %,反射式X光管之K特性輻射約佔總能譜8.91 %,結果顯示穿透式X光管具有較顯著的K特性輻射輸出。反射式及穿透式X光管使用PTW 31010於水中相對劑量量測中,反射式X光管於Y軸劑量剖面可觀測到跟效應的現象,而穿透式X光管於X、Y軸劑量剖面呈現對稱性的不平坦分佈,此分佈可能是由於靶材設計或輻射前向輸出特性所造成;於水中百分深度劑量量測中,50 mm深度以前穿透式X光管具較顯著之劑量下降梯度,此較顯著的劑量下降梯度是由於穿透式X光管低能K特性輻射為主的能譜,產生較顯著衰減所造成。
本研究亦使用TLD-100H、XR-QA2,以及蒙地卡羅模擬驗證PTW 31010游離腔於水中所測得之相對劑量,於劑量剖面量測中PTW 31010、TLD-100H、XR-QA2,以及蒙地卡羅模擬結果在90 %內劑量區間誤差均小於3 %,而半影區由於相異劑量計之延軸直徑差異產生之劑量平均效應,所造成之顯著劑量誤差已由蒙地卡羅模擬驗證;於水中百分深度劑量研究中,PTW 31010於量測點往空腔淺層偏移1.7 mm後,相較於Markus 23343量測及蒙地卡羅模擬結果於5 mm深度前誤差小於3 %,於5 mm深度後誤差小於2 %,而TLD-100H使用蒙地卡羅修正相異深度之能依性後,相較於PTW 31010量測結果,於5 mm深度後誤差小於2 %。
本研究亦使用TLD-100H、XR-QA2,以及蒙地卡羅模擬驗證PTW 31010游離腔於水中所測得之相對劑量,於劑量剖面量測中PTW 31010、TLD-100H、XR-QA2,以及蒙地卡羅模擬結果在90 %內劑量區間誤差均小於3 %,而半影區由於相異劑量計之延軸直徑差異產生之劑量平均效應,所造成之顯著劑量誤差已由蒙地卡羅模擬驗證;於水中百分深度劑量研究中,PTW 31010於量測點往空腔淺層偏移1.7 mm後,相較於Markus 23343量測及蒙地卡羅模擬結果於5 mm深度前誤差小於3 %,於5 mm深度後誤差小於2 %,而TLD-100H使用蒙地卡羅修正相異深度之能依性後,相較於PTW 31010量測結果,於5 mm深度後誤差小於2 %。
Transmission-target X-ray tube due to its special thin target design,
the radiation output characteristics of the transmission-target X-ray tube are different from those of the traditional reflection-target X-ray tube. The aims of this study were to compare the differences of radiation output
dose、spectrum and relative dose distribution between these two X-ray tubes under the same conditions. For these two X-ray tubes, the dose output and the mAs appeared to have good linear relations(R2=1); the dose output and kVp variations also had positive relations(R2≧0.999). However, under the same parameters, the dose output of transmission-target X-ray tube was 2.64-3.21 times higher than the reflection-target X-ray tube, implying that the transmission-target system had a higher X-ray production rate. The K characteristic radiations reach 22.96% and 8.91% of the spectrum in transmission-target and reflection-target, respectively. The spectrum measurements showed that the transmission-target X-ray tube had more obvious output of K characteristic radiation. The relative dose distribution in water were measured for these two X-ray tubes using PTW 31010 ion chamber, the results of dose profile showed that the reflection-target X-ray tube had heel effect along Y-axis, and the transmission-target X-ray tube had uneven symmetry distribution along both axes due to its metal target design or the characteristics of radiation output forward from the metal target; The results of percentage depth dose showed that transmission-target X-ray tube had greater dose gradient attenuation than reflection-target X-ray tube above 50 mm depth in water. The transmission-target X-ray tube spectrum mainly composed of low energy K characteristic radiation, resulting in more attenuation above 50 mm depth.
In this study, the measured relative doses by PTW 31010 ion chamber were also compared with TLD-100H、XR-QA2 film and Monte Carlo simulation. The results of dose profile showed that the differences between all measurements and simulations were less than 3 % inside the 90 % dose region, and the significant dose differences in penumbra due to dose average effect which caused by different dosimeter diameter were verified by Monte Carlo simulations.
The percentage depth doses measured by PTW 31010 ion chamber with the point of measurement shifted to shallower depth by 1.7 mm was compared with Markus 23343 measurements and Monte Carlo simulations. The results of percentage depth dose showed that the differences between ion chamber measurements and simulations were less than 3 % in the first 5 mm, and less than 2 % below 5 mm in water; After TLD-100H corrected the energy dependence varied with depth by Monte Carlo, the differences between TLD-100H and PTW 31010 ion chamber measurements were less than 2% below 5 mm in water.
the radiation output characteristics of the transmission-target X-ray tube are different from those of the traditional reflection-target X-ray tube. The aims of this study were to compare the differences of radiation output
dose、spectrum and relative dose distribution between these two X-ray tubes under the same conditions. For these two X-ray tubes, the dose output and the mAs appeared to have good linear relations(R2=1); the dose output and kVp variations also had positive relations(R2≧0.999). However, under the same parameters, the dose output of transmission-target X-ray tube was 2.64-3.21 times higher than the reflection-target X-ray tube, implying that the transmission-target system had a higher X-ray production rate. The K characteristic radiations reach 22.96% and 8.91% of the spectrum in transmission-target and reflection-target, respectively. The spectrum measurements showed that the transmission-target X-ray tube had more obvious output of K characteristic radiation. The relative dose distribution in water were measured for these two X-ray tubes using PTW 31010 ion chamber, the results of dose profile showed that the reflection-target X-ray tube had heel effect along Y-axis, and the transmission-target X-ray tube had uneven symmetry distribution along both axes due to its metal target design or the characteristics of radiation output forward from the metal target; The results of percentage depth dose showed that transmission-target X-ray tube had greater dose gradient attenuation than reflection-target X-ray tube above 50 mm depth in water. The transmission-target X-ray tube spectrum mainly composed of low energy K characteristic radiation, resulting in more attenuation above 50 mm depth.
In this study, the measured relative doses by PTW 31010 ion chamber were also compared with TLD-100H、XR-QA2 film and Monte Carlo simulation. The results of dose profile showed that the differences between all measurements and simulations were less than 3 % inside the 90 % dose region, and the significant dose differences in penumbra due to dose average effect which caused by different dosimeter diameter were verified by Monte Carlo simulations.
The percentage depth doses measured by PTW 31010 ion chamber with the point of measurement shifted to shallower depth by 1.7 mm was compared with Markus 23343 measurements and Monte Carlo simulations. The results of percentage depth dose showed that the differences between ion chamber measurements and simulations were less than 3 % in the first 5 mm, and less than 2 % below 5 mm in water; After TLD-100H corrected the energy dependence varied with depth by Monte Carlo, the differences between TLD-100H and PTW 31010 ion chamber measurements were less than 2% below 5 mm in water.