可偵測多根骨科固定螺絲鬆緊度之被動無線感測裝置之優化

Improvement of passive wireless sensors for tightness monitoring at different locations on orthopedic fixation screw system

檔案

日期

2016

類型

語言

中文

摘要

椎弓根螺釘系統常見於骨科脊椎融合手術中,係以固定椎骨的方式來使患者之椎骨癒合,而骨頭癒合情況是否良好,則與椎弓根螺釘之固定狀況有直接關係。但因術後椎骨之融合需要一段時間,目前臨床上,大約都是術後六至八周才進行x光檢測,以檢視患者之椎骨融合情形,而椎弓根螺釘若在這段期間內發生鬆動之情形,以x光照片是不易判斷出來的,但卻有很大機率使癒合之情形不如預期。
本實驗室在104年度已提出一單頻段之可偵測椎弓根螺釘鬆緊度之被動無線感測裝置。可藉由量測感測器之共振頻率的變化,得知椎弓根螺釘鎖緊力的狀況,但其感應距離僅有30 mm,而在實際應用上感應距離至少需40-80 mm,且需同時量測體內多根骨釘之固定狀況,因此仍具改善之空間,為此本實驗室在104年也已透過電感和其寄生電容之模擬,設計出不同頻段之電感,然其性能仍須確認。
本研究配合先前之設計,製造出六根不同頻段之感測器,可同時監控體內不同位置之椎弓根螺釘固定狀況,並藉由讀取端和感測端電感之最佳化之實驗,將其無線感測距離分別優化至56-72 mm,而無線感應距離最遠之感測器(72 mm),配合椎骨鎖進豬骨內後,其鎖緊力之變化量(7.8N)與直接量測的鎖緊力變化(7.4N)相比,誤差約0.4N,顯示在術後若經由共振頻率的變化來估算鎖緊力之變化是具有良好之準確性,而本研究也對多頻段之感測器進行干擾測試,以確認不同頻段之感測器間在不同擺放位置下彼此間產生的干擾影響,本研究有望突破以往x光照片無法判斷體內螺絲鬆緊度之困境,幫助醫生更有效率的掌握患者術後狀況,在螺絲發生鬆動現象時能夠及早進行處理。
Pedicle screw is a typical fixation device in spine fusion surgery for osteoporosis patients.
Spine fusion condition is highly depends on whether pedicle screw is tightly fixed on spine, if not, spine fusion will not work properly. But it’s normally about six-eight weeks to take first radiologic images after the surgery to observe the spine fusion condition. And pedicle screw’s fixed condition can’t be observe from radiologic images. Hence we develop a passive, wireless tightness sensor to improve conventional monitoring method. It can be read without the battery installed and by measuring changes of frequency of the sensor, we can know the fixing condition of the pedicle screw.
Our lab has proposed a passive wireless sensor with one band . By using LC circuit principle, we can use the change of frequency to estimate the fixing force of pedicle screw through this device. But the wireless sensing distance in this design is only 30 mm and one detect band. It’s not practical for clinical applications. For this reason, we also through simulation to design different band of sensor .However, it’s performance still need to be confirmed .
In this research, we successfully using the design to develop six sensors with different bands. It can monitor different pedicle screw on different locations simultaneously. And by optimizing the reading and sensing antenna, the wireless sensing distance of these sensors are 56-72 mm. The sensor which wireless sensing distance is 72 mm will combine pedicle screw locked into pork bone, and the errors of force variation estimated from measured frequency(7.4 N) compare to directly measured force variation (7.8 N) is 0.4 N. It shows that via measured frequency to estimate pedicle screw’s locking force has good accuracy. And we also do the interference test with different sensors to confirm the influences of different placement. We hope this sensor can help physician monitor pedicle screw’s locking force more efficiently. If physician find pedicle screw loosening after surgery, it can be treated as soon as possible.