應用於無導線式心臟節律器之圓管狀駐極體能量擷取器之製作與測試
Fabrication and Testing of Cylindrical Vibrational Energy Harvester for Leadless Intracardiac Pacemakers
摘要
各類植入式裝置中,心臟節律器可控制與調節患者心律,其中無導線式裝置用以降低手術風險與複雜度,使用傳輸系統沿著股靜脈進入心臟心室內並植入固定。但此類醫療器材的生命週期會因電池的電量而有上限,最終需要進行開胸手術替換。若發展出半永久性的能量擷取器,植入心臟後將改善此問題。近年研究已發表的心內能量擷取器多以振盪質量塊作為能量轉換源,然而此方式需要佔據較大體積,而使裝置無法微小化或與處理晶片整合,心臟節律器或電路需要另外以別的植入裝置放置在體內。本文延續實驗室經驗設計圓管狀靜電式能量擷取器並提出微小化方案,分別以不鏽鋼及鈦金屬作為內外電極材料,裝置管狀中間若為空心,則可容納節律器的處理電路。
裝置包含內電極、圓管狀駐極體、外電極,使用同軸封裝。不鏽鋼電極裝置的內電極為空心不鏽鋼管,外電極使用遮罩與濕蝕刻製作成具孔洞的不鏽鋼管。鈦電極大尺寸裝置與微縮裝置的內電極皆實心鈦棒,鈦電極大尺寸裝置的外電極外徑與不鏽鋼電極裝置相同,鈦電極微縮裝置的外電極則選用較小的外徑,並濕蝕刻至適當厚度再鑽孔。
血壓可經由孔洞施加在駐極體外壁,使駐極體產生週期性的向心或離心位移,此時兩電極上出現感應電荷,再接上一負載電阻在兩電極之間,即產生輸出功率。圓管狀PDMS是使用不鏽鋼管或高速鋼(HSS)管浸泡於PDMS中,取出後固化而成,再沉積 1"μ" m厚的 Parylene C後即為駐極體。將駐極體置於另一不鏽鋼管中,以線徑 36awg的漆包線為線電極,以不同數值的高壓片電阻限流。表面電荷的穩定度,在放電實驗完成 459天後,表面電位約 53.5%。將內外電極與駐極體以PDMS封裝後,置於不同流體環境中量測。操作頻率 1.0Hz,不鏽鋼電極裝置在空氣環境下量測得負載 5 GΩ、施壓 - 40mmHg,峰值功率約 1.56"μ" W,在食鹽水環境下負載 1GΩ峰值功率為 0.273"μ" W。鈦電極大尺寸裝置在負壓 100mmHg、頻率 1Hz時,空氣中負載 23 GΩ峰值輸出約 1.8"μ" W,食鹽水中負載 43GΩ峰值輸出約 0.7"μ" W,自來水中訊號去雜訊後,負載 63G峰值輸出約 1.23"μ" W。
裝置包含內電極、圓管狀駐極體、外電極,使用同軸封裝。不鏽鋼電極裝置的內電極為空心不鏽鋼管,外電極使用遮罩與濕蝕刻製作成具孔洞的不鏽鋼管。鈦電極大尺寸裝置與微縮裝置的內電極皆實心鈦棒,鈦電極大尺寸裝置的外電極外徑與不鏽鋼電極裝置相同,鈦電極微縮裝置的外電極則選用較小的外徑,並濕蝕刻至適當厚度再鑽孔。
血壓可經由孔洞施加在駐極體外壁,使駐極體產生週期性的向心或離心位移,此時兩電極上出現感應電荷,再接上一負載電阻在兩電極之間,即產生輸出功率。圓管狀PDMS是使用不鏽鋼管或高速鋼(HSS)管浸泡於PDMS中,取出後固化而成,再沉積 1"μ" m厚的 Parylene C後即為駐極體。將駐極體置於另一不鏽鋼管中,以線徑 36awg的漆包線為線電極,以不同數值的高壓片電阻限流。表面電荷的穩定度,在放電實驗完成 459天後,表面電位約 53.5%。將內外電極與駐極體以PDMS封裝後,置於不同流體環境中量測。操作頻率 1.0Hz,不鏽鋼電極裝置在空氣環境下量測得負載 5 GΩ、施壓 - 40mmHg,峰值功率約 1.56"μ" W,在食鹽水環境下負載 1GΩ峰值功率為 0.273"μ" W。鈦電極大尺寸裝置在負壓 100mmHg、頻率 1Hz時,空氣中負載 23 GΩ峰值輸出約 1.8"μ" W,食鹽水中負載 43GΩ峰值輸出約 0.7"μ" W,自來水中訊號去雜訊後,負載 63G峰值輸出約 1.23"μ" W。
Among all kinds of medical implants, cardiac pacemakers are especially used for maintain and control beat rate of human heart. Generally, surgeons utilize the transmission system to place leadless pacemkers through femeral vein to ventrical, in order to reduce the rick and complexity of surgical operation. The advantage of intracardiac harvesters is that no open-chest surgery is needed so risks can be significantly reduced while the lifetime were limited by batteries.
However, published intracardiac energy harvesters often employ volumetric converting mechanism such as oscillating mass. Therefore, most volume of the harvesters could be occupied. We developed several harvester structures, the structure of each harvester includes inner electrode, electret membrane and outer electrode. Components are coaxially assembled. To allow blood pressure applying to the flexible electret membrane, we etch holes unifomally through the surface of outer electrode. The blood pressure variation moves the electret membrane in and out to induce charge variation on the two collecting electrodes so that output current can flow through external load. Then 1 μm parylene was deposited on PDMS tube surfaces as an electret. The PDMS/parylene tube was inserted into another stainless or HSS tube and a corona line discharge. The measured charge stability is good after 459 days decay, the surfaces potential is still 53.5%.
Then the three tubular components were assembled and sealed with PDMS. The peak power measured of the prototype which was made of stainless steel at 5 GΩ load, -40 mmHg pressure variation at 1.0 Hz is 1.56μW in air, while 0.273μW measured in saline.One output power of prototype made of titanium under -100mmHg pressure variation at 1.0Hz is about 1.8μW measured in air at 23GΩ load, another measured in saline is about 0.7μW at 43GΩ load, and the other measured in water is about 1.23μW at 63GΩ load.
However, published intracardiac energy harvesters often employ volumetric converting mechanism such as oscillating mass. Therefore, most volume of the harvesters could be occupied. We developed several harvester structures, the structure of each harvester includes inner electrode, electret membrane and outer electrode. Components are coaxially assembled. To allow blood pressure applying to the flexible electret membrane, we etch holes unifomally through the surface of outer electrode. The blood pressure variation moves the electret membrane in and out to induce charge variation on the two collecting electrodes so that output current can flow through external load. Then 1 μm parylene was deposited on PDMS tube surfaces as an electret. The PDMS/parylene tube was inserted into another stainless or HSS tube and a corona line discharge. The measured charge stability is good after 459 days decay, the surfaces potential is still 53.5%.
Then the three tubular components were assembled and sealed with PDMS. The peak power measured of the prototype which was made of stainless steel at 5 GΩ load, -40 mmHg pressure variation at 1.0 Hz is 1.56μW in air, while 0.273μW measured in saline.One output power of prototype made of titanium under -100mmHg pressure variation at 1.0Hz is about 1.8μW measured in air at 23GΩ load, another measured in saline is about 0.7μW at 43GΩ load, and the other measured in water is about 1.23μW at 63GΩ load.