通過橈動脈位置檢測及類比訊號處理以提高非侵入式光學血壓陣列感測器之測量精度
Towards Maximizing Measuring Accuracy of a Optical Blood Pressure Non-invasive Cuffless Array Sensor via Radial Artery Position Detection and Analog Signal Processing
摘要
本論文目的為設計一新型光學式血壓感測器用於監測連續性血壓,此血壓感測器引入光體積描述法以檢測血管內寫容量體積變化並且計算血壓值。此系統採用多波長發光二極體,其波長為530 nm、660 nm、940 nm,此感測器由4組發光二極體及4個感光器陣列所組成。前端電路包含前置放大器、帶通濾波器、可編程增益放大器、LED驅動電路、微處理器和無線模組,帶通濾波器之頻帶範圍為0.3Hz~7.2Hz,可編程增益放大器則有8個不同增益可作調變,脈搏位置偵測演算法可偵測使用者橈動脈位置並將偵測到之連續血壓和用戶統計資料及結果分析顯示於電腦上。在實驗中,受測者以本論文之血壓感測器量測結果將與市售血壓感測器OMRON HEM-7310做比較。實驗結果的最大誤差為±6 mmHg,小於±8 mmHg,符合“醫療器材促進發展協會”(AAMI)的要求,小於±8 mmHg。
A new optical PPG sensor for continuous blood pressure (BP) monitoring is successfully developed by this thesis. BP device introduces the principle of photoplethysmography (PPG) to detect intravascular blood volume changes and calculate blood pressure. The real-time system adopts LEDs with muti-wavelength of 530, 660 and 940 nm. The analog front end (AFE) circuit includes a preamplifier, a band-pass filter, a programmable gain amplifier (PGA), LED driver, a microprocessor unit (MCU), and a wireless module. The pass-band frequency of filter is from 0.3 to 7.2 Hz. The PGA of adjustable gain has 8 channels to adjust. A computer is also used to display the continuous BPs, detect the pulse position and record statistical analysis/results for users. As results, 10 subjects in the experimental validation, in which the obtained BPs are compared with the results from a commercial BP monitor by OMRON HEM-7310. The maximum error of experimental results is ± 6 mmHg, which is less than ±8 mmHg conforming to the requirement by the Association for the Advancement of Medical Instrumentation (AAMI).