基於遠距離體積描述儀的無袖袋影像血壓計
Cuffless Image Blood Pressure Monitor by Remote-PPG
摘要
根據台灣高血壓學會訊息指出,全台 624 萬高血壓風險族群裡,每 5 人中僅 1 人每周量
測血壓。若想要有效控制高血壓,必須先養成固定量測血壓之習慣。並且高血壓風險族群若
能在輕鬆、無壓力的環境下量測血壓,則可以避免像白袍高血壓等造成血壓誤判之症狀發生。
在此,本文提出無袖袋/非接觸式的影像血壓計。藉由一般的網路攝影機達到輕鬆、無壓力的
狀況下監測血壓,並且改善傳統手臂式/手腕式袖袋血壓計無法連續監控血壓的問題。其採用
的方式為利用網路攝影機擷取脈波傳導時間,並配合生理資料庫及機器學習之方式生成血壓
模型及預測。經實驗後,其收縮壓 RMSE 可達到 4.56 mmHg;舒張壓 RMSE 可達到 3.97
mmHg,且其精準度達到英國高血壓協會所認定的 grade C 水準。
測血壓。若想要有效控制高血壓,必須先養成固定量測血壓之習慣。並且高血壓風險族群若
能在輕鬆、無壓力的環境下量測血壓,則可以避免像白袍高血壓等造成血壓誤判之症狀發生。
在此,本文提出無袖袋/非接觸式的影像血壓計。藉由一般的網路攝影機達到輕鬆、無壓力的
狀況下監測血壓,並且改善傳統手臂式/手腕式袖袋血壓計無法連續監控血壓的問題。其採用
的方式為利用網路攝影機擷取脈波傳導時間,並配合生理資料庫及機器學習之方式生成血壓
模型及預測。經實驗後,其收縮壓 RMSE 可達到 4.56 mmHg;舒張壓 RMSE 可達到 3.97
mmHg,且其精準度達到英國高血壓協會所認定的 grade C 水準。
According to the information from Taiwan Hypertension Society, there are 6.24 million
populations have potential hypertension risk. Only one out of every 5 people measures blood
pressure every week. If we want to avoid hypertension, it is critical to have a habit of measuring
blood pressure in every day. In addition, if measuring blood pressure under relaxes condition, we
could keep white coat hypertension away. In this paper, we provide cuffless/contactless image
sphygmomanometer. By off-the-shelf web camera, we can monitor blood pressure every time,
everywhere. With image base solution, we can overcome cannot continue measuring problem by
traditional blood pressure meter. We capture pulse transit time (PTT) from camera and base the
model from physiological database with machine learning method. After experimental, the SBP
RMSE is 4.56 mmHg; the DBP RMSE is 3.97 mmHg. And the accuracy can reach grade C under
the standard British Hypertension Society (BHS) protocol.
populations have potential hypertension risk. Only one out of every 5 people measures blood
pressure every week. If we want to avoid hypertension, it is critical to have a habit of measuring
blood pressure in every day. In addition, if measuring blood pressure under relaxes condition, we
could keep white coat hypertension away. In this paper, we provide cuffless/contactless image
sphygmomanometer. By off-the-shelf web camera, we can monitor blood pressure every time,
everywhere. With image base solution, we can overcome cannot continue measuring problem by
traditional blood pressure meter. We capture pulse transit time (PTT) from camera and base the
model from physiological database with machine learning method. After experimental, the SBP
RMSE is 4.56 mmHg; the DBP RMSE is 3.97 mmHg. And the accuracy can reach grade C under
the standard British Hypertension Society (BHS) protocol.