應用於下視網膜植入具電荷泵升壓電路之四向共用電極180奈米互補式金氧半480像素感測及雙向電流刺激晶片設計
顯示簡易詮釋資料 統計資料| 欄位 | 內容 | 語言 |
|---|---|---|
dc.contributor.advisor |
吳重雨 | |
dc.contributor.advisor |
Wu, Chung-Yu | |
dc.contributor.author |
廖容興 | |
dc.contributor.author |
Liao, Jung-Hsing | |
dc.contributor.college |
電機學院 | |
dc.contributor.committee |
謝志成 | |
dc.contributor.committee |
邱進峯 | |
dc.contributor.committee |
Hsieh, Chih-Cheng | |
dc.contributor.committee |
Chiu, Chin-Fong | |
dc.contributor.department |
電子研究所 | |
dc.contributor.studentid |
0450205 | |
dc.date.NCLavailable |
2023/05/17 | |
dc.date.USTavailable |
2023/05/17 | |
dc.date.issued |
2018 | |
dc.date.issuedROC |
107 | |
dc.date.schoolyear |
106 | |
dc.description.abstract |
本篇論文提出及分析應用於下視網膜互補式金氧半太陽能電池480像素刺激晶片移植系統,晶片架構基於分區供電方式(DPSS)和主動式像素元件(APS)改良輸出刺激電流之效率。實驗之晶片三十二區分區供電為總數480的光二極體陣列以及控制訊號產生電路。而晶片採用台積電0.18um互補式金氧半CIS製程工藝。最終的晶片尺寸為3.1mm x 3.1mm,起初晶片無法正常運作,但經過電路修改程序後(FIB),此晶片在 505.4 lux 之訊號光以及94mW/cm2 之紅外線背景光照射下量測,負載為10-kΩ時控制訊號頻率為38赫茲,輸出的雙向刺激電流有9.0uA;負載為生物模型時控制訊號頻率為30赫茲,輸出的雙向刺激電流有9.0uA,累積電荷量為9.8nC。此量測結果驗證了晶片修改過後可以正常操作。 | |
dc.description.abstract |
A photovoltaic-cell-powered CMOS 480-pixel implantable chip is proposed for subretinal prostheses. In the proposed chip, the divisional power supply scheme (DPSS) and the active pixel sensor (APS) are adopted to improve the efficiency of output stimulation currents and the image sensitivity. The proposed chip consists of totally 480 photodiode array with 32 DPSS divisions, control signal generator circuits, and photovoltaic cells. It is designed and fabricated in 180-nm CMOS image sensor (CIS) technology. The chip size is 3.1mm x 3.1mm. At first, the chip have not any output function. After FIB, this chip measured frequency of 32-phase control signals is 30 Hz under signal light intensity of 505.4 lux and background IR intensity of 94 mW/cm2. The measured output stimulation current is 9.0 μA under 10-kΩ load. Under the equivalent electrode impedance load, the measured frequency of 32-phase control signals is 38 Hz. The measured peak output stimulation current is 9.0 μA and the amount of injected charges per pixel is 9.8 nC. The measurement results have verified the correct function of the proposed subretinal implant chip after FIB. | |
dc.description.abstractopendate |
||
dc.description.degree |
碩士 | |
dc.description.tableofcontents |
ABSTRACT (CHINESE) iABSTRACT (ENGLISH) iiACKNOWLEDGEMENTS ivCONTENTS vTABLE CAPTIONS viiFIGURE CAPTIONS viiiCHAPTER 1 INTRODUCTION 11.1 Background 11.2 Review on the Subretinal Prostheses 41.3 Motivation 101.4 Design Goals 111.5 Main Results and Thesis Organization 12CHAPTER 2 CHIP ARCHITECTURE AND CIRCUIT DESIGN 132.1 System And Chip Architecture 132.2 Photovoltaic Cell Layout Design 182.3 Circuit Design .212.3.1 Four Directional Sharing Electrodes (FDSEs) 212.3.2 Adaptive Background Cancellation Circuit (ABCC) 262.3.3 Pixel Circuit 312.3.4 Constant GM, Clock Generator And Level Shifter 332.3.5 Control Signal Generation Curcuit 362.3.6 Charge Pump Circuit [27] 392.3 Post Simulation Results 43CHAPTER 3 EXPERIMENTAL RESULTS 453.1 Measurement Results of Photovoltaic Cell Test-keys 453.2 Measurement Results of Subretinal Chips 523.2.1 Measurement Setup 523.2.2 Measurement Results 553.2.3 Discussion 653.2.4 Redesign 68CHAPTER 4 CONCLUSIONS AND FUTURE WORK 704.1 Conclusions 704.2 Future Work 71REFERENCES 72VITA 78 | |
dc.identifier.citation |
[1] C.Hamel,”Retinitis pigmentosa,” Orphanet Journal of Rare Diseases, vol. 1, no.1 , article no. 40 ,2006.[2] D. S. Friedman, B. J. O'Colmain, B. Munoz, S. C. Tomany, C. McCarty et al., “Prevalence of age-related macular degeneration in the United States,” Arch Ophthalmol, vol. 122, no. 4, pp. 564-572, April 2004.[3] G. Brindley, W. Lewin "The sensation produced by electrical stimulation of the visual cortex". Journal of Physiology 196, pp. 479–93, 1968.[4] A. Majji, M. Humayun, J. Weiland, S. Suzuki, S. D’Anna, E. deJuan Jr. (1999). "Long-term histological and electrophysiological results of an inactive epiretinal electrode array implantation in dogs". Investigative Ophthalmology and Visual Science 40 (9): 2073–81.[5] P. Walter, P. Szurman, M. Vobig, H. Berk, H. Ludtke-Handjery, H.Richter, et. al. (1999). "Successful long-term implantation of electrically inactive epiretinal microelectrode arrays in rabbits". Retina 19 (6): 546–52.[6] K. Sooksood, E. Noorsal, J. Becker, and M. Ortmanns, " A neural stimulator front-end with arbitrary pulse shape, HV compliance and adaptive supply requiring 0.05mm2 in 0.35μm HVCMOS," IEEE International Solid-State Circuits Conference, pp. 306-308,2011.[7] C.-C. Chiao, Y.-T. Yang, C. Wan, W.-C. Yang, L.-J. Lin, P.-K. Lin, and C.-Y. Wu,” Responses of rabbit retinal ganglion cells to subretinal electrical stimulation with a silicon-based microphotodiode array,” ARVO Abstract. Invest. Ophtha. Vis. Sci., vol. 33, no. 4, May 2010, pp. 1048-1992.[8] K. Mathieson, J. Loudin, G. Goetz, P. Huie, L. Wang, T. I. Kamins, L. Galambos, R. Smith, J. S. Harris, A. Sher and D. Palanker, “Photovoltaic retinal prosthesis with high pixel density,” Nature Photonics, vol. 6, pp. 391-397, May. 2012.[9] David Boinagrov, Xin Lei, Georges Goetz, Theodore I. Kamins, Keith Mathieson, Ludwig Galambos, James S. Harris, Jr., and Daniel Palanker, “Photovoltaic Pixels for Neural Stimulation: Circuit Models and Performance,” IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, VOL. 10, NO. 1, 2016.[10] C.-L. Lee and C.-C. Hsieh, “A 0.8-V 4096-Pixel CMOS Sense-and-Stimulus Imager for Retinal Prosthesis,” IEEE Transactions on Electron Devices, vol. 60, pp. 1162-1168, 2013.[11]P.-K. Lin, P.-H. Kuo, Y.-C. Tsai, M.-J. Sui, C.-C. Chiao, T. Noda, J. Ohta, and C.-Y. Wu, “The ex vivo and in vivo electrophysiological investigations of a subretinal photovoltaic prosthesis embedded with solar cells and divisional-power-supply-scheme,” in TEATC (World Research Congress: The Eye and The Chip), 2014.[12] S. Oh, J.-H. Ahn, S. Lee, H. Ko, J. M. Seo, Y.-S. Goo, and D. Cho, “Light-Controlled Biphasic Current Stimulator IC Using CMOS Image Sensors for High-Resolution Retinal Prosthesis and In Vitro Experimental Results With rd1 Mouse,” IEEE Transactions on Biomedical Engineering, vol. 62, pp. 70-79, 2015.[13] K. Chen, Z. Yang, L. Hoang, J. Weiland, M. Humayun, and W. Liu, “An integrated 256-channel epiretinal prosthesis,” IEEE J. Solid-State Circuits, vol. 45, pp. 1946-1956, Sept. 2010.[14] K. Chen, Z. Yang, L. Hoang, J. Weiland, M. Humayun, and W. Liu, “A 37.6mm2 1024-Channel High-Compliance -Voltage SoC for Epiretinal Prostheses,” in IEEE ISSCC Dig. Tech. Papers, 2013, pp. 294-296.[15] C.-Y. Wu, W.-J. Sung, P.-H. Kuo, C.-K. Tzeng, C.-C. Chiao, and Y.-C. Tsai, “The design of CMOS self-powered 256-pixel implantable chip with on-chip photovoltaic cells and active pixel sensors for subretinal prostheses,” IEEE Biomedical Circuits and Systems Conference (BioCAS), 2015.[16] A. Rothermel., “A 1600-pixel subretinal chip with DC-free terminals and ±2V supply optimized for long lifetime and high stimulationefficiency,” ISSCC Dig. Tech. Papers, pp. 144-602, Feb. 2008.[17] Liu Liu, Jiirgen Wiinschmann, Naser Pour Aryan, Amr Zohny, Michael Fischer,Steffen Kibbelt , Albrecht Rothermel, “An ambient light adaptive subretinal stimulator,” Conference of the IEEE ESSCIRC, 10 November 2009.[18] Brandon Bosse, Eberhart Zrenner, Robert Wilk, “Standard ERG Equipment Can Be Used to Monitor Functionality of Retinal Implants,” 33rd Annual International Conference of the IEEE EMBS Boston, Massachusetts USA, August 30 - September 3, 2011.[19] Christian Brendler, Naser Pour Aryan, Viola Rieger, and Albrecht Rothermel, “Wireless Power Delivery for a Biomedical Retinal Prosthesis,” IEEE 20th International Conference on Electronics, Circuits, and Systems (ICECS), December 2013.[20] Stingl K, Bartz-Schmidt KU, Besch D, Braun A, Bruckmann A, et al , “ Artificial vision with wirelessly powered subretinal electronic implant alpha-IMS,” Proceedings of the Royal Society B: Biological Sciences, 280:20130077, 20 February 2013.[21] Albrecht Rothermel, “Recent Results With Subretinal Stimulation,” IEEE Biomedical Circuits and Systems Conference (BioCAS), 2014.[22] C.-Y. Wu, P.-H. Kuo, P.-K. Lin, P.-C. Chen, W.-J. Sung, Jun Ohta, Takashi Tokuda, and Toshihiko Noda, “A CMOS 256-Pixel Photovoltaics-Powered Implantable Chip with Active Pixel Sensors and Iridium-Oxide Electrodes for Subretinal Prostheses,” Sensors and Materials, Vol. 30, No. 2, 2018.[23] J. Ohta, T. Tokuda, K Kagawa, “Laboratory investigation of microelectronics-based stimulators for large-scale suprachoroidal transretinal stimulation (STS),” J Neural Eng. 2007;4:S85–S91.[24] T. Tokuda, T. Ito, T. Kitao, T. Noda. “CMOS-based smart-electrode-type retinal stimulator with bullet-shaped bulk Pt electrodes,” 33rd Annual International Conference of the IEEE EMBS Boston, Massachusetts USA, August 30 - September 3, 2011.[25] P. B. Matteucci, P. Byrnes-Preston, S. C. Chen, N. H. Lovell, and G. J. Suaning, "ARM-based visual processing system for prosthetic vision," IEEE Engineering in Medicine and Biology Society, pp. 3921-3924, 2011.[26] C.-Y. Wu, P.-K. Lin, J. Lin, C. Yang, C. Wan, “Power controlling apparatus applied to biochip and operating method thereof.” U.S. Patent 7622 702, Mar. 12, 2009.[27] O.-Y. Wong, P.-H. Chen, C.-Y. Wu, “A Fully-Integrated Charge Pump for Self-Powered Implantable Retinal Prostheses,” IEEE International Symposium on Radio-Frequency Integration Technology (RFIT), 2016.[28] G. Yang, X. Yu, S. Lun, “A new analytical method for the extraction of series resistance of solar cell,” International Conference on Mechatronics and Control (ICMC), July 3 - 5, 2014.[29] Z. Chen, M.-K. Law, P.-I. Mak, and Rui P. Martins, “A Single-Chip Solar Energy Harvesting IC Using Integrated Photodiodes for Biomedical Implant Applications,” TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, VOL. 11, NO. 1, 2017.[30] S. Negi, R. Bhandari, L. Rieth, and F. Solzbacher, “In vitro comparison of sputtered iridium oxide and platinum-coated neural implantable microelectrode arrays,” Biomed.Mater., vol. 5, 2010.[31] Sungjin Oh, Jae-Hyun Ahn, Sangmin Lee, Hyoungho Ko, Jong Mo Seo, Yong-Sook Goo, and Dong-il “Dan” Cho, “Light-Controlled Biphasic Current Stimulator IC Using CMOS Image Sensors for High-Resolution Retinal Prosthesis and In Vitro Experimental Results With rd1 Mouse,” TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL. 62, NO. 1, JANUARY 2015.[32] Viola Rieger , Fabian Buntz , Christian Feller and Albrecht Rothermel, “Low-Latency System for Evaluation of Image-Enhancement-Algorithms on Patients using Subretinal Implants ,” Engineering in Medicine and Biology Society (EMBC), 05 November 2015.[33] F. Gekeler, K. Kobuch, H. N. Schwahn, A. Stett, K. Shinoda, and E. Zrenner: Graef. Arch. Clin. Exp. Ophthalmol. 242 (2004) 587. | |
dc.identifier.uri |
https://hdl.handle.net/20.500.14371/68280 | |
dc.language.iso |
中文 | |
dc.relation.alma |
990056576280206772 | |
dc.relation.page |
78 | |
dc.subject.keyword |
下視網膜 | |
dc.subject.keyword |
雙向電流刺激晶片 | |
dc.subject.keyword |
四向共用電極 | |
dc.subject.keyword |
電荷泵升壓電路 | |
dc.subject.keyword |
太陽能電池 | |
dc.subject.keyword |
SUBRETINAL PROSTHESIS | |
dc.subject.keyword |
BIPHASIC CURRENT STIMULATION CHIPS | |
dc.subject.keyword |
FOUR DIRECTIONAL SHARING ELECTRODES | |
dc.subject.keyword |
CHARGE PUMP | |
dc.subject.keyword |
PHOTOVOLTAIC CELL | |
dc.title |
應用於下視網膜植入具電荷泵升壓電路之四向共用電極180奈米互補式金氧半480像素感測及雙向電流刺激晶片設計 | zh-TW |
dc.title |
THE DESIGN OF 180-NM CMOS 480-PIXEL SENSING AND BIPHASIC CURRENT STIMULATION CHIPS WITH FOUR DIRECTIONAL SHARING ELECTRODES AND CHARGE PUMP FOR SUBRETINAL PROSTHESIS | en |
dc.type |
Thesis | zh-TW |
檔案
- 01.pdf (4.56 MB, Adobe PDF)
所屬典藏: 電子研究所