攀藤PM2.5感測器之實場評估校正
Field evaluation and calibration of low-cost Plantower PM2.5 sensors
摘要
手動採樣的聯邦參考方法(Federal Reference Method, FRM) 及自動監測的聯邦等效方法(Federal Equivalent Method, FEM)所量測的PM2.5質量濃度測值較準確,但因儀器的價格較昂貴且體積龐大而導致測站數量不足。目前已有許多研究針對可大量佈置、體積較小且價格低廉的PM2.5微型感測器進行與FRM及FEM的比對,進行其準確度的評估。期望未來可以應用PM2.5感測器建立廣泛、即時的空氣品質監測網。
本團隊在國立交通大學環境工程研究所頂樓不同期間進行三種不同型號攀藤PM2.5感測器(PMS5003、PMS7003及PMS A003)與TEOM-FDMS(Tapered Element Oscillating Microbalance with Filter Dynamics Measurement System, Thermo, 1405-DF)和BAM-1020(Beta Attenuation Monitor, Met One, 1020)進行大氣PM2.5小時平均質量濃度比對測試。結果顯示,PMS5003、PMS7003及PMS A003與TEOM-FDMS未揮發修正之測值平均標準化誤差(Mean Normailized Error,MNE)分別為39.14±32.89%、88.52±56.92%及63.89±56.14%;PMS5003、PMS7003及PMS A003與TEOM-FDMS揮發修正之測值MNE分別為24.06±26.19%、59.37±48.35%及50.07±57.98%;PMS A003與BAM-1020之測值MNE為40.39±35.63%。進一步探討後發現,攀藤PM2.5感測器與TEOM-FDMS及BAM-1020的量測誤差受環境相對溼度影響。並透過結合環境因子(環境溫度、相對濕度)校正攀藤PM2.5感測器,使其與TEOM-FDMS未揮發修正測值之MNE降至17.26%~32.01%;與TEOM-FDMS揮發修正測值之MNE降至17.68~32.06%。
綜合上述,雖然經校正後的攀藤PM2.5感測器與TEOM-FDMS及BAM-1020仍有是17~32%的誤差,但已有足夠的準確性來作為補充用的空氣品質監測網。
本團隊在國立交通大學環境工程研究所頂樓不同期間進行三種不同型號攀藤PM2.5感測器(PMS5003、PMS7003及PMS A003)與TEOM-FDMS(Tapered Element Oscillating Microbalance with Filter Dynamics Measurement System, Thermo, 1405-DF)和BAM-1020(Beta Attenuation Monitor, Met One, 1020)進行大氣PM2.5小時平均質量濃度比對測試。結果顯示,PMS5003、PMS7003及PMS A003與TEOM-FDMS未揮發修正之測值平均標準化誤差(Mean Normailized Error,MNE)分別為39.14±32.89%、88.52±56.92%及63.89±56.14%;PMS5003、PMS7003及PMS A003與TEOM-FDMS揮發修正之測值MNE分別為24.06±26.19%、59.37±48.35%及50.07±57.98%;PMS A003與BAM-1020之測值MNE為40.39±35.63%。進一步探討後發現,攀藤PM2.5感測器與TEOM-FDMS及BAM-1020的量測誤差受環境相對溼度影響。並透過結合環境因子(環境溫度、相對濕度)校正攀藤PM2.5感測器,使其與TEOM-FDMS未揮發修正測值之MNE降至17.26%~32.01%;與TEOM-FDMS揮發修正測值之MNE降至17.68~32.06%。
綜合上述,雖然經校正後的攀藤PM2.5感測器與TEOM-FDMS及BAM-1020仍有是17~32%的誤差,但已有足夠的準確性來作為補充用的空氣品質監測網。
The measurements of Federal Reference Method(FRM) and Federal Equivalent Method(FEM) are more accurate but the expensive and bulky instruments make them hard to widely use. In order to built the extensive air quality monitoring network, the low-cost and compact sensors have been evaluated in the field and laboratory test.
In this study, three models of Plantower sensors were compared with TEOM-FDMS and BAM-1020 at NCTU in three differenct sampling periods (2017/03/31-2017/04/20, 2018/01/01-2018/05/05, 2018/08/10-2019/01/14). The results show that the Mean Normailized Error(MNE) of the PMS5003, PMS7003 and PMS A003 comparing with TEOM-FDMS base model were 39.14±32.89%, 88.52±56.92% and 63.89±56.14%; the MNE of the PMS5003, PMS7003 and PMS A003 comparing with TEOM-FDMS base-reference model were 24.06±26.19%, 59.37±48.35% and 50.07±57.98%; the MNE PMS A003 comparing with BAM-1020 was 40.39±35.63%. . The measurement of low-cost Plantower PM sensors is significantly affected with the ambient RH compared to reference monitors due to lack of heater and dryer to reduce the RH of incoming aerosols before the PM sensors inlet
The multiple variable regression equation is used to calibrate the PM2.5 concentration measured by Plantower PM2.5 sensors. After Plantower PM2.5 sensors were calibrated, the MNE between the calibrated PM2.5 sensor and TEOM-FDMS base model reduce to 17.26~32.01%; the MNE between the calibrated PM2.5 sensor and TOEM-FDMS base- reference model reduce to 17.68~32.06%.
In this study, three models of Plantower sensors were compared with TEOM-FDMS and BAM-1020 at NCTU in three differenct sampling periods (2017/03/31-2017/04/20, 2018/01/01-2018/05/05, 2018/08/10-2019/01/14). The results show that the Mean Normailized Error(MNE) of the PMS5003, PMS7003 and PMS A003 comparing with TEOM-FDMS base model were 39.14±32.89%, 88.52±56.92% and 63.89±56.14%; the MNE of the PMS5003, PMS7003 and PMS A003 comparing with TEOM-FDMS base-reference model were 24.06±26.19%, 59.37±48.35% and 50.07±57.98%; the MNE PMS A003 comparing with BAM-1020 was 40.39±35.63%. . The measurement of low-cost Plantower PM sensors is significantly affected with the ambient RH compared to reference monitors due to lack of heater and dryer to reduce the RH of incoming aerosols before the PM sensors inlet
The multiple variable regression equation is used to calibrate the PM2.5 concentration measured by Plantower PM2.5 sensors. After Plantower PM2.5 sensors were calibrated, the MNE between the calibrated PM2.5 sensor and TEOM-FDMS base model reduce to 17.26~32.01%; the MNE between the calibrated PM2.5 sensor and TOEM-FDMS base- reference model reduce to 17.68~32.06%.