設計趨流式微流體晶片進行高品質活力精蟲篩選
Design a Rheotaxis-based of Microfluidic Chip for High-quality Actively Motile Sperms Sorting
摘要
現代人不孕的比例,隨著男性精子品質的下降而日趨嚴重。因此如何快速篩檢精蟲品質、篩選優良精子,便成了一個重要的課題。由於男性精蟲品質不佳是造成不孕症的主要因素,因此在人工輔助生殖技術 (Assisted reproductive technology, ART) 中對於精蟲篩選越來越受到重視。現有的精蟲相關診斷、治療之方式,不論是體外人工受精 (In Vitro Fertilization ,IVF) 或是單一精蟲顯微注射 (Intracytoplasmic Sperm Injection,ICSI) ,都需要對精蟲樣本做前處理程序。當中繁瑣的離心步驟及試劑的添加,都會對精蟲造成物理性及化學性的損傷。本研究提出以微流體晶片的方法,進行精蟲泳動能力高低的分選。目前已開發出一套新穎之微流體晶片,其設計是以微流體減速流場之物理特性,進行溫和的精蟲逆游篩選,同時去除掉細胞碎屑、白血球…等雜質。相比過往之相關技術,此微流道晶片,能將嚴重患者檢體中之活耀精蟲比例,從 2 % 大幅提高至 97%。相較現有技術的 22% 及 38%,有著極為顯著的效率提升。同時,由於此無需額外添加可能傷害精蟲之化學品,也無需要進行高強度的離心過程,因此能減少分選過程中對精蟲產生的傷害,大幅降低形成畸胎的機率。此外,在晶片動力的設計上,採用重力做為流體驅動的來源,能簡化流體推動系統,達到操作便捷之目的。分選後之精蟲各項數據皆遠優於市售產品,具有快速、高效率回收、高通量…等特性,可廣泛應用於IVF與ICSI療程,此全能型人工生殖輔助技術平台,可望提高試管嬰兒的成功率,改善少子化問題。
Infertility is a serious globe issue, especially in the developed countries, over 20% infertile male has sperm problems, that include instance immotile, abnormal morphology and low sperm quantity. Therefore, how to isolate high quality sperm for clinical use is extremely required.
Up to date, activity sperm sorting is usually using Swiming-up and Density gradient centrifugation approaches. Unfortunately, both methods are detrimental to the sperm viability and elicits production of reactive oxygen species. Separation sperm by microfluidics has been put emphasis on and studied; however, it faced some challenge such as low throughput and an unacceptable recover rate. Therefore, we developed a new type of microfluidic chip with utilizing the characteristic of sperm anadromous instinct for high-throughput sperm separation. The process can be finished in 10 min, which his 10 times faster than the conventional MFC design. Computer Assisted Semen Analysis (CASA) System was utilized to calculate sperm quality and visualize the sperm by spermac stain. Experiment reveals that the qualified sperm motility was raised up from 30% to 99% and linearity (LIN) up from 0.2 to 0.85 via sorting by our microfluidic chip (MFC chip).
In this novel device, high quality sperm were sorted by their anadromous phenomenon and results in an ultra-rapid and effective sorting ability. Compared with traditional Swinging up, Density gradient or other microfluidic separation methods. Using the CASA system to analyze the sperm motion trail, we found sperm sorted by microfluidic have high linear velocity and linearity. The anadromous sperm grading chip provide high-throughput separation with live high quality sperm.
Up to date, activity sperm sorting is usually using Swiming-up and Density gradient centrifugation approaches. Unfortunately, both methods are detrimental to the sperm viability and elicits production of reactive oxygen species. Separation sperm by microfluidics has been put emphasis on and studied; however, it faced some challenge such as low throughput and an unacceptable recover rate. Therefore, we developed a new type of microfluidic chip with utilizing the characteristic of sperm anadromous instinct for high-throughput sperm separation. The process can be finished in 10 min, which his 10 times faster than the conventional MFC design. Computer Assisted Semen Analysis (CASA) System was utilized to calculate sperm quality and visualize the sperm by spermac stain. Experiment reveals that the qualified sperm motility was raised up from 30% to 99% and linearity (LIN) up from 0.2 to 0.85 via sorting by our microfluidic chip (MFC chip).
In this novel device, high quality sperm were sorted by their anadromous phenomenon and results in an ultra-rapid and effective sorting ability. Compared with traditional Swinging up, Density gradient or other microfluidic separation methods. Using the CASA system to analyze the sperm motion trail, we found sperm sorted by microfluidic have high linear velocity and linearity. The anadromous sperm grading chip provide high-throughput separation with live high quality sperm.