非侵入式散射係數量測及其在非侵入式量測血糖上之應用
Non-invasive scattering coefficient measurement and its application on non-invasively monitoring blood glucose changes
摘要
In this thesis, we try to compose a low-cost, portable non-invasive blood glucose device with simple light source and minor structure. We take the variety of mean wavelength number per unit distance as the coefficient which help us distinguish the concentration of our sample. First, we simulate the tiny concentration difference of the blood glucose by gradually adding small amount of glucose liquid into the intralipid liquid. We have the fixed intralipid liquid as the human tissue, then simulate the blood glucose concentration change from 70 mg/dL to 140 mg/dL. This outcome reveals that the standard deviation of all concentration are below the standard of a commercial glucose machine when the integration time is 3 minutes. The intralipid is homogeneous and borderless which is easier for the experiment and analysis, but the defect is that it cannot perfectly present the structure of human skin.
After we confirm that we can tell the difference of slightly sugar concentration change from the former experiment, we wonder if it works on human and we try to make the coefficient numerical if it is feasible. We choose the back of the hand as the experimental sample. Starting from time before breakfast, eating breakfast, until an hour after it started.
After we confirm that we can tell the difference of slightly sugar concentration change from the former experiment, we wonder if it works on human and we try to make the coefficient numerical if it is feasible. We choose the back of the hand as the experimental sample. Starting from time before breakfast, eating breakfast, until an hour after it started.
本篇論文嘗試利用簡易光源與微小的架構組成低成本又可攜式的非侵入式血糖監測裝置。我們使用每單位距離產生的光譜之重心波長變化來當作我們的指標,協助我們比較不同的散射濃度之樣品。首先我們在散射劑溶液中,依序加入微量的葡萄糖溶液來模擬人體血糖的微量變化。我們利用固定濃度的散射劑來代表人體組織,接著模擬了從人體空腹的最低值70 mg/dL,不斷提升濃度至接近飯後最高值的140 mg/dL。散射劑實驗結果顯示當積分時間為3分鐘時,所有濃度的標準差會低於一定的標準,換算成解析度則符合市售血糖機的標準,不過由於散射劑溶液具有均質性且無邊界性,雖然量測與分析都較為容易,但缺點是是無法完美呈現人體組織的分層。
確認假體實驗可以量測到微小糖濃度變化後,人體量測的部分則是希望可以對量測到的參數做進一步的量化分析,推估出我們的量測數值與血糖機量測到的數值做比較。在人體量測部分我們選擇在血管最明顯的部位,也就是手背的量測,從早餐前空腹狀態開始量測,十分鐘後開始進食,量測時間持續一小時。結果顯示人體的數據明顯震盪較大,趨勢不如散射劑實驗的明顯,要從中進行分析的難度提升許多。
確認假體實驗可以量測到微小糖濃度變化後,人體量測的部分則是希望可以對量測到的參數做進一步的量化分析,推估出我們的量測數值與血糖機量測到的數值做比較。在人體量測部分我們選擇在血管最明顯的部位,也就是手背的量測,從早餐前空腹狀態開始量測,十分鐘後開始進食,量測時間持續一小時。結果顯示人體的數據明顯震盪較大,趨勢不如散射劑實驗的明顯,要從中進行分析的難度提升許多。