分區參數化支架設計之下顎大範圍骨缺損特製化植入物開發

Development of Large Patient-specific Mandible Prosthesis with Regional Parametric Structure

檔案

全文未公開。

日期

2020

類型

語言

中文

摘要

臨床上下顎大範圍骨缺損治療目標為恢復病患外觀型態及正常咬合之生理功能。然而目前腓骨皮瓣重建手術之治療,無法完整填補缺損部位和提供足夠高度以供植牙,骨板亦有力學強度不足之疑慮。新興之金屬積層製造特製化補骨塊雖可完好恢復外觀,但未經結構設計重量過重,可能導致植入物崩塌或骨頭負荷過大;部分人工牙根一體成型之補骨塊則存在細菌感染風險,未考量下顎各區段之不同咬合力學條件。本研究之目的為考量下顎骨各區段不同力學條件,分區針對補骨塊結構進行有限元素分析、拓樸最佳化與拓樸密度加權計算,將各區最佳結構歸納為參數化支架,設計一可恢復美觀、可允許植牙、輕量化兼顧力學強度之下顎骨分區補骨塊。
臨床資料收集階段,依據國際內固定協會(AO foundation),先針對下顎骨分區為C區頦端(Symphyseal and parasymphyseal)、B區下顎骨體(Body)與A區下顎角(Angle and ramus),收集105位下顎骨醫學影像進行尺寸量測與統計,進而篩選出最危險案例作為分析設計對象,並建立下顎尺寸資料庫及定義特製化尺寸允收範圍。針對最危險案例設計預留移植骨空間之C/B/A+B三區特製化補骨塊,各區分別進行正向力及軸向力拓樸最佳化,三區以不同力學加權比例係數計算二種最佳化結果並統整為參數支架結構,B+C+B/A+B+C區補骨塊由已參數化之C/B/A+B區補骨塊組合獲得,再次將五區參數化支架補骨塊進行有限元素力學驗證分析,分別與各區實心/空心補骨塊進行重量與應力值(von-Mises stress)之比較,以驗證參數化支架之輕量化及力學強度。將C/B/A+B區參數化支架之特製化補骨塊進行金屬積層製造,分別針對各區補骨塊長度(L)、高度(H)、厚度(T)及孔徑(D)進行製造尺寸精準度測試,製造誤差目標為小於10%;生物力學測試分別以補骨塊及市售骨板重建下顎骨C/B/A+B區缺損模型,模擬下顎術後六個月咬合情況(20~200N,250000次),與市售重建骨板進行力學強度比較。
經拓樸加權最佳化分析,C/B/A+B區加權後最佳結構分別可參數化為2(分別位於第二小臼齒與第一大臼齒下方)、2(分別位於側門齒下方)及4支(其中2支支架位於第二小臼齒與第二大臼齒下方,另2支支架位於下顎枝中心,且相互約呈15°)支架,其尺寸及位置分別以補骨塊寬度W及牙齒位置定義,並經由力學驗證模擬分析,顯示參數化支架模型在重量及應力上比實心模型分別下降50~70%及5~30%,且在跨區域時此下降趨勢愈加明顯,證實參數化支架達成輕量化兼具更佳力學強度之目標。尺寸精準度測試結果,各區補骨塊長度、高度、厚度及孔徑尺寸之製造誤差皆小於10%,符合製造目標;在生物力學測試結果,各區補骨塊重建之下顎模型皆完全通過250000次負載而未產生明顯破壞,相較之下重建骨板組產生骨釘崩壞、骨板斷裂及下顎髁頭斷裂等嚴重破壞,荷重元(load cell)及植入物本體之位移值也皆高於補骨塊組,證實參數化支架設計補骨塊力學強度優於市售重建骨板。
本研究已建立可恢復外觀及提供植牙之分區支架參數化補骨塊,建立下顎資料庫並定義下顎特製化尺寸允收範圍,篩選最危險案例之拓樸加權最佳化分析,由力學驗證分析證實其力學強度優於實心/空心補骨塊,由生物力學測試驗證其力學強度優於市售重建骨板。
The main goal of mandible reconstruction for mandible large defect caused by tumor or trauma is to restore facial appearance and occlusal function. However, the free fibula flap reconstruction cannot achieve the goal due to the difference in appearance between fibula flap and mandible, insufficient height for dental implant and insufficient strength for the traditional reconstruction plate. Although the application of 3D printing in medical implants has flourished in recent years, the consideration of occlusal function, structure optimization, overweight and bacterial infections are needed to be addressed. The aim of this study is to develop a large patient-specific mandible prosthesis with regional parametric structure under the consideration of the different mechanical conditions of each segment of the mandible by design, analysis, manufacture and verification of the worst case.
In this study, the CT images of the patients’ mandible were measured in region C (Symphyseal and parasymphyseal)、region B (Body) and region A (Angle and ramus) based on AO foundation classification, and were statistic analyzed to establish a clinical database and define the patient-specific range. The worst case was selected to design the region C/B/A+B patient-specific mandible prosthesis on which reserve space for bone grafting for dental implants. By the finite element analysis and topology optimization for positive force and axial force, the two topological optimization results in each region were calculated with different mechanical weighted coefficients to obtain the best structure which could be unified into a parametric structure. B+C+B/A+B+C patient-specific prosthesis was combined by the parameterized C/B/A+B patient-specific prosthesis. The C/B/A+B/B+C+B/A+B+C large patient-specific mandible prosthesis with regional parametric structure was analyzed again to be compared with solid/hollow prosthesis of each region in weight and stress (von-Mises stress). The C/B/A+B patient-specific prosthesis was manufactured by metal 3D printing, and the length (L), height (H), thickness (T) and aperture (D) of the prostheses were measured respectively. In biomechanical testing, ABS defect mandible models were reconstructed by the C/B/A+B patient-specific prosthesis and commercially reconstruction plate to simulate the mechanical conditions of the mandible for six months (20~200 N, 250,000 times) and to compare the mechanical strength between the patient-specific prosthesis and commercially reconstruction plate.
After topological optimization and weighted calculation, the weighted optimal structure of the C/B/A+B prosthesis could be parameterized into 2, 2, and 4 stents, respectively. The dimension and the position of the stent structure were defined by the width of the prosthesis (W) and the tooth. In the mechanical verification analysis, the parameterized prosthesis had a 50-70% and 5-30% decrease in weight and stress in comparison with the solid model. In the dimensional accuracy test, the manufacturing errors of the length, height, thickness, and aperture size of each patient-specific prosthesis were less than 10%. In the biomechanical test, all mandible models reconstructed by the patient-specific prosthesis passed without significant damage. In contrast, the model reconstructed by commercial plate occurred bone nail collapse, bone plate fracture, and fracture of condyle of the mandible model. The displacement of the load cell and the reconstruction plate was higher than those of patient-specific prosthesis, confirming that the patient-specific prosthesis with regional parametric structure had better strength and stability than the reconstruction plate.
In this study, the clinical mandible database has been collected to define the patient-specific range and select the worst case for design, analysis and verification. The large patient-specific mandible prosthesis with regional parametric structure has been established by the finite element analysis, topology optimization and weighted calculation, and the mechanical strength has been verified by the mechanical verification analysis and biomechanical testing.
顯示完整詮釋資料 統計資料

引用此條目

蔡瑋姮 (2020). 分區參數化支架設計之下顎大範圍骨缺損特製化植入物開發. https://thesis.lib.nycu.edu.tw/handle/20.500.14371/75404

BibTeX RIS

引用文字依 APA 第 7 版格式自動產生,使用前請再核對。

分享

永久網址:https://thesis.lib.nycu.edu.tw/handle/20.500.14371/75404