可見光與近紅外光雙波段光纖化頻域式光學同調斷層掃描攝影術之研發
Development of Visible and Near-infrared Light Dual-band All-fiber Based Spectral-domain Optical Coherence Tomography
摘要
可見光光譜式光學同調斷層掃描攝影術(Visible-light spectroscopic optical coherence tomography, vis-OCT)是OCT一項重要的延伸應用,基於組織內吸光物質在可見光範圍內的衰減係數µt(λ),套用短時傅立葉轉換(Short-time Fourier transform, STFT)以解析頻譜中不同波段的深度影像資訊,較短的中心波長(λ0 = 575 nm)也提供更佳的軸向解析度,主要用於分析活體血液的血氧飽和度(Blood oxygen saturation, SO2),而SO2已被許多研究指出,能夠在視網膜病變以及神經退化性疾病等應用上提供臨床診斷的指標,幫助醫師判斷病程。
然而,可見光波段卻因為穿透深度不夠限制了其應用,因此我們開發了一套新穎的單光源雙波段光學同調斷層影像掃描術,加入NIR(λ0 = 1275 nm)系統後,不僅使結構影像更加完整,在應用層面也將更為廣泛。架構上相較於free-space的做法,我們採用光纖化的設計以縮小架構體積,藉此提升系統穩定度與操作便利性;在樣品端,則利用二色分光鏡(Dichroic mirror)將兩個波段整合於同一道光路,並使用單一掃描振鏡組同步擷取影像,節省掃描時間,是一個相當具有發展潛力的影像工具。
然而,可見光波段卻因為穿透深度不夠限制了其應用,因此我們開發了一套新穎的單光源雙波段光學同調斷層影像掃描術,加入NIR(λ0 = 1275 nm)系統後,不僅使結構影像更加完整,在應用層面也將更為廣泛。架構上相較於free-space的做法,我們採用光纖化的設計以縮小架構體積,藉此提升系統穩定度與操作便利性;在樣品端,則利用二色分光鏡(Dichroic mirror)將兩個波段整合於同一道光路,並使用單一掃描振鏡組同步擷取影像,節省掃描時間,是一個相當具有發展潛力的影像工具。
Visible-light spectroscopic optical coherence tomography (vis-OCT) is one of the extensions of conventional OCT. Based on the extinction coefficient of absorbers, we applied a short-time Fourier transform (STFT) to analyze the depth information with different wavelength ranges. Moreover, shorter central wavelength (λ0 = 575 nm) provides a better axial resolution. This method was widely used to extract the blood oxygen saturation (SO2) of living blood. Also, more and more papers pointed out that SO2 can provide some indicators for clinical diagnosis in retinopathy or neurodegenerative disease. This information is helpful for physicians to judge disease progression.
However, vis-OCT was limited by its insufficient penetration depth. To solve this shortcoming, we developed a novel single light source dual-band OCT by adding another near-infrared (NIR) setup (λ0 = 1275 nm) into the vis-OCT. This integrated system not only makes the structural imaging more intact but also has a widespread application. Compare to free-space; we adapted fiber-based practice to reduce the size of the setup; Also, system stability and operational convenience were greatly improved. In the sample arm, we used a dichroic mirror to integrate two light paths of each subsystem into one. To acquire imaging simultaneously and reduce the scanning time, we used a single galvo scanner instead of two for individual use. This proposed platform is a potential imaging tool.
However, vis-OCT was limited by its insufficient penetration depth. To solve this shortcoming, we developed a novel single light source dual-band OCT by adding another near-infrared (NIR) setup (λ0 = 1275 nm) into the vis-OCT. This integrated system not only makes the structural imaging more intact but also has a widespread application. Compare to free-space; we adapted fiber-based practice to reduce the size of the setup; Also, system stability and operational convenience were greatly improved. In the sample arm, we used a dichroic mirror to integrate two light paths of each subsystem into one. To acquire imaging simultaneously and reduce the scanning time, we used a single galvo scanner instead of two for individual use. This proposed platform is a potential imaging tool.
所屬典藏
生醫光電研究所引用此條目
Mi-Sheng Dai (2020). Development of Visible and Near-infrared Light Dual-band All-fiber Based Spectral-domain Optical Coherence Tomography. https://thesis.lib.nycu.edu.tw/handle/20.500.14371/75469
引用文字依 APA 第 7 版格式自動產生,使用前請再核對。
分享
永久網址:https://thesis.lib.nycu.edu.tw/handle/20.500.14371/75469