新型口內壓力治療裝置對阻塞性睡眠呼吸中止症患者之 呼吸道功能及結構之影響研究

Investigation of Airway Function and Structure of Obstructive Sleep Apnea Syndrome Patients under Treatment of A Novel Oral Pressure Therapy Device

檔案

日期

2020

類型

語言

英文

摘要

摘要
睡眠呼吸疾病 (Sleep disordered breathing, SDB) 是在睡眠中肺部氣體交換減少或不足的狀況,其中以阻塞性睡眠呼吸疾病 (Obstructive sleep apnea, OSA) 最為常見。阻塞性睡眠呼吸中止症反復間歇性發生之呼吸事件導因於咽部局部或全面塌陷致使呼吸氣流受到間歇性局限或阻斷。阻塞性睡眠呼吸中止症對身體健康的影響不僅於影響睡眠品質,其經過證實會導致、影響或加劇許多疾病,例如:高血壓與心血管疾病、糖尿病、代謝症候群甚至會增加癌症之致死率。陽壓呼吸器 (Continuous positive airway pressure devices) 是阻塞性睡眠呼吸中止症的標準治療方式,雖足具治療效力,但治療順從性卻十分不佳,其他治療阻塞性睡眠呼吸中止症之替代方式是醫療的一大缺口。時至今日,有一間歇性負壓器 [intermittent negative air pressure device, iNAP Sleep Therapy System (iNAP)] 之新創療法,但其臨床安全性、有效性、及其作用機制尚未被研究與定義。
本論文透過臺灣臨床試驗數據回溯(N=48) 與已經在美國合法上市之比對品 (predicate device) 比對其安全性,揭露此間歇性負壓器之安全疑慮甚小,並發現此器材對患者之口部組織造成之不適感 (oral tissue discomfort) 低於比對品的一半 (47.95% vs. 18.42%)。再者,數據顯示受試者基值睡眠窒息指數 (Apnea-Hypopnea Index, AHI) 具統計力之顯著下降,基值睡眠窒息指數平均為每小時41.4 ± 24.3次,配戴此器材治療下受試者基值睡眠窒息指數改善比率中位數為 −25%。血氧不飽和事件 (oxygen desaturation events) 也呈現具統計力之顯著下降並伴隨著平均血氧飽和度 (Mean SpO2) 和最低血氧飽和度 (lowest SpO2) 之顯著提升。此外,睡眠覺醒指數(arousal index) 也顯著下降,提升患者的睡眠穩定度。因此,間歇性負壓器可降低阻塞性睡眠呼吸中止症患者之呼吸事件,重建病人血氧之穩定性,並降低覺醒次數幫助病患達到較穩定的睡眠狀況。為了解此器材之適用族群,本論文先用受試者基值睡眠窒息指數與身體品質指數 (Body Mass Index, BMI)繪製睡眠窒息指數改變比例與經治療睡眠窒息指數散佈圖 (scatter-chart),並利用Sher 標準圈選出對器材治療較有反應之族群。再利用基值睡眠窒息指數與身體品質指數之亞群分析以清晰定義適應症標準之界線,依三種臨床常用治療阻塞性睡眠呼吸中止症反應標準,最後圈選出此創新性間歇性負壓器之適用族群為:基值睡眠窒息指數介於每小時5~50次之間或身體品質指數小於25之東方種族阻塞性睡眠呼吸中止症患者。
為達到個別睡眠呼吸中止症病患之精準醫療,本論文進一步分析臺灣臨床試驗受試者之鼻咽理學檢查、內視鏡檢查與X光影像側顱結構,發現下列指標可用以預測個別阻塞性睡眠呼吸中止症患者之嚴重程度:馬氏分級指標(Mallampati score)、軟顎後空間分級指標(retropalatal size)、舌後空間分級指標 (retroglossal size)、鼻前脊柱至鼻後脊柱距離、舌骨尖端至下顎平面距離與中段咽旁間隙。上述指標中,唯有鼻前脊柱至鼻後脊柱之距離與阻塞性睡眠呼吸中止症嚴重程度之負相關指標,其餘皆為正相關指標。另一方面,肥厚之口內組織、較壅塞之軟顎後空間、舌後空間與較長的舌頭為此間歇性負壓器治療阻塞性睡眠呼吸中止症療效之反指標。這些結果與磁共振成像(Magnetic Resonance Imaging, MRI)發現間歇性負壓器對軟顎後空間而非舌後空間較有改善力相互呼應。再者,本論文嘗試利用適配回歸模型(fit regression model)、向前選取法(forward selection model)、向後消去法(backward elimination model)、逐步回歸法(stepwise regression model),將各式生理指標、結構指標、睡眠監測報告指標做系統性之整合,以建立預測新型器材療效之模型。最終成功的選用向後消去法所建構的預測模型,模型中使用受試者基值睡眠窒息指數的決定係數(R2)高達 97%,但去掉受試者基值睡眠窒息指數之模型的決定係數也仍持有94%。當將基值睡眠窒息指數此療效預測最重要之因子從預測模型移除時,鼻前脊柱至鼻後脊柱距離(相關係數為-4.724)、舌頭長度(相關係數為30.27)以及舌頭厚度(相關係數為65.7)取代扮演前三大療效預測決定因子,以避免使用耗時費工之睡眠監測之相關參數。
本論文研究發現新型間歇性負壓器能降低臨床呼吸事件、穩定血中含氧量與有效提升睡眠品質,並利用核磁共振成像證實新型間歇性負壓器確實能有效擴增自然睡眠狀態下患者之上呼吸道。最後透過回歸模型的建構,得以預測個別阻塞性睡眠呼吸中止症患者使用新型間歇性負壓器之療效,經模型精準且有效率的預測可提升患者之療效成功率,以增加病患嘗試新型器材治療的意願。尤以其安全性、舒適性與便利性,將對於長期阻塞性睡眠呼吸中止症患者之照護與疾病控制呈現較佳之順應性與穩定性。
ABSTRACT
Sleep disordered breathing (SDB) are conditions that diminish the exchange of air of lungs during sleep. Obstructive sleep apnea (OSA), the most common of these conditions, produces repetitive intermittent respiratory events during which airflow is temporarily restricted or blocked due to partial or complete pharyngeal airway obstruction. OSA does not only impair the sleep quality but have proven to predispose or lead to lots of diseases, such as high blood pressure and heart comorbidity, diabetes, metabolic syndrome and even exacerbate of cancer mortality. Continuous positive airway pressure (CPAP) devices are the standard treatment for OSA. Despite the effectiveness of CPAP, suboptimal adherence to CPAP is universal. Alternative treatment for OSA patients who have difficulties in compliance with CPAP is in highly demand. To this end, a novel intermittent negative air pressure device, iNAP Sleep Therapy System (iNAP), has been designed. Nonetheless, its safety, efficacy and mechanism have not yet been investigated.
By analyzing the iNAP feasibility study clinical data, which were conducted on Asian Taiwanese, and compared with the clinical results of the predicate device, the iNAP has been disclosed with minimum safety concerns and a favorable comparison of the two devices has shown that the oral tissue discomfort rate of iNAP was less than half of the predicate device (47.95% vs. 18.42%). Secondly, the baseline apnea-hypopnea index (AHI) was significently decreased from an average baseline at 41.4 ± 24.3 events/hour with median of percentage change at − 25%. The oxygen desaturation events were also observed to be significantly diminished with both average of oxygen saturation (Mean SpO2 saturation) and the lowest SpO2 being increased significantly. Besides, the arousal index was significantly reduced providing better sleep stability for the treated patients. Thus, the iNAP was demonstrated to not only reduce respiratory events, restore the oxygen saturation stability but significantly reduce the arousals during sleep, providing a more stable sleep for OSA patients. In order to define the indication for this new device, the responders were marked out by Sher criteria with AHI reduction rate and treated AHI on the scatter plots of age, body mass index (BMI) and baseline AHI. To further clarify the borderline of the indication range, three clinically in practice criteria of treatment responsiveness to OSA treatment were applied on subgroups divided by baseline AHI and by BMI. Finally, the indication range was suggested as OSA patients with baseline AHI range between 5~50 events/hour and BMI below 25kg/m2 for oriental ethnicity.
In pursuing precision medicine for individual’s OSA treatment, the nasopharyngeal, nasopharyngoscopy and lateral cephalometric data of the subjects of the Taiwan feasibility study were further investigated. Below parameters were found to be able to predict the severity of OSA: Mallampati score, retropalatal size and retroglossal size, the distance of anterior nasal spine to posterior nasal spine (ANS-PNS), hyoid bone tip to mandibular plane (H-GoMe) and middle pharyngeal space (MPS). Among all, only the ANS-PNS was a negative predictor for the baseline AHI. On the other hand, the redundancies of the oral tissue, reflected by larger scale of redundancy of uvula and the soft palate (RUS), retropalatal size, retropalatal Müller grade and retroglossal Müller grade, as well as the longer tongue-length would predict unfavorable treatment outcome of the iNAP. These findings were consistent with upper airway characterization with magnetic resonance imaging (MRI). The increment of the volume of retropalatal area was more obvious under iNAP treatment (p=0.024). Further to systemically coordinate all physical, structural and polysomnography (PSG) parameters and produce a single regression model for prediction the treatment outcome of the novel devices, the fit regression model, forward selection model, backward elimination model and stepwise regression model were tried. In the end, using backward elimination approach, the regression model used to predict the iNAP-treated AHI with the baseline AHI parameter achieved an R2 of 97%. Further, when deducing the baseline AHI from the prediction model, an R2 of 94% were astonishingly built. When reducing decisive baseline AHI parameter from the prediction model, ANS-PNS (coefficient = -4.724), tongue length (coefficient = 30.27) and tongue thickness (coefficient = 65.7) took the first three places in making it feasible to bypass the time and manual-consuming PSG parameters in prediction of the treatment response.
In summary, the iNAP has clinically proven effective in terms of lowering respiratory events, stabilizing blood oxygen level as well as improving sleep quality. Besides, the iNAP was also demonstrated to be capable of widening the upper airway in MRI analysis even when patients were naturally asleep. With building of regression models in predicting treatment responses to the novel approach, an increase of individual treatment response rate shall in term triggers patients’ willingness to try novel treatment. Adding up the strengthens of the iNAP in its low safety concerns, comfort and convenience, a substantially improvement of the long-tern compliance and OSA control would be likely foreseeable.