梯度導模共振於螢光檢測系統之開發
Development of Fluorescence Detection System Based on Gradient Guided-Mode Resonance Filter
摘要
本研究目的使用具導模共振特性的導模共振濾波器與感光耦合元件開發出一套低成本的螢光檢測系統,主要拿來量測與生物分子標定有關的螢光染劑。週期梯度或是厚度梯度導模共振是一維具有濾波特性的導模共振濾波器,本研究使用週期梯度導膜共振器當作生物螢光檢測系統的分光元件,其檢測波長為500至700 nm,在TE共振下解析度為1 nm。
我們利用週期梯度導模共振濾波器在不同週期位置共振波長不同的特性,將導模共振濾波器緊貼於感光耦合元件(CCD)上。當單一波長入射時,感光耦合元件在產生共振的週期正後方之像素會呈現最低的感光強度,整理每個單一波長入射時,線性感光耦合元件的量測結果,我們便可以得到像素與波長的關係,對一未知螢光入射時,便可從螢光檢測系統得知入射光波峰與強度,並將螢光檢測系統量測的結果與市售光譜儀量測結果進行比較。
本研究各別地量測了單一生物螢光分子6-FAM (6-Carboxyfluorescein) 與 5-TAMRA (5-Carboxytetramethylrhodamine),並將兩者混合後激發量測。各別量測的情況下,6-FAM最小可量測濃度為10-7 M,5-TAMRA最小可量測濃度為10-7 M,而混合後的雙螢光則無法準確地以螢光檢測系統量測到各別濃度。
最後,本研究將不同濃度的人血清白蛋白(Human Serum Albumin)以人工模擬尿液進行稀釋後再加入可與HSA做專一性結合之螢光(Albumin Blue 580),利用生物螢光檢測系統量測不同濃度之螢光強度,其最低檢測極限為40.99 mg/L,藉此模擬受測者腎臟過濾白蛋白的程度,以達到生醫分子檢測之目的。
我們利用週期梯度導模共振濾波器在不同週期位置共振波長不同的特性,將導模共振濾波器緊貼於感光耦合元件(CCD)上。當單一波長入射時,感光耦合元件在產生共振的週期正後方之像素會呈現最低的感光強度,整理每個單一波長入射時,線性感光耦合元件的量測結果,我們便可以得到像素與波長的關係,對一未知螢光入射時,便可從螢光檢測系統得知入射光波峰與強度,並將螢光檢測系統量測的結果與市售光譜儀量測結果進行比較。
本研究各別地量測了單一生物螢光分子6-FAM (6-Carboxyfluorescein) 與 5-TAMRA (5-Carboxytetramethylrhodamine),並將兩者混合後激發量測。各別量測的情況下,6-FAM最小可量測濃度為10-7 M,5-TAMRA最小可量測濃度為10-7 M,而混合後的雙螢光則無法準確地以螢光檢測系統量測到各別濃度。
最後,本研究將不同濃度的人血清白蛋白(Human Serum Albumin)以人工模擬尿液進行稀釋後再加入可與HSA做專一性結合之螢光(Albumin Blue 580),利用生物螢光檢測系統量測不同濃度之螢光強度,其最低檢測極限為40.99 mg/L,藉此模擬受測者腎臟過濾白蛋白的程度,以達到生醫分子檢測之目的。
The purpose of this study is to develop a fluorescence detection system using gradient grating period guided-mode resonance (GGP-GMR) filter and charge-coupled device (CCD). GGP-GMR is a one-dimensional optical filter exhibiting spatially variable resonance wavelengths. The GGP-GMR used in this work has detection range of 500-700 nm with resolution of 1 nm under TE polarization. By attaching a GGP-GMR to a CCD, for a specific wavelength illumination, the light resonates at a specific period and reflects back at this period and transmitted through at other periods such that a minimum intensity is measured underneath this period. By scanning a range of wavelength, we can correlate the minimum intensity corresponding pixel with incident wavelength. For an unknown fluorescence illumination, the peak and intensity can be obtained from the fluorescent detection system.
In this study, detection of 6-FAM (6-Carboxyfluorescein) and 5-TAMRA (5-Carboxytetramethylrhodamine) were first demonstrated separately and concentration of 10-7 M for both fluorophores was achieved. By contrast, the concentration of the mixture cannot be determined accurately. Finally, human serum albumin (HSA) of different concentrations diluted with artificial urine were used to simulate the degree of albumin filtering by the patients’ kidneys. The fluorophores of Albumin 580 was used to specifically bind with HSA where the intensity can be correlated with the HSA concentration. The developed fluorescence detection system can achieve the detection range of 6.25 – 200 mg/L with detection limit of 40.99 mg/L.
In this study, detection of 6-FAM (6-Carboxyfluorescein) and 5-TAMRA (5-Carboxytetramethylrhodamine) were first demonstrated separately and concentration of 10-7 M for both fluorophores was achieved. By contrast, the concentration of the mixture cannot be determined accurately. Finally, human serum albumin (HSA) of different concentrations diluted with artificial urine were used to simulate the degree of albumin filtering by the patients’ kidneys. The fluorophores of Albumin 580 was used to specifically bind with HSA where the intensity can be correlated with the HSA concentration. The developed fluorescence detection system can achieve the detection range of 6.25 – 200 mg/L with detection limit of 40.99 mg/L.