摘要

競爭型ELISA是現今量測抗原的手法,但目前市面上販售的競爭型ELISA Kit,偵測極限卻相當高,無法檢測低濃度的抗原,尤其在非侵入型檢測(唾液、尿液)時,濃度相當低,檢測不易。現今雖有其他方式能夠改善偵測極限,卻也因此產生熱效應、價格昂貴等等問題,能夠簡單操作且便利有效的改善方式,目前仍未出現。
βCTX是檢測骨折所需要的指標中的一種,需要使用競爭型ELISA做為量測手段,且目前唾液中的βCTX因濃度太低無法使用市面上販售之競爭型ELISA KIT進行量測,因此本篇研究在開發出新技術後,將以能夠量測到唾液中此抗原作為目標。
本實驗旨在改善競爭型ELISA偵測極限,預計將偵測極限從98pg/ml降低至20pg/ml,並量測出唾液中βCTX濃度。本研究提出兩種降低偵測極限的方法,第一種為增加反應時抗原濃度,藉由增加反應時抗原體積、減少反應時競爭物體積來達成,且會讓競爭物在反應時維持與標準流程相同總量;第二種為做較多組測試組,將標準流程的三組,增加至五到八組,同時可以藉由選擇中位數及洽大於及小於中位數的三組測試組做出偵測極限較低的檢量線。
透過此兩種方法量測後可以得知,抗原、競爭物皆為標準體積各50 ul及標準濃度,在增加至七組測試組進行量測的情況下,成功將偵測極限低至0.089~12.696 pg/ml,偵測極限在各種體積中及各種組數情況下是降最低的,但有成本較高的缺點。抗原75 ul、競爭物25 ul的體積,在增加至五組測試組進行量測情況下為各體積中表現最佳,成功將偵測極限降低至3.485~48.833 pg/ml,同時兼顧低偵測極限、偵測極限區間穩定及較低成本的優點。
本實驗也成功量測出17受試者中12位唾液中βCTX的濃度:10位健康受試者唾液βCTX濃度,七位分布298.827~2404.51pg/ml,一位超出量測濃度區間無法計算,其餘兩位濃度過低無法量測;7位病患受試者唾液βCTX濃度,五位分布16.861~365.916 pg/ml,兩位濃度過低無法量測。
Competitive ELISA is a method for measuring antigens. However, the popular tool around the world, ELISA KIT, cannot measure low concentration of antigens because of its high limit of detection, especially in non-invasive detection. Although there are some methods that can improve limit of detection, there are also some problems, such as thermal effects and expensive products. The convenient, effective methods that can be easily operated have not created.
βCTX is one of the indicators for detecting fractures, but the concentration of βCTX in saliva is too low, and cannot be measured by current type of competitive ELSA KIT.
The goal of this research is to create an innovative method to improve the limit of detection in competitive ELISA KIT from 98pg/ml to 20pg/ml, and then use this method to measure the concentration of βCTX in saliva. Here we propose two methods. The first method is to increase the concentration of antigens in competitive ELISA KIT during the reaction. We can achieve this goal by increasing the volume of the antigens and decreasing the volume of the competitor during the reaction, and remain amount of .the competitor. The second method is to do more measuring groups. By increasing three measuring groups to between five and eight measuring groups, choosing median and two number near the median of all measuring groups and concentrations, then we can get the calibration curve with the lower limit of detection
By using 50 ul of antigen and competitor and increasing measure group to seven groups, this research successfully decreases limit of detection to 0.089~12.696 pg/ml. This situation can get the lowest limit od detection, but it is more expensive. By using 75 ul of antigen and 25 ul of competitor and increasing measure group to five groups, this research successfully decreases limit of detection to 3.485~48.833 pg/ml. Although this situation can not get the limit of detection like the previous situation, but it is cheaper and stabler.
The concentration of βCTX in the saliva of 17 subjects was successfully measured: The concentration of βCTX in the saliva of 10 healthy subjects:298.827~2404.51 pg/ml in seven healthy subjects, and the concentration of βCTX in the saliva of one of the healthy subjects exceeds the detecting range, and the concentration of βCTX in the saliva of the other two healthy subjects cannot be detected. The concentration of βCTX in the saliva of 7 patients:16.861~365.916 pg/ml in five patients, and the concentration of βCTX in the saliva of the other two patients cannot be detected.