以頂空注射器結合氣相層析質譜儀進行食品中二氧化硫殘留檢測及方法確效評估

Method Validation of Determination of Sulfur Dioxide Residues by Headspace-Gas Chromatography Mass Spectrometry

檔案

全文未公開。

日期

2020

類型

語言

中文

摘要

食品中的二氧化硫來自於亞硫酸鹽類,其為被廣泛使用之合法的食品添加物,具有漂白、防腐、殺菌、防止酵素性褐變與非酵素性褐變反應及抑制氧化等作用,然而過量的亞硫酸鹽會對身體造成不同程度之過敏反應,對亞硫酸鹽有過敏之患者,可能產生氣喘、喉嚨腫脹、呼吸困難,甚至死亡之症狀。
傳統之二氧化硫檢測方法為通氣蒸餾法,此方法雖然設備簡單、成本低,但裝置複雜、費時,且最後滴定步驟易受人為因素影響,造成實驗誤差。本研究利用頂空搭配氣相層析質譜儀(HS-GC-MS)建立二氧化硫殘留檢測之方法,其步驟簡單、快速、靈敏度高、基質影響小且可減少人為因素造成之實驗誤差。為建立本方法之最佳化條件,分別對頂空分析儀之平衡時間及平衡溫度與氣相層析儀之分流比進行探討。結果顯示烘箱以80℃加熱15分鐘,且Loop及Transfer Line之溫度設定130℃及140℃,GC分流比為25:1可獲得最佳果。並以此條件進行分析方法確效評估,其在檢量線範圍為5~250 μg/mL之濃度範圍下,線性迴歸方程式為y=9882.34x-5518.34,相關係數(R)可達0.999,回收率在82.6%~107.4%之間,變異係數低於4%,偵測極限為1 μg/mL,定量極限為5 μg/mL。最後藉由不同類型之樣品檢測並評估其回收率,其結果皆有達80%,證明本研究具有良好之適用性及可行性。
Sulfur dioxide in food comes from sulfites, and it has been wildly used in food additives for bleaxhing, preventing spoilage, inhibiting growth on bactorials, controlling enzymatic and non-enzymatic browning and avoiding oxidationand. However, excessive sulfites may cause allergic reactions such as asthma, swollen throat, dyspnea and evendeath.
Traditional method for sulfite detection is aeration distillation. Although this method use simple equipment with low cost, it is a complicated process and time consuming. Furthermore, it can be easily influenced by artificial factors and cause experimental errors. In this study, a procedure to detect sulfur dioxide residue by Headspace-Gas Chromatography Mass Spectrometry (HS-GC-MS) was developed. The high efficiency, sensitivity, simplicity and low matrix effect reduce experimental errors due to the artificial factors. In order to establish the optimal conditions, the equilibrium time and the equilibrium temperature in Headspace and the split ratio in GC were investigated. In this study, the oven equilibrium temperature at 80℃ for 15 minutes and the temperatures of the loop and the transfer line were 130℃ and 140℃, respectively. The split ratio was 25:1 in GC. Under the conditions, the calibration curve range was between 5 and 250 μg/mL, and linear regression equation was y=9882.34x-5518.34 (R=0.999). The recovery was between 82.6% and 107.4%, and the coefficient of variation values was less than 4%. The limit of detection was 1 μg/mL, and the limit of quantification was 5 μg/mL. Moreover, real samples were examined and the recovery was above 80%. The results proved that the method can be potentially used in real world samples.
顯示完整詮釋資料 統計資料

引用此條目

張凱婷 (2020). 以頂空注射器結合氣相層析質譜儀進行食品中二氧化硫殘留檢測及方法確效評估. https://thesis.lib.nycu.edu.tw/handle/20.500.14371/76282