採集大氣粗細微粒的無彈跳慣性衝擊器開發

Development of a non-bouncing inertial impactor for collecting ambient coarse and fine particles

檔案

日期

2021

類型

語言

中文

摘要

VSCC和WINS為目前美國環保署認可的PM2.5微粒分徑器,廣泛運用在FRM手動採樣及FEM自動監測儀,而長時間採樣會造成微粒彈跳及微粒負載,導致細微粒質量不準確性,且也無法得知粗微粒濃度。
本研究設計出一套無彈跳慣性衝擊器(Non-Bouncing Impactor, NBI),參考電子式多階衝擊器(EMCI,柯, 2017)其中兩階噴孔片 (Dpa50 = 5.6、2.5 μm)以及底部注水式衝擊板的設計,不僅減少微粒負載、微粒彈跳對採樣影響,未來可將沖洗的廢水進行離子分析。
設計慣性衝擊器,須符合標準規範。任一階噴孔片1和2差別為噴孔片1的噴嘴之間有間隔,而噴孔片2則無。實驗室收集效率測試中,無彈跳慣性衝擊器(NBI)第一階噴孔片1和2與一般或底部注水式的衝擊板 (Dpa50=5.57~5.74 μm、GSD = 1.13~1.16),第二階噴孔片1和2與一般或底部注水式的衝擊板(Dpa50=2.45~2.56 μm、GSD = 1.11~1.27),任一階的噴孔片1和2皆符合慣性衝擊器的規範。在微粒損失方面,無彈跳慣性衝擊器(NBI)的任一階噴孔片2比噴孔片1減少5~20 %的微粒損失。最後,將上述收集效率和微粒損失的結果與三種理想大氣分佈結合,得到總體損失的結果,結果顯示噴孔片2比噴孔片1減少5 ~ 8 %,確定實地採樣方面主要任一階以噴孔片2為主,減少因微粒損失可能會造成實場採樣的影響。
實地採樣方面,首先,設立PM10及PM2.5標準採樣道,三道PM10進口交叉比對和VSCC與FRM WINS比對結果,結果顯示誤差皆於範圍內。由於APS的粒徑偏移,將無彈跳慣性衝擊器第二階噴孔片2(Dpa50=2.5 μm)與VSCC進行採樣比對濃度偏差,結果顯示平均採樣誤差-0.91±4.88 %,證實第二階噴孔片2去除2.5μm微粒的效能,雖採樣誤差未超出美國環保署標準PM2.5微粒分徑器的標準5 %,但採樣期間都在低濃度(4.72~13.86 μg/m3)狀況下,可能是導致標準差較高的原因。無彈跳慣性衝擊器兩階噴孔片與表準採樣比對,使用濾紙夾放置鐵氟龍濾紙,確認實地採樣質量偏差,結果顯示PM10和PM2.5~10(Coarse particle)負偏差高,主要原因可能與慣性衝擊器噴嘴上的微粒損失有關,藉由總體損失的修正,質量濃度比對誤差只有-0.15±6.25 %和3.99±12.45 %。最後,進行無彈跳慣性衝擊器比對,更換成注水式衝擊板,粗微粒皆被沖洗,而細微粒的比對偏差為-1.71±4.45 %,證實無彈跳慣性衝擊器在實場採樣有很高的質量準確性。另外,測試將無彈跳慣性衝擊器沖洗的廢水與濾紙萃取分析進行比對,結果顯示雖文獻中提到玻璃纖維濾紙可進行離子分析,但較高的背景值嚴重干擾分析結果。
尋找可替代玻璃纖維濾紙,改善背景值干擾的影響後,將無彈跳慣性衝擊器進行粗微粒水溶性無機離子分析,發展成粗微粒水溶性無機離子即時監測系統,與平行板濕式固氣分離器−液體採樣器(PPWD-PILS)或本實驗開發的細微粒水溶性無機離子即時監測系統平行板濕式固氣分離器−半乾式靜電集塵器(PPWD-SDEP)結合,發展可完整即時監測粗細微粒與前驅氣體水溶性無機離子的系統。
VSCC and WINS are currently Inertial impactor approved by the U.S. Environmental Protection Agency. They are widely used in FRM manual sampling and FEM automatic monitors. Long-term sampling will cause particle bounce and loading, resulting in inaccuracy in the quality of fine particles, it is impossible to know the concentration of coarse particles.
In research, a set of non-bouncing inertial impactor (NBI) was designed, referring to the Electrical micro-orifice cascade impactor (EMCI, Ke, 2017) in which the two-stage orifice sheet (Dpa50 = 5.6, 2.5 μm) and the design of the water-injection on the bottom impaction plate not only reduces the impaction of particle loading and particle bouncing on sampling, but the flushing wastewater can be subjected to ion analysis in the future.
The design of the inertial impactor must comply with the standard specifications. The difference between nozzle 1 and 2 at any stage is that there is a gap between nozzles of nozzle 1, while nozzle 2 does not. In the collection efficiency test, the non-bouncing inertial impactor (NBI) first stage nozzles 1 and 2 and the general or bottom- injection impaction plate (Dpa50=5.57~5.74 μm, GSD = 1.13~1.16), the second stage nozzle 1 and 2 and general or bottom-injection impaction plates (Dpa50=2.45~2.56 μm, GSD = 1.11~1.27), any nozzle 1 and 2 of any stage meet the specifications of inertial impactors. In terms of particle loss, the nozzle 2 of any stage of the non-bouncing inertial impactor (NBI) reduces the particle loss by 5-20% compared with the nozzle 1. Finally, combining the above collection efficiency and particle loss results with the three ideal atmospheric distributions, the overall loss results are obtained. The results show that the nozzle 2 is 5-8% less than the nozzle 1. It is determined that any stage in the field sampling is mainly nozzle 2 is mainly used to reduce the impaction factor of real-time sampling due to particle loss.
In terms of field sampling, first, set up PM10 and PM2.5 standard sampling impactor, three PM10 inlet cross-comparisons, and VSCC and FRM WINS comparison results. The results showed that the errors were all within the standard range. Due to the particle size deviation of APS, the second stage nozzle 2 (Dpa50=2.5 μm) of the non-bouncing inertial impactor was compared with VSCC.The concentration deviation showed that the average sampling error was -0.91±4.88%, which confirmed the effectiveness of the second stage nozzle 2 in removing 2.5 μm particles, although the sampling error does not exceed the standard of the U.S.A Environmental Protection Agency standard PM2.5 impactor by 5%, but during the sampling period it is in a low concentration (4.72~13.86 μg/m3) condition,it may be the reason for the higher standard deviation. The two-stage nozzle of the non-bouncing inertial impactor was compared with the standard sampling PM10 and PM2.5 impactor, and the teflon filter was placed with a filter holder to confirm the quality deviation of the field sampling. The result showed that the negative deviation of PM10 and PM2.5~10 (Coarse particle) is high, the reason may be related to the particle loss on the inertial impactor nozzle. With the correction of the overall loss, the mass concentration comparison bias is only -0.15±6.25 % and 3.99±12.45 %. Finally, the non-bouncing inertial impactor was compared and replaced with a water-injected impaction plate. The coarse particles were washed, and the deviation of the comparison of fine particles was -1.71±4.45 %, which proved that the non-bouncing inertial impactor was very useful in field sampling. In addition, the ion analysis test compares the waste water from the non-bouncing inertial impactor with the filter. The results show that although the glass fiber filter paper can be used for ion analysis in the paper, the higher background value seriously interferes with the results.
After looking for an alternative glass fiber filter paper to improve the influence of background value interference, the non-bouncing inertial impactor is used to analyze water-soluble inorganic ions of the coarse particle, and developed into a semi-continuous monitoring of water-soluble ions for coarse particle, which is combined from the parallel plate wet denuder –particle-into-liquid sampler (PPWD-PILS) or the semi-continuous monitoring of water-soluble ions for fine particle of a parallel plate wet denude - a semi-dry electrostatic precipitator (PPWD-SDEP), can be developed in a complete semi-continuous monitoring of water-soluble ions for precursor gas、coarse and fine particle.

所屬典藏

環境工程研究所
顯示完整詮釋資料 統計資料

引用此條目

莊君皜 (2021). 採集大氣粗細微粒的無彈跳慣性衝擊器開發. https://thesis.lib.nycu.edu.tw/handle/20.500.14371/76958

BibTeX RIS

引用文字依 APA 第 7 版格式自動產生,使用前請再核對。

分享