應用於智慧型隱形眼鏡之光學式連續血糖監測系統設計與開發

Design and Development of An Optical Continuous Blood Glucose Monitoring System for Smart Contact Lenses

檔案

日期

2021

類型

語言

中文

摘要

本研究提出一測量房水血糖之量測系統,以光學架構為基礎,發展長時間監控血糖之非侵入式量測方式。
在本研究之系統量測中,本研究將整體系統分為光源端以及光接收端,先以相關論文佐證於隱形眼鏡上架構實現的可行性,再使用外部元件進行相關驗證系統的設計,以商用功率計量測出濃度與光功率之關係及偏振角對訊號線性度的影響,之後在隱形眼鏡架構驗證系統實驗中,以商用放大電路進行構建,電路中以光感測器感測功率變化,並佐以轉換電路及儀表放大器,並由壓控振盪器轉換為頻率輸出並以計頻器量測,成功測量出約20 mg/dl之解析度,確立了於隱形眼鏡上實現之可行性。
元件方面利用微機電製程方式來製作能與光源及15通道之驅動晶片進行封裝之元件,以聚一氯對二甲苯為主要結構,金為底下金屬走線,使用Mini-LED作為光源封裝之測試元件與底板接合,之後使用覆晶封裝方式將底板與驅動晶片結合,並成功封入隱形眼鏡之中,驗證了光源端於隱形眼鏡架構中的可行性。
This study proposes a measurement system for measuring blood glucose in aqueous humor. Based on an optical architecture, a non-invasive measurement method for long-term blood glucose monitoring is developed.
In the system measurement, this study divides the overall system into a light source end and a light receiving end. The feasibility of the architecture implementation on the contact lens is supported by related papers, and then the external components are used to design the related verification system. Commercial power meter is used to measure the relationship between concentration and optical power and the influence of polarization angle on signal linearity. Later, in the experiment of the contact lens architecture verification system, a commercial amplifier circuit is used to construct the circuit, and a light sensor is used to sense the power change. With the conversion circuit and instrumentation amplifier, the frequency output is converted from a voltage-controlled oscillator and measured with a frequency counter. The resolution of about 20 mg/dl is successfully measured, which confirms the feasibility of implementation on contact lenses.
In terms of components, micro-electromechanical process methods are used to produce components that can be packaged with light sources and 15-channel driver chips. The main structure is poly-p-xylene, gold is the underlying metal trace, and the Mini-LED is used as the light source package for testing. The components are joined to the bottom plate, and then the bottom plate and the driver chip are combined using flip-chip packaging, and successfully sealed in the contact lens, verifying the feasibility of the light source end in the contact lens structure.