以紫外光誘發臭氧氧化法移除一氧化氮對硫化氫監測系統分析偏差的影響
Reduction of NO-triggered Detection Bias by a UV-Induced Ozone Oxidation Method in a H2S Monitoring System
摘要
半導體晶圓廠空氣中的硫化氫(H2S)由於易與含銅薄膜進行反應產生缺陷,因此長期準確監測H2S氣體濃度成為廠區控制產品良率的重要需求。目前普遍用來監測H2S的儀器為紫外光螢光法分析儀,其主要是利用H2S氧化物-二氧化硫(SO2)的螢光發光值回推H2S含量,然而空氣中的一氧化氮(NO)卻會發出與SO2螢光光譜重疊的光譜,因而導致分析偏差與錯誤示警。為改善NO對H2S分析儀的干擾,本研究以紫外光誘發臭氧氧化法將NO轉化為NO2以降低NO干擾並提高分析儀對H2S濃度響應的準確性。實驗結果顯示,廠區空氣中含有38-67 ppbv NO與1.1-1.6 ppbv H2S,NO確實對分析儀造成正偏差影響,且響應值與NO濃度呈正相關(y=0.0149x+0.3581, R2=0.9938),透過紫外光誘發臭氧氧化法可大幅移除99.28 – 99.42% 的NO,但對H2S測值的影響僅0.5-1.8%,系統的方法偵測極限可達0.44 ppbv,相較於一般以活性炭移除NO但卻同步吸附98% H2S,紫外光誘發臭氧氧化法可有效降低NO干擾且低度影響H2S測值。當紫外光燈管輸出功率為75%時,對NO有最快移除速率,其速率常數可達1.64 min-1,且連續15天維持99%以上去除效率,將裝置安裝於H2S監測系統持續7天,錯誤示警次數可從3次降低為0次,估計一年能省下2,340,000元之採樣分析費用。
As H2S easily reacts Cu to cause defects, demand for H2S monitoring has be becoming emergent in chip factory in order to reach high product quality control. Ultraviolet-fluorescence analyzers are one of commonly used monitors for H2S detection. They mainly measure fluorescence of an oxidized product of H2S, SO2, to reflect concentration of H2S. However, NO in the air emits fluorescence within the fluorescence spectrum of SO2, thus causing analytical bias and false alarms. To reduce the NO interference, an UV-induced ozone oxidation method was developed to convert NO into NO2. Results showed that it contained 38-67 ppbv NO and 1.1-1.6 ppbv H2S in the ambient air in the plant. The H2S analyzer had a positive and linear response to NO concentrations (y=0.0149x+0.3581, R2=0.9938, where x and y are NO concentration and signal output, respectively). The UV-induced ozone oxidation greatly reduced 99.28-99.42% NO but had insignificant influence on H2S measurement (0.5-1.8%). The H2S detection limit was down to 0.44 ppbv with the UV-induced ozone oxidation. Compared to activated carbons, which adsorbed 98% H2S while NO removal, the UV-induced ozone oxidation effectively reduced the bias caused by NO and introduced a low impact on the H2S measurement. When the power of the UV lamp was 75%, NO removal rate was the highest and the rate constant was 1.64 min-1. The NO elimination equipment was able to continuously convert 99% NO for 15 days and reduced the number of false alarms for the H2S monitoring system from 3 to 0 within 7 days. With improved accuracy of the H2S monitoring system by the UV-induced ozone oxidation, approximate 2,340,000 NTD per year for H2S sampling and analysis can be saved.