以互補式總體經驗模態分解法實現心阻抗描記術的設計與實作
System design and implementation of impedance cardiography with CEEMD
摘要
心臟是人體的一個幫浦,驅動氧氣及養分的輸送來幫助人體完成血液循環,使各個器官可以正常運作,臨床上會藉由觀察心搏量(Stroke volume, SV)、心輸出量(Cardiac output)及左心室射血時間等血流動力學參數來瞭解心臟的健康情況。過去多採用侵入性的方式來量測血流動力學參數,Kubicek於1966年提出心阻抗描記術(Impedance cardiography, ICG),ICG以非侵入式的方法來進行連續性量測及SV估計。傳統上,ICG系統是由巴特沃斯濾波器(Butterworth filter, BF)來進行ICG訊號處理,BF拆解過程會造成訊號的相位偏移(Phase shift),影響動態量測下SV估算的結果;經驗模態分解法(Empirical mode decomposition, EMD)是由黃鄂院士於1998年提出的訊號拆解法,此方法善於處理非線性、非穩態的訊號,且拆解過程不具有相位差的問題,為了改善ICG在動態量測所遇到的問題,我們將此方法應用在ICG訊號處理上。
此研究中,實做出ICG系統並藉由BF及互補式總體經驗模態分解法(Complementary ensemble EMD, CEEMD)來進行ICG訊號處理的比較。本實驗共招募了5名受試者,實驗設計為三分鐘的基線量測,接著依序進行五分鐘不同頻率的Paced breathing,每分鐘6次、12次、20次及30次的呼吸循環(每次呼吸循環包含吸氣和吐氣),及騎乘腳踏車實驗,實驗過程中同步進行ICG、呼吸綁帶及都普勒超聲波(Doppler ultrasound, DUS)的量測,用以比對ICG系統在靜態、動態下的準確性並由呼吸綁帶獲得的呼吸訊號及呼吸率(Breathing rate, BR)來觀察心臟及呼吸調控之間的關聯性。
結果顯示:(1)由ICG訊號處理結果可以看出,經BF拆解得出的ICGBF有許多小波動,而經CEEMD所得出的ICGCEEMD較平滑,可以更容易辨別出ICG的特徵點,且BF的方式還伴隨著相位偏移的問題,而CEEMD則具有零相位差的特性,足以證明ICGCEEMD的拆解結果更優於ICGBF;(2) 將ICGBF及ICGCEEMD各別估算之SV和DUS估算之SV進行史皮爾曼等級相關的檢定,在相關性檢定的結果顯示,SVCEEMD與SVDUS在基線、12/20/30 cycle的Paced breathing及騎乘腳踏車各階段的相關係數rs分別為1.00, 0.90, 0.79, 0.80, 0.76,皆呈現高度相關,6-cycle的Paced breathing則呈中度相關;SVBF與SVDUS則僅有基線與6-cycle的Paced breathing呈高度相關,由此結果可以看出SVCEEMD與SVDUS具有較高的吻合姓;(3)為了驗證ICGCEEMD及DUS兩種方法的一致性,使用Bland-Altman plot來進行分析,結果顯示,40筆數據中只有一筆數據落在一致性界線範圍外,故兩種方法是具有一致性的。另外,此研究進一步討論由BF與CEEMD從阻抗訊號提取呼吸成分的可能性,CEEMD選取的呼吸成分和呼吸訊號在交互相關檢定中,皆具有高度相關性;並由頻譜分析及HR/BR比率觀察到呼吸對於心搏量變異度的影響。
總結此研究,在特定BR之呼吸調控與腳踏車騎乘的實驗中,透過自行設計之ICG量測電路來獲得原始數據,比較CEEMD和BF在訊號拆解上的差異,藉此觀察不同呼吸率下之SV並探討呼吸對於SV之關聯性。
此研究中,實做出ICG系統並藉由BF及互補式總體經驗模態分解法(Complementary ensemble EMD, CEEMD)來進行ICG訊號處理的比較。本實驗共招募了5名受試者,實驗設計為三分鐘的基線量測,接著依序進行五分鐘不同頻率的Paced breathing,每分鐘6次、12次、20次及30次的呼吸循環(每次呼吸循環包含吸氣和吐氣),及騎乘腳踏車實驗,實驗過程中同步進行ICG、呼吸綁帶及都普勒超聲波(Doppler ultrasound, DUS)的量測,用以比對ICG系統在靜態、動態下的準確性並由呼吸綁帶獲得的呼吸訊號及呼吸率(Breathing rate, BR)來觀察心臟及呼吸調控之間的關聯性。
結果顯示:(1)由ICG訊號處理結果可以看出,經BF拆解得出的ICGBF有許多小波動,而經CEEMD所得出的ICGCEEMD較平滑,可以更容易辨別出ICG的特徵點,且BF的方式還伴隨著相位偏移的問題,而CEEMD則具有零相位差的特性,足以證明ICGCEEMD的拆解結果更優於ICGBF;(2) 將ICGBF及ICGCEEMD各別估算之SV和DUS估算之SV進行史皮爾曼等級相關的檢定,在相關性檢定的結果顯示,SVCEEMD與SVDUS在基線、12/20/30 cycle的Paced breathing及騎乘腳踏車各階段的相關係數rs分別為1.00, 0.90, 0.79, 0.80, 0.76,皆呈現高度相關,6-cycle的Paced breathing則呈中度相關;SVBF與SVDUS則僅有基線與6-cycle的Paced breathing呈高度相關,由此結果可以看出SVCEEMD與SVDUS具有較高的吻合姓;(3)為了驗證ICGCEEMD及DUS兩種方法的一致性,使用Bland-Altman plot來進行分析,結果顯示,40筆數據中只有一筆數據落在一致性界線範圍外,故兩種方法是具有一致性的。另外,此研究進一步討論由BF與CEEMD從阻抗訊號提取呼吸成分的可能性,CEEMD選取的呼吸成分和呼吸訊號在交互相關檢定中,皆具有高度相關性;並由頻譜分析及HR/BR比率觀察到呼吸對於心搏量變異度的影響。
總結此研究,在特定BR之呼吸調控與腳踏車騎乘的實驗中,透過自行設計之ICG量測電路來獲得原始數據,比較CEEMD和BF在訊號拆解上的差異,藉此觀察不同呼吸率下之SV並探討呼吸對於SV之關聯性。
The heart is a pump of the human body, which drives the delivery of oxygen and nutrients to help the human body complete blood circulation, so that various organs can operate normally. In clinic, we will observe the hemodynamic parameters such as stroke volume (SV), cardiac output and left ventricular ejection time to understand the health of the heart. In the past, invasive methods were used to measure hemodynamic parameters. In 1966, Kubicek proposed impedance cardiography (ICG). ICG was used for continuous measurement and SV estimation in a non-invasive way. Traditionally, ICG system uses Butterworth filter (BF) to process ICG signal. The decomposition process of BF will cause phase shift of signal, which will affect the result of SV estimation under dynamic measurement; Empirical mode decomposition (EMD) is a signal decomposition method proposed by N. E. Huang in 1998. This method is suitable for dealing with non-linear and non-stationary signals, and there is no phase difference in the decomposition process. In order to improve the problems of ICG in dynamic measurement, we apply this method to ICG signal processing.
In this study, the ICG system is implemented, and the ICG signal processing is compared by using BF and the complementary empirical mode decomposition (CEEMD). A total of 5 subjects were recruited in this study. The experiment was designed as a three minute baseline measurement, followed by a five minute paced breathing (6, 12, 20 and 30 breathing cycles per minute, each including inhalation and exhalation) and biking experiment, during the experiment, ICG, respiratory inductive plethysmography (RIP) and doppler ultrasound (DUS) were measured simultaneously. DUS was used to compare the static and dynamic accuracy of ICG system, and the correlation between heart and respiratory regulation was observed by respiratory signal and breathing rate (BR) obtained by RIP.
The results show that: (1) It can be seen from the ICG signal processing results that ICGBF decomposed by BF has many ripples, but ICGCEEMD obtained by CEEMD is smoother so that it can more easily identify the characteristic points of ICG. There are phase shift problems in BF, and CEEMD has the characteristics of zero phase difference, it is enough to prove that the decomposition result of ICGCEEMD is better than ICGBF; (2)The SV estimated by ICGBF and ICGCEEMD and the SV estimated by DUS are tested for Spearman rank correlation coefficient. The results of the correlation test show that SVCEEMD and SVDUS are in the baseline, 12/20/30 cycle Paced breathing and biking. The correlation coefficients rs of each stage are 1.00, 0.90, 0.79, 0.80, 0.76, all showing high correlation, 6-cycle Paced breathing is moderately correlated; SVBF and SVDUS only have a high correlation in baseline and 6-cycle paced breathing. From this result, it can be seen that SVCEEMD and SVDUS have a higher coincidence; (3) In order to verify the agreement of ICGCEEMD and DUS, Bland Altman plot is used for analysis. The results show that only one of 40 data falls outside the consistency boundary, so there is agreement between ICG and DUS. In addition, this study further discusses the possibility of extracting respiratory components from impedance signals by BF and CEEMD. The respiratory components extracted by CEEMD are highly correlated with respiratory signals in cross-correlation test; The effect of breathing on stroke volume variation was observed by spectrum analysis and HR to BR ratio.
Summarizing this study, in the specific BR breathing regulation experiment and biking experiment, the raw data is obtained through the own ICG measurement circuit to compare the difference between CEEMD and BF signal decomposition, and then estimating SV under different BR to discuss the relationship between breathing and SV.
In this study, the ICG system is implemented, and the ICG signal processing is compared by using BF and the complementary empirical mode decomposition (CEEMD). A total of 5 subjects were recruited in this study. The experiment was designed as a three minute baseline measurement, followed by a five minute paced breathing (6, 12, 20 and 30 breathing cycles per minute, each including inhalation and exhalation) and biking experiment, during the experiment, ICG, respiratory inductive plethysmography (RIP) and doppler ultrasound (DUS) were measured simultaneously. DUS was used to compare the static and dynamic accuracy of ICG system, and the correlation between heart and respiratory regulation was observed by respiratory signal and breathing rate (BR) obtained by RIP.
The results show that: (1) It can be seen from the ICG signal processing results that ICGBF decomposed by BF has many ripples, but ICGCEEMD obtained by CEEMD is smoother so that it can more easily identify the characteristic points of ICG. There are phase shift problems in BF, and CEEMD has the characteristics of zero phase difference, it is enough to prove that the decomposition result of ICGCEEMD is better than ICGBF; (2)The SV estimated by ICGBF and ICGCEEMD and the SV estimated by DUS are tested for Spearman rank correlation coefficient. The results of the correlation test show that SVCEEMD and SVDUS are in the baseline, 12/20/30 cycle Paced breathing and biking. The correlation coefficients rs of each stage are 1.00, 0.90, 0.79, 0.80, 0.76, all showing high correlation, 6-cycle Paced breathing is moderately correlated; SVBF and SVDUS only have a high correlation in baseline and 6-cycle paced breathing. From this result, it can be seen that SVCEEMD and SVDUS have a higher coincidence; (3) In order to verify the agreement of ICGCEEMD and DUS, Bland Altman plot is used for analysis. The results show that only one of 40 data falls outside the consistency boundary, so there is agreement between ICG and DUS. In addition, this study further discusses the possibility of extracting respiratory components from impedance signals by BF and CEEMD. The respiratory components extracted by CEEMD are highly correlated with respiratory signals in cross-correlation test; The effect of breathing on stroke volume variation was observed by spectrum analysis and HR to BR ratio.
Summarizing this study, in the specific BR breathing regulation experiment and biking experiment, the raw data is obtained through the own ICG measurement circuit to compare the difference between CEEMD and BF signal decomposition, and then estimating SV under different BR to discuss the relationship between breathing and SV.