三維動態單光子電腦斷層半定量造影技術研究
Dynamic three-dimensional semi-quantitative analysis for SPECT imaging
摘要
研究背景與目的: 核子醫學影像有別於眾多醫學影像,藉由注射微量放射性示蹤劑,再由伽瑪閃爍攝影機擷取影像,可得到病患當下的身體最立即真實的生理定性描述甚至定量數據評估;目前臨床傳統核子醫學動態功能性影像分析,幾乎是以2D平面方式執行造影,並且以原廠後處理軟體加以分析。,然而人體器官的幾何形狀屬於立體且非規則形狀,只收取單一方向容易導致半定量的影像資訊會有缺失。本研究為了要突破目前以各原廠造影模式與分析方法所得到具有功能性半定量數值與影像,嘗試以現有傳統伽瑪閃爍攝影機模擬3D動態單光子電腦斷層(dynamic SPECT, D-SPECT)掃描模式,得到更加正確的功能性影像以及半定量分析數據。
方法: 本研究屬於前導行研究先以擬真器官假體能模擬動態流速功能,胃排空造影以3D列印製成胃臟動態假體,材質是聚乳酸(polylactic acid, PLA),胃假體模擬正常進食後放射活性會隨時間減少;唾液腺造影選擇壓克力材料製成唾液腺假體,是因為模擬人體唾液腺隨著時間活性隨之增加,並且模擬唾液腺受到刺激可以在短時間內快速排空的情況,因此時假體本身密度必須要足夠以確保排空時不會滲漏造成實驗誤差。針對兩類型假體,皆分別以2D平面動態造影與3D動態單光子電腦斷層(dynamic SPECT, D-SPECT)掃描模式進行造影,再以MATLAB所撰寫相對應之處理程式進行數據分析。
結果: 本實驗無論胃排空、唾液腺檢查,分別以2D平面動態造影與動態單光子電腦斷層造影模式皆進行造影與分析,兩者結果經由統計軟體SPSS以皮爾森相關性係數( Pearson’s correlation coefficient) 分析其相關性;布蘭德奧特曼差異圖(Bland-Altman difference plot)分析其兩種造影模式誤差值;並且於”模擬腸道活性重疊干擾胃排空正確性”實驗中,以”3D SPECT VOI ”概念特別針對三維空間”標的器官(target organ)”分析,與原本受到腸道活性干擾造成錯誤殘存百分比,以統計軟體SPSS使用成對樣品t檢定(paired sample t test)來分析,顯示其t統計:5.532、雙尾的P-value = 0.012,而能得到正確三維動態胃排空殘餘量活性曲線(time activity curve, TAC)及殘存百分比。
結論: 傳統伽瑪閃爍攝影機可以成功模擬動態單光子電腦斷層造影模式。動態單光子電腦斷層可以解決因空間上因器官重疊所造成半定量數據之誤差,進而可以取代傳統平面動態造影結果。本研究可以提供核醫科醫師、醫事放射師在每日例行臨床半定量造影程序中,在無需更改現有造影設備系統、不增加額外的臨床作業情況之下,進行三維動態單光子電腦斷層半定量造影。
關鍵字:
動態單光子電腦斷層、半定量分析、時間活度曲線、感興趣體積、
快速旋轉機架
方法: 本研究屬於前導行研究先以擬真器官假體能模擬動態流速功能,胃排空造影以3D列印製成胃臟動態假體,材質是聚乳酸(polylactic acid, PLA),胃假體模擬正常進食後放射活性會隨時間減少;唾液腺造影選擇壓克力材料製成唾液腺假體,是因為模擬人體唾液腺隨著時間活性隨之增加,並且模擬唾液腺受到刺激可以在短時間內快速排空的情況,因此時假體本身密度必須要足夠以確保排空時不會滲漏造成實驗誤差。針對兩類型假體,皆分別以2D平面動態造影與3D動態單光子電腦斷層(dynamic SPECT, D-SPECT)掃描模式進行造影,再以MATLAB所撰寫相對應之處理程式進行數據分析。
結果: 本實驗無論胃排空、唾液腺檢查,分別以2D平面動態造影與動態單光子電腦斷層造影模式皆進行造影與分析,兩者結果經由統計軟體SPSS以皮爾森相關性係數( Pearson’s correlation coefficient) 分析其相關性;布蘭德奧特曼差異圖(Bland-Altman difference plot)分析其兩種造影模式誤差值;並且於”模擬腸道活性重疊干擾胃排空正確性”實驗中,以”3D SPECT VOI ”概念特別針對三維空間”標的器官(target organ)”分析,與原本受到腸道活性干擾造成錯誤殘存百分比,以統計軟體SPSS使用成對樣品t檢定(paired sample t test)來分析,顯示其t統計:5.532、雙尾的P-value = 0.012,而能得到正確三維動態胃排空殘餘量活性曲線(time activity curve, TAC)及殘存百分比。
結論: 傳統伽瑪閃爍攝影機可以成功模擬動態單光子電腦斷層造影模式。動態單光子電腦斷層可以解決因空間上因器官重疊所造成半定量數據之誤差,進而可以取代傳統平面動態造影結果。本研究可以提供核醫科醫師、醫事放射師在每日例行臨床半定量造影程序中,在無需更改現有造影設備系統、不增加額外的臨床作業情況之下,進行三維動態單光子電腦斷層半定量造影。
關鍵字:
動態單光子電腦斷層、半定量分析、時間活度曲線、感興趣體積、
快速旋轉機架
Background and Purpose: Nuclear medicine imaging is different from many medical images. By injecting a small amount of radioactive tracer, and then capturing the image by a gamma scintillation camera, the patient's immediate body status, true physiological qualitative description can be obtained and even quantitative data evaluation; In the current traditional clinical nuclear medicine dynamic functional image analysis, imaging is almost performed in a two-dimensional planar mode and analyzed by original equipment manufacturer’s post-processing software. However, the geometric shape of human organs is three-dimensional and irregular. Projection from a single direction can easily lead to the lack of semi-quantitative image information. In order to break through the current limit of functional semi-quantitative values and images obtained with various original imaging modes and analysis methods, this research simulated a 3D dynamic single-photon computed tomography (dynamic SPECT, D-SPECT) scans with existing traditional gamma scintillation cameras mode to obtain more accurate functional images and semi-quantitative analysis data.
Methods: This research is a pioneer study. Firstly, a virtual organ phantom was used to simulate the dynamic flow rate function. 3D printed gastric phantom was used to aid the dynamic gastric emptying. The material used in the phantom is polylactic acid (PLA), and the gastric phantom simulates normal eating situation: after eating, the radioactivity in the stomach will decrease over time。In the salivary gland imaging, we chose acrylic material to make the salivary gland phantom because the activity of the simulated human salivary glands increases with time, and the simulated acrylic salivary glands can be emptied quickly in a short time after being stimulated. Therefore, the density of the phantom itself must be sufficient to ensure that it will not leak during emptying and cause experimental errors. For the two types of phantoms, both 2D planar dynamic imaging and dynamic single-photon computed tomography (dynamic SPECT, D-SPECT) scanning modes were used for imaging, and then the corresponding processing programs and data analysis were written by MATLAB.
Results: In this experiment, both the gastric emptying scintigraphy and salivary gland scintigraphy were analyzed in the 2D planar dynamic imaging and the D-SPECT. The results were processed in statistical software SPSS, and the Pearson correlation coefficient was used to analyze the correlation; Bland-Altman plot was used to analyze the deviation. In the simulated gastric emptying experiement with intestinal activity, the concept of "3D SPECT VOI" was used to analyze the three-dimensional target organ and the percentage of residual error caused by the interference with intestinal activity. We used SPSS to do the paired sample t test for statistical analysis, it showed that its t: 5.532, the two-tailed P-value = 0.012, the accurate three-dimensional dynamic gastric emptying time activity curve (TAC) and residual volume were obtained.
Conclusion: The traditional gamma scintillation camera can successfully simulate the dynamic single-photon computed tomography mode. D-SPECT can solve the errors in semi-quantitative data caused by overlapping organs in space, and can replace traditional planar dynamic imaging results. This research can provide nuclear physicians and technologists clinical semi-quantitative information in the imaging procedure in the daily routine, and three-dimensional D-SPECT semi-quantitative technology is performed without changing the existing imaging system and without adding additional clinical work.
Keywords:
Dynamic single-photon emission computed tomography, Semi-quantitative analysis, time‐activity curve (TAC), Volume of Interest (VOI), Rapid-rotating gantry
Methods: This research is a pioneer study. Firstly, a virtual organ phantom was used to simulate the dynamic flow rate function. 3D printed gastric phantom was used to aid the dynamic gastric emptying. The material used in the phantom is polylactic acid (PLA), and the gastric phantom simulates normal eating situation: after eating, the radioactivity in the stomach will decrease over time。In the salivary gland imaging, we chose acrylic material to make the salivary gland phantom because the activity of the simulated human salivary glands increases with time, and the simulated acrylic salivary glands can be emptied quickly in a short time after being stimulated. Therefore, the density of the phantom itself must be sufficient to ensure that it will not leak during emptying and cause experimental errors. For the two types of phantoms, both 2D planar dynamic imaging and dynamic single-photon computed tomography (dynamic SPECT, D-SPECT) scanning modes were used for imaging, and then the corresponding processing programs and data analysis were written by MATLAB.
Results: In this experiment, both the gastric emptying scintigraphy and salivary gland scintigraphy were analyzed in the 2D planar dynamic imaging and the D-SPECT. The results were processed in statistical software SPSS, and the Pearson correlation coefficient was used to analyze the correlation; Bland-Altman plot was used to analyze the deviation. In the simulated gastric emptying experiement with intestinal activity, the concept of "3D SPECT VOI" was used to analyze the three-dimensional target organ and the percentage of residual error caused by the interference with intestinal activity. We used SPSS to do the paired sample t test for statistical analysis, it showed that its t: 5.532, the two-tailed P-value = 0.012, the accurate three-dimensional dynamic gastric emptying time activity curve (TAC) and residual volume were obtained.
Conclusion: The traditional gamma scintillation camera can successfully simulate the dynamic single-photon computed tomography mode. D-SPECT can solve the errors in semi-quantitative data caused by overlapping organs in space, and can replace traditional planar dynamic imaging results. This research can provide nuclear physicians and technologists clinical semi-quantitative information in the imaging procedure in the daily routine, and three-dimensional D-SPECT semi-quantitative technology is performed without changing the existing imaging system and without adding additional clinical work.
Keywords:
Dynamic single-photon emission computed tomography, Semi-quantitative analysis, time‐activity curve (TAC), Volume of Interest (VOI), Rapid-rotating gantry