光伏供電下視網膜晶片電極架構與影像動態範圍之改進設計及其植入應用
THE IMPROVED DESIGN OF PHOTOVOLTAIC-POWERED SUBRETINAL PROSTHETIC CHIPS ON IMAGE DYNAMIC RANGE AND ELECTRODE STRUCTURES FOR PROSTHETIC SYSTEM APPLICATIONS
摘要
本篇論文提出應用於下視網膜植入晶片系統,晶片採用台積電0.18um互補式金氧半CIS製程工藝,並在產出後在晶片之刺激電極濺鍍上了二氧化銥(IrOx)。在本篇論文提出的新設計中,電壓箝制電路被加在四倍電荷泵升壓電路的輸出,用來防止電刺激的過壓,並同時起到穩定刺激電流和刺激頻率的效果。基於晶片采用分區供電方式,在CHIP II和CHIP III的設計中我們提出了電極共用的架構,此架構通過共用相臨像素的電極,達到增加電極面積和降低電極阻抗的效果;同時設計了主動式背景光消除電路,用感測環境中的光強度,將其從感光像素感測到的影像中扣除,防止感光像素飽和,進而提昇感光範圍。
經量測,在100mW/cm2 的IR照射下CHIP I可以穩定輸出26Hz的電刺激,感光範圍為150~740Lux,並可在740Lux的光訊號下輸出8.05nC的最大刺激電荷。在94mW/cm2 的IR照射下CHIP II可以穩定輸出30.4Hz的電刺激,在主動式背景光消除電路的運作下感光範圍為416~7346Lux,並可輸出9.8nC的最大刺激電荷。在74mW/cm2 的IR照射下CHIP III可以穩定輸出17.8Hz的電刺激,在主動式背景光消除電路的運作下感光範圍為290~9393Lux,並可輸出13.3nC的最大刺激電荷。
在生物體外實驗中,新一版的晶片也可成功刺激野生小鼠和視網膜退化之小鼠( rd1 )之視網膜神經節細胞(RGCs)並利用膜片箝制(Patch Clamp)技術記錄小鼠的電生理訊號。
經量測,在100mW/cm2 的IR照射下CHIP I可以穩定輸出26Hz的電刺激,感光範圍為150~740Lux,並可在740Lux的光訊號下輸出8.05nC的最大刺激電荷。在94mW/cm2 的IR照射下CHIP II可以穩定輸出30.4Hz的電刺激,在主動式背景光消除電路的運作下感光範圍為416~7346Lux,並可輸出9.8nC的最大刺激電荷。在74mW/cm2 的IR照射下CHIP III可以穩定輸出17.8Hz的電刺激,在主動式背景光消除電路的運作下感光範圍為290~9393Lux,並可輸出13.3nC的最大刺激電荷。
在生物體外實驗中,新一版的晶片也可成功刺激野生小鼠和視網膜退化之小鼠( rd1 )之視網膜神經節細胞(RGCs)並利用膜片箝制(Patch Clamp)技術記錄小鼠的電生理訊號。
The proposed CMOS photovoltaic-powered implantable chips with active pixel sensors and iridium-oxide electrodes for subretinal prostheses are designed and fabricated by 180nm 1P6M CIS technology. In the design, voltage clamping circuit is added to improve the safety issue cause by water window and the stimulation stability. The sharing electrode scheme is proposed to lower the loading of stimulation and raise the pixel number. The adaptive background cancellation circuit is designed to widen the sensing range of active pixel sensor. From the electrical measurement results, the measured maximum output stimulation current of Chip I is 6.6 μA. The measured stimulation frequency is 26 Hz and the amount of injected charges at each pixel is 8.05 nC. The detectable image illuminance is 150 Lux to 740 Lux under IR light intensity of 100 mW/cm2. The measured maximum output stimulation current of CHIP II is 9 uA. The measured stimulation frequency is 30.4 Hz and the amount of injected charges at each pixel is 9.8 nC. The detectable image illuminance with ABCC is 416 Lux to 7346 Lux under IR light intensity of 94 mW/cm2. The measured maximum output stimulation current of CHIP III is 13.4 uA. The measured stimulation frequency is 17.8 Hz and the amount of injected charges at each pixel is 13.3 nC. The detectable image illuminance with ABCC is 290 Lux to 9393 Lux under IR light intensity of 74 mW/cm2. The function of the proposed chip (CHIP II and III) has been further validated successfully with the ex vivo experimental results by recording the electrophysiological responses of retinal ganglion cells (RGCs) of retinas from wild type (C57BL/6) and retinal degeneration (rd1) mice with a patch clamp. Both of chips show that during the trials of stimulation, the evoked spikes behave consistently.