噴墨列印式軟性硫化氫感測電極設計與製造
Design and fabrication of flexible inkjet printed hydrogen sulfide sensing electrode
摘要
在生理環境中硫化氫主要經由多種含硫分子,如含硫氨基酸、蛋氨酸、半胱氨酸和胱氨酸
等,經由 Cystathionine-β-synthase(CBS)與 cystathionine γ−lyase (CSE)兩種酶催化而
生成。在人體血液中硫化氫濃度為 30-100μM,而近年研究發現在許多不同病症病患的體內硫
化氫濃度也有所不同,故硫化氫已被視為繼一氧化碳(CO)與一氧化氮(NO)後第三種被發現的氣
體信號因子。而現行用於量測生物環境中硫化氫的方式多為需要進行化學預處理後量測,耗時
且易破壞樣品。本研究利用噴墨式列印技術於軟性基版 Kapton 上實現利用恆電位庫倫分析法
(amperometry)以直接量測樣品中硫化氫水平。透過噴墨列印技術所實現硫化氫感測器,其特
色包含 1.於軟性基版上噴印奈米銀顆粒(Ag NPs)所製成電極,可利用還原氧化石墨烯修飾電
極表面與於作為抗干擾層的納菲薄膜參雜氧化還原對六胺釕以進一步增加電極敏感度與可靠
度. 2.該電極可整合噴墨式列印技術製程之噴墨銀/氯化銀做為參考電極以實現微型化感測
器。本研究針對工作電極抗干擾能力、量測回復性、連續監測可能性與參考電極穩定度進行討
論,研究發現利用石墨烯修飾銀電極與於納菲薄膜薄膜中參雜氧化還原對之工作電極可以使
靈敏度提升至 59.71nA/μM/cm2,而電極偵測極限為 0.26μM。此外經過測試發現該電極可
有效阻絕抗壞血酸、尿素、多巴胺所造成訊號干擾。且在長時間量測的應用方面,本研究更提
出於電極閒置於溶液期間利用低電位以防止硫化物累積於工作電極表面造成電流偏差;在參考
電極方面於不同濃度硫化氫溶液中電極所造成電流偏差僅為 0.3%,最後本研究也嘗試量測雞
蛋加熱過程中含硫氨基酸受熱氧化所釋放出硫化氫,以展示該感測電極實時量測可行性。透過噴墨式列印製程實現製程簡單、低成本、高控制性與重複性之軟性感測電極,將可使未來使用
場域更加多元並利於感測器微型化。
等,經由 Cystathionine-β-synthase(CBS)與 cystathionine γ−lyase (CSE)兩種酶催化而
生成。在人體血液中硫化氫濃度為 30-100μM,而近年研究發現在許多不同病症病患的體內硫
化氫濃度也有所不同,故硫化氫已被視為繼一氧化碳(CO)與一氧化氮(NO)後第三種被發現的氣
體信號因子。而現行用於量測生物環境中硫化氫的方式多為需要進行化學預處理後量測,耗時
且易破壞樣品。本研究利用噴墨式列印技術於軟性基版 Kapton 上實現利用恆電位庫倫分析法
(amperometry)以直接量測樣品中硫化氫水平。透過噴墨列印技術所實現硫化氫感測器,其特
色包含 1.於軟性基版上噴印奈米銀顆粒(Ag NPs)所製成電極,可利用還原氧化石墨烯修飾電
極表面與於作為抗干擾層的納菲薄膜參雜氧化還原對六胺釕以進一步增加電極敏感度與可靠
度. 2.該電極可整合噴墨式列印技術製程之噴墨銀/氯化銀做為參考電極以實現微型化感測
器。本研究針對工作電極抗干擾能力、量測回復性、連續監測可能性與參考電極穩定度進行討
論,研究發現利用石墨烯修飾銀電極與於納菲薄膜薄膜中參雜氧化還原對之工作電極可以使
靈敏度提升至 59.71nA/μM/cm2,而電極偵測極限為 0.26μM。此外經過測試發現該電極可
有效阻絕抗壞血酸、尿素、多巴胺所造成訊號干擾。且在長時間量測的應用方面,本研究更提
出於電極閒置於溶液期間利用低電位以防止硫化物累積於工作電極表面造成電流偏差;在參考
電極方面於不同濃度硫化氫溶液中電極所造成電流偏差僅為 0.3%,最後本研究也嘗試量測雞
蛋加熱過程中含硫氨基酸受熱氧化所釋放出硫化氫,以展示該感測電極實時量測可行性。透過噴墨式列印製程實現製程簡單、低成本、高控制性與重複性之軟性感測電極,將可使未來使用
場域更加多元並利於感測器微型化。
In the physiological environment, hydrogen sulfide is mainly catalyzed by a variety of sulfur
containing molecules, such as sulfur-containing amino acids, methionine, cysteine and cystine,
etc., and is catalyzed by two enzymes, Cystathionine-β-synthase (CBS) and cystathionine γ
lyase (CSE). The level of hydrogen sulfide in human blood is 30-100μM. Recent researches
have shown that the levels of hydrogen sulfide in patients with many different diseases are
different and it has been regarded as the third gas signal factor to be discovered after carbon
monoxide (CO) and nitric oxide (NO). The present methods used to measure hydrogen sulfide
in the biological environment mostly require chemical pre-treatment and time-consuming,
which easily cause detrimental effects on tested samples. In this thesis, we successfully develop
inkjet printed flexible hydrogen sulfide sensors facilitating directly measure the hydrogen
sulfide level based on the constant potential coulomb analysis method (amperometry). Several
critical technology features of the as-developed sensor include 1. the sensing electrode made
of printed silver particles (Ag NPs) on a soft substrate can exhibit excellent sensitivity and
reliability via the surface characteristics electrochemically modified by the printed reduced
graphite oxide and the Nafion film doped with hexamine ruthenium redox pairs. 2. the printed electrode can be monolithically integrated with the inkjet silver/silver chloride reference
electrode for the realization of miniaturized sensors. In this study, the anti-interference ability
reliability, the feasibility of continuous monitoring and the stability of the printed sensing and
reference electrode have been fully investigated and it has been found that the printed silver
electrode modified can exhibit a sensitivity of 59.71nA/μM/cm2 and 0.26μM of LoD.
Meanwhile, experimental results show the sensing electrode can effectively block signal
interference resulting from ascorbic acid, urea acid, and dopamine, etc.., the sensing electrode
biased at a reduction potential during non-operational period in the solution can also effectively
prevent sulfide formation on the surface of the sensing electrode, and the printed reference
electrode can exhibit excellent stability only with 0.3% sensing current deviation while being
placed in the solutions with different concentrations of hydrogen sulfide. A measurement of
the hydrogen sulfide released by the thermal oxidation of sulfur-containing amino acids from
heated eggs has demonstrated the feasibility of real-time measurement of the sensing electrode.
We think the developed sensor with simple manufacturing process, low cost, high
controllability and repeatability and the corresponding planar electrodes characteristic on a
flexible substrate will have not only more extensively sensing applications but also better size
scaling ability for sensor miniaturization.
containing molecules, such as sulfur-containing amino acids, methionine, cysteine and cystine,
etc., and is catalyzed by two enzymes, Cystathionine-β-synthase (CBS) and cystathionine γ
lyase (CSE). The level of hydrogen sulfide in human blood is 30-100μM. Recent researches
have shown that the levels of hydrogen sulfide in patients with many different diseases are
different and it has been regarded as the third gas signal factor to be discovered after carbon
monoxide (CO) and nitric oxide (NO). The present methods used to measure hydrogen sulfide
in the biological environment mostly require chemical pre-treatment and time-consuming,
which easily cause detrimental effects on tested samples. In this thesis, we successfully develop
inkjet printed flexible hydrogen sulfide sensors facilitating directly measure the hydrogen
sulfide level based on the constant potential coulomb analysis method (amperometry). Several
critical technology features of the as-developed sensor include 1. the sensing electrode made
of printed silver particles (Ag NPs) on a soft substrate can exhibit excellent sensitivity and
reliability via the surface characteristics electrochemically modified by the printed reduced
graphite oxide and the Nafion film doped with hexamine ruthenium redox pairs. 2. the printed electrode can be monolithically integrated with the inkjet silver/silver chloride reference
electrode for the realization of miniaturized sensors. In this study, the anti-interference ability
reliability, the feasibility of continuous monitoring and the stability of the printed sensing and
reference electrode have been fully investigated and it has been found that the printed silver
electrode modified can exhibit a sensitivity of 59.71nA/μM/cm2 and 0.26μM of LoD.
Meanwhile, experimental results show the sensing electrode can effectively block signal
interference resulting from ascorbic acid, urea acid, and dopamine, etc.., the sensing electrode
biased at a reduction potential during non-operational period in the solution can also effectively
prevent sulfide formation on the surface of the sensing electrode, and the printed reference
electrode can exhibit excellent stability only with 0.3% sensing current deviation while being
placed in the solutions with different concentrations of hydrogen sulfide. A measurement of
the hydrogen sulfide released by the thermal oxidation of sulfur-containing amino acids from
heated eggs has demonstrated the feasibility of real-time measurement of the sensing electrode.
We think the developed sensor with simple manufacturing process, low cost, high
controllability and repeatability and the corresponding planar electrodes characteristic on a
flexible substrate will have not only more extensively sensing applications but also better size
scaling ability for sensor miniaturization.