基於時序共享之時間數位轉換器架構實現可程式化電容感測陣列應用於細胞檢測
Programmable Capacitive Sensor Array with Ts-TDC Architecture for Cell Detection
摘要
近幾年來,實驗室晶片(Lab-on-a-Chip, LOC)技術迅速發展,已開始應用於有關生物的檢測,其提高了檢測的效率及速率,並具有節省試劑消耗、易整合性和可攜帶性。
因此,在此篇論文中,我們提出一款具460,800 像素電極之可程式化生醫晶片應用於觀察細胞型態,透過可程式化系統控制電容感測陣列(Capacitive Sensor Array, CSA),每一電極單元(Electrode Cell) 均具有8×8 個像素電極(Pixel Electrode)、行列開關、感測電路和內建自我測試(Built-In-Self-Testing, BIST) 掃描鍊,並且縮小像素電極面積至2.25×2.25 μm2 實現較高的空間解析度,解決傳統生醫晶片解析度不足的瓶頸。同時,使用時序共享之時間數位轉換器(Time-Sharing Time-to-Digital Converter, Ts-TDC) 架構在同一時脈週期下對電容感測陣列進行運作,來簡化電路設計的複雜度且提高時間效率。此外,我們透過一個直觀的圖形化使用者介面將感測結果傳入,呈現樣品的型態以達成即時的量測平台。此設計使用台積電0.18 μm CMOS 製程。從實驗結果可看出液滴型態在使用者介面上和顯微鏡下相互對應關係,提供使用者更好的觀察和操作體驗,本文所提出之電容感測陣列生醫晶片能實現可程式化、攜帶性和易觀察性於整合式自動化檢測平台。
因此,在此篇論文中,我們提出一款具460,800 像素電極之可程式化生醫晶片應用於觀察細胞型態,透過可程式化系統控制電容感測陣列(Capacitive Sensor Array, CSA),每一電極單元(Electrode Cell) 均具有8×8 個像素電極(Pixel Electrode)、行列開關、感測電路和內建自我測試(Built-In-Self-Testing, BIST) 掃描鍊,並且縮小像素電極面積至2.25×2.25 μm2 實現較高的空間解析度,解決傳統生醫晶片解析度不足的瓶頸。同時,使用時序共享之時間數位轉換器(Time-Sharing Time-to-Digital Converter, Ts-TDC) 架構在同一時脈週期下對電容感測陣列進行運作,來簡化電路設計的複雜度且提高時間效率。此外,我們透過一個直觀的圖形化使用者介面將感測結果傳入,呈現樣品的型態以達成即時的量測平台。此設計使用台積電0.18 μm CMOS 製程。從實驗結果可看出液滴型態在使用者介面上和顯微鏡下相互對應關係,提供使用者更好的觀察和操作體驗,本文所提出之電容感測陣列生醫晶片能實現可程式化、攜帶性和易觀察性於整合式自動化檢測平台。
In recent years, Lab-on-a-Chip technology has rapidly developed and been applied for biological detection. It enhances the efficiency and speed of detection, and also has the advantages of saving reagent consumption, easy integration, and portability.
Therefore, in this thesis, we propose a programmable biomedical chip with 460,800 pixel electrodes for observing cell morphology through the programmable system control Capacitive Sensor Array (CSA). Each Electrode Cell has 8×8 Pixel Electrodes, row/column switches, sensing circuits and Built-In-Self-Testing (BIST) scan chains. We also reduce the area of the pixel electrode to 2.25×2.25 μm2 to achieve higher spatial resolution, which solves the bottleneck of insufficient resolution of traditional biomedical chips. At the same time, we use Time-Sharing Time-to-Digital Converter (Ts-TDC) architecture to simplify circuit complexity and boost time efficiency by operating on CSA within the same clock period. Additionally, we utilize an intuitive graphical user interface to input the sensing results and visually present the morphology of the sample for achieving real-time measurement platform.
This system has been implemented with TSMC 0.18 μm CMOS standard process. The experimental results present a correspondence between the droplet morphology observed under the microscope and on the user interface, which provides the better observation and operation experience for users. The proposed CSA biomedical chip can achieve programmability, portability and observability in an integrated automated detection platform.
Therefore, in this thesis, we propose a programmable biomedical chip with 460,800 pixel electrodes for observing cell morphology through the programmable system control Capacitive Sensor Array (CSA). Each Electrode Cell has 8×8 Pixel Electrodes, row/column switches, sensing circuits and Built-In-Self-Testing (BIST) scan chains. We also reduce the area of the pixel electrode to 2.25×2.25 μm2 to achieve higher spatial resolution, which solves the bottleneck of insufficient resolution of traditional biomedical chips. At the same time, we use Time-Sharing Time-to-Digital Converter (Ts-TDC) architecture to simplify circuit complexity and boost time efficiency by operating on CSA within the same clock period. Additionally, we utilize an intuitive graphical user interface to input the sensing results and visually present the morphology of the sample for achieving real-time measurement platform.
This system has been implemented with TSMC 0.18 μm CMOS standard process. The experimental results present a correspondence between the droplet morphology observed under the microscope and on the user interface, which provides the better observation and operation experience for users. The proposed CSA biomedical chip can achieve programmability, portability and observability in an integrated automated detection platform.