手持式週期梯度導模共振檢測系統整合
Integration of handheld gradient guided-mode resonance detection system
摘要
早期的疾病檢測需要依賴大型且昂貴的儀器,並由專業人員進行繁瑣檢驗,較難滿足定點照護、即時檢測的需求,因此目標是開發具備操作簡易、即時檢測、低成本、可攜式的手持生物感測器平台。本研究感測晶片為利用導模共振原理製造出的週期梯度導模共振晶片,結合發光二極體、濾光片、偏振片等光路架設,並使用開發板STM32F767控制感光耦合元件進行數據接收,整合為一光學式生物感測平台。
以上述元件進行蔗糖水實驗量測得出靈敏度為2520.95 µm/RIU,檢測極限為3.23 x 10-3 RIU,接著進行肌酸酐量測,經由four parameter logistic (4PL) regression擬合後,得出檢測極限為60.34 µg/ml,表現此系統在生物分子研究方面的潛力。最後將開發板與感光耦合元件整合成印刷電路板,未來進行光路元件方面的優化後,與印刷電路板進行整合,成功檢測生物分子後,將有望達成一完整的手持裝置且兼具良好的檢測極限,達到最初的定點照護的目標。
以上述元件進行蔗糖水實驗量測得出靈敏度為2520.95 µm/RIU,檢測極限為3.23 x 10-3 RIU,接著進行肌酸酐量測,經由four parameter logistic (4PL) regression擬合後,得出檢測極限為60.34 µg/ml,表現此系統在生物分子研究方面的潛力。最後將開發板與感光耦合元件整合成印刷電路板,未來進行光路元件方面的優化後,與印刷電路板進行整合,成功檢測生物分子後,將有望達成一完整的手持裝置且兼具良好的檢測極限,達到最初的定點照護的目標。
The detection of disease in the past needs to rely on large and expensive equipment, and it always need to be operated by professionals. Therefore, it couldn’t reach the goal of “point of care” or “instant detection”. So, our goal is to develop a handheld biosensor detection platform with easy operation, instant detection and low cost. In this research, a gradient grating period guided-mode resonant (GGP-GMR) filter was used as a biosensor. The detection platform consists of a light-emitting diodes, filter, polarizer. A homemade circuit board integrated with a charge coupled device (CCD) was developed to read out the signal intensity from the CCD.
The measurement results from various sucrose solutions indicate the sensitivity can reach 2520.95 µm/RIU, and the limit of detection (LOD) can achieve 3.23 x 10-3 RIU. In addition, the creatinine detection based on immunoassay shows that the LOD of 60.34 µg/ml can be achieved. Lastly, we developed a circuit board with an integrated CCD. In the near future, we will test the circuit board for sucrose and biomolecule measurement. We believe the complete system can be used for point of care applications.
The measurement results from various sucrose solutions indicate the sensitivity can reach 2520.95 µm/RIU, and the limit of detection (LOD) can achieve 3.23 x 10-3 RIU. In addition, the creatinine detection based on immunoassay shows that the LOD of 60.34 µg/ml can be achieved. Lastly, we developed a circuit board with an integrated CCD. In the near future, we will test the circuit board for sucrose and biomolecule measurement. We believe the complete system can be used for point of care applications.