基於慣性感測元件之即時癲癇偵測系統與電路設計
IMU-based Real-Time Epilepsy Detection Processor Design
摘要
本論文提出僅需使用一個慣性感測元件的即時癲癇偵測系統,在二到六天的長時間監測下,患者每次發作不超過五分鐘。
透過對發作訊號的觀察,巨觀上,癲癇發作的趨勢能將大部分的非發作時間濾除;微觀上,癲癇發作的特徵波形可以提升靈敏度(sensitivity)。
經由發作時間的比對,本論文呈現75.64%靈敏度(sensitivity)、99.81%準確性(accuracy)、99.81%特異度(specificity)和0.19%偽陽性率(false positive rate),本論文不僅保證了可靠性,也為診斷提供量化分析的成果。此系統採用台積電0.18 μm製程進行晶片實作,以證實發展成穿戴式裝置的可行性,電路的工作頻率是438kHz,功率消耗為2.1678mW,晶片面積為3 × 3 mm2。
透過對發作訊號的觀察,巨觀上,癲癇發作的趨勢能將大部分的非發作時間濾除;微觀上,癲癇發作的特徵波形可以提升靈敏度(sensitivity)。
經由發作時間的比對,本論文呈現75.64%靈敏度(sensitivity)、99.81%準確性(accuracy)、99.81%特異度(specificity)和0.19%偽陽性率(false positive rate),本論文不僅保證了可靠性,也為診斷提供量化分析的成果。此系統採用台積電0.18 μm製程進行晶片實作,以證實發展成穿戴式裝置的可行性,電路的工作頻率是438kHz,功率消耗為2.1678mW,晶片面積為3 × 3 mm2。
In this thesis, we proposed a real-time epilepsy detection system with a single inertial measurement unit (IMU). Under the long-term monitoring of two to six days, patients does not exceed five minutes per seizure.
Through the observation of seizure signals, macroscopically, the trend of epileptic seizures can filter out most of the non-seizure time; microscopically, the characteristic waveforms of epileptic seizures can increase the sensitivity.
Through the comparison of seizure time, the thesis achieved 75.64% sensitivity, 99.81% accuracy, 99.81% specificity, and 0.19% false positive rate, it not only ensures reliability but also provides a quantification analysis for diagnosis. The proposed method has been implemented into a chip using TSMC 0.18 μm process, which proves the feasibility of mobile device to the real-time epilepsy detection system (RED system). The clock frequency is 438 kHz, the power consumption is 2.1678 mW, and the chip area is 3 × 3 mm2.
Through the observation of seizure signals, macroscopically, the trend of epileptic seizures can filter out most of the non-seizure time; microscopically, the characteristic waveforms of epileptic seizures can increase the sensitivity.
Through the comparison of seizure time, the thesis achieved 75.64% sensitivity, 99.81% accuracy, 99.81% specificity, and 0.19% false positive rate, it not only ensures reliability but also provides a quantification analysis for diagnosis. The proposed method has been implemented into a chip using TSMC 0.18 μm process, which proves the feasibility of mobile device to the real-time epilepsy detection system (RED system). The clock frequency is 438 kHz, the power consumption is 2.1678 mW, and the chip area is 3 × 3 mm2.