應用於低雜訊心電圖監測的類比前端感測晶片

Design of Low-Noise Analog Front-End Sensing Chip for Electrocardiography Monitoring Applications

檔案

全文未公開。

日期

2023

類型

語言

英文

摘要

近年來,由於飲食習慣的變化、人口老化以及新冠病毒的廣泛傳播,導致患有心臟問題和心率異常的患者數量不斷增加。 為了鼓勵人們持續監測和記錄自己的健康狀況,研究穿戴式裝置變得至關重要,其中心電圖是最能反應心臟問題的方法。
心電圖訊號擷取系統需要具備多個關鍵特性:可調增益以適應未知大小的心電訊號、足夠低的頻寬以覆蓋心電訊號的頻率範圍,以協助醫生進行病症診斷、高輸入阻抗以防止電極間的不匹配影響訊號擷取,並具備足夠大的共模抑制比以抵禦環境中的共模干擾。
本論文提出了一種低雜訊且低功耗的心電訊號感測電路。 其中,類比前端感測電路是實現高性能心電訊號感測電路的關鍵,該設計採用具有高輸入阻抗和出色共模抑制比特性的儀表放大器作為核心元件,提供了 23~dB的增益。此外,還加入了可調增益放大器,將整體增益範圍從 23~dB 至~65 dB。最後再透過直流伺服迴路、正回授迴路和斬波器分別產生 0.3~Hz 的低頻高通濾波、將輸入阻抗提升至 4.6~M$\Omega$ 以及在頻率範圍 0.5~Hz 至 150~Hz 內將雜訊降低至 2.25~\textmu V$_{rms}$,以確保清晰放大從人體上採集到的心電訊號。
Over the past few years, there has been a surge in patients with heart problems and cardiac arrhythmias due to changes in dietary habits, an aging population, and the widespread impact of COVID-19. Portable devices have become essential in research and healthcare. Among these devices, electrocardiogram (ECG) monitoring is particularly effective in assessing cardiac problems.
An ECG signal acquisition system needs to have an adjustable gain to accommodate varying signal amplitudes and a bandwidth sufficient to cover the frequency range of ECG signals, aiding in diagnosing symptoms. High input impedance is necessary to prevent signal attenuation caused by electrode mismatches. Additionally, a significant common-mode rejection ratio is crucial to eliminate interference from common-mode noise in the environment during signal acquisition.
This thesis presents a low-noise and low-power ECG signal sensing circuit. The analog front-end sensing circuit is crucial for high-performance ECG signal sensing. This design uses a high-performance instrumentation amplifier with high input impedance and a high common-mode rejection ratio as the core component, providing a gain of 23~dB. An adjustable gain amplifier is also incorporated, extending the overall gain range from 23 dB~to~65 dB. Further enhancements are achieved through a DC servo loop, a positive feedback loop, and a chopper. These separately provide a low-frequency high-pass cutoff at 0.3~Hz, increase the input impedance to 4.6~M$\Omega$, and reduce noise to 2.25~\textmu V$_{rms}$ within the frequency range of 0.5 Hz to 150 Hz.
Over the past few years, there has been a surge in patients with heart problems and cardiac arrhythmias due to changes in dietary habits, an aging population, and the widespread impact of COVID-19. Portable devices have become essential in research and healthcare. Among these devices, electrocardiogram (ECG) monitoring is particularly effective in assessing cardiac problems.
An ECG signal acquisition system needs to have an adjustable gain to accommodate varying signal amplitudes and a bandwidth sufficient to cover the frequency range of ECG signals, aiding in diagnosing symptoms. High input impedance is necessary to prevent signal attenuation caused by electrode mismatches. Additionally, a significant common-mode rejection ratio is crucial to eliminate interference from common-mode noise in the environment during signal acquisition.
This thesis presents a low-noise and low-power ECG signal sensing circuit. The analog front-end sensing circuit is crucial for high-performance ECG signal sensing. This design uses a high-performance instrumentation amplifier with high input impedance and a high common-mode rejection ratio as the core component, providing a gain of 23~dB. An adjustable gain amplifier is also incorporated, extending the overall gain range from 23 dB~to~65 dB. Further enhancements are achieved through a DC servo loop, a positive feedback loop, and a chopper. These separately provide a low-frequency high-pass cutoff at 0.3~Hz, increase the input impedance to 4.6~M$\Omega$, and reduce noise to 2.25~\textmu V$_{rms}$ within the frequency range of 0.5 Hz to 150 Hz.
近年來,由於飲食習慣的變化、人口老化以及新冠病毒的廣泛傳播,導致患有心臟問題和心率異常的患者數量不斷增加。 為了鼓勵人們持續監測和記錄自己的健康狀況,研究穿戴式裝置變得至關重要,其中心電圖是最能反應心臟問題的方法。
心電圖訊號擷取系統需要具備多個關鍵特性:可調增益以適應未知大小的心電訊號、足夠低的頻寬以覆蓋心電訊號的頻率範圍,以協助醫生進行病症診斷、高輸入阻抗以防止電極間的不匹配影響訊號擷取,並具備足夠大的共模抑制比以抵禦環境中的共模干擾。
本論文提出了一種低雜訊且低功耗的心電訊號感測電路。 其中,類比前端感測電路是實現高性能心電訊號感測電路的關鍵,該設計採用具有高輸入阻抗和出色共模抑制比特性的儀表放大器作為核心元件,提供了 23~dB的增益。此外,還加入了可調增益放大器,將整體增益範圍從 23~dB 至~65 dB。最後再透過直流伺服迴路、正回授迴路和斬波器分別產生 0.3~Hz 的低頻高通濾波、將輸入阻抗提升至 4.6~M$\Omega$ 以及在頻率範圍 0.5~Hz 至 150~Hz 內將雜訊降低至 2.25~\textmu V$_{rms}$,以確保清晰放大從人體上採集到的心電訊號。

所屬典藏

電機工程學系
顯示完整詮釋資料 統計資料

引用此條目

Liu, W.-C. (2023). Design of Low-Noise Analog Front-End Sensing Chip for Electrocardiography Monitoring Applications. https://thesis.lib.nycu.edu.tw/handle/20.500.14371/847

BibTeX RIS

引用文字依 APA 第 7 版格式自動產生,使用前請再核對。

分享

永久網址:https://thesis.lib.nycu.edu.tw/handle/20.500.14371/847