Accuracy Optimization of Personalized Cochlear Implant Finite Element Models Based on Evoked Compound Action Potentials
基於誘發複合動作電位之個人化人工耳蝸有限元模型精度優化研究
Files
The full text is not publicly available.
Date
2025
Type
Department
Language
英文
Abstract
部分重度聽力損失患者由於耳蝸內部將聲波轉化為電生理信號的絨毛細胞因為各種原因受損,導致無法正常感知聲音。人工耳蝸作為一種植入式醫療裝置,能繞過受損的絨毛細胞,直接以電刺激方式作用於聽覺神經,是目前臨床上有效恢復重度聽力損失患者聽力的治療手段之一。本研究旨在優化基於電誘發複合動作電位(Evoked Compound Action Potentials, ECAP)反應的個人化人工耳蝸有限元素模型(Finite Element Model, FEM),以提升模擬結果對臨床數據的貼合度。
本研究根據實驗室既有研究成果、廠商提供之元件資料及實際臨床測量數據,建構人工耳蝸有限元素模型,並透過調整網格解析度參數,建立兩種不同精度的模型版本:一為約600萬網格節點的標準解析度模型,另一為約 1900 萬網格節點的高解析度模型,並以此兩者進行模擬,取得對應之 ECAP 響應結果,與臨床實測數據進行比對分析。此外,本文亦根據模型中被激活的聽神經數量,預估使用者於各電極位置的最佳舒適度(Most Comfortable Loudness Level, MCL),並亦與臨床量測結果進行比較。
模擬結果顯示,高解析度模型所產生之 ECAP 響應與臨床量測值吻合度較為接近,其所預測之 MCL數值亦與臨床數據呈現更高的一致性。此結果顯示,提升模型之網格精度有助於更準確地模擬個體聽覺神經的電刺激反應,為人工耳蝸模擬模型之個人化優化與臨床應用提供潛在依據。
本研究根據實驗室既有研究成果、廠商提供之元件資料及實際臨床測量數據,建構人工耳蝸有限元素模型,並透過調整網格解析度參數,建立兩種不同精度的模型版本:一為約600萬網格節點的標準解析度模型,另一為約 1900 萬網格節點的高解析度模型,並以此兩者進行模擬,取得對應之 ECAP 響應結果,與臨床實測數據進行比對分析。此外,本文亦根據模型中被激活的聽神經數量,預估使用者於各電極位置的最佳舒適度(Most Comfortable Loudness Level, MCL),並亦與臨床量測結果進行比較。
模擬結果顯示,高解析度模型所產生之 ECAP 響應與臨床量測值吻合度較為接近,其所預測之 MCL數值亦與臨床數據呈現更高的一致性。此結果顯示,提升模型之網格精度有助於更準確地模擬個體聽覺神經的電刺激反應,為人工耳蝸模擬模型之個人化優化與臨床應用提供潛在依據。
Some patients with severe hearing loss are unable to perceive sound properly due to damage in the hair cells of the cochlea, which are responsible for converting sound vibrations into electrophysiological signals. A cochlear implant, as an implantable medical device, bypasses these damaged hair cells and directly stimulates the auditory nerve using electrical signals. It has been clinically proven to be one of the most effective treatments for restoring hearing in patients with profound hearing loss. This study aims to optimize personalized cochlear implant finite element models (FEM) based on responses from evoked compound action potentials (ECAP), in order to improve the alignment between simulation results and clinical data.
The cochlear implant FEMs were constructed based on prior research from our laboratory, component specifications provided by the manufacturer, and actual clinical measurement data. Two model versions with different mesh resolutions were developed by adjusting mesh parameters: one with approximately 6 million elements representing the standard-resolution model, and the other with around 19 million elements representing the high-resolution model. Simulations were performed on both versions to obtain corresponding ECAP responses, which were then compared against clinical measurements. In addition, the number of activated auditory nerve fibers in the models was used to estimate the most comfortable loudness level (MCL) at each electrode position, and these predictions were also compared to clinically measured values.
Simulation results show that the ECAP responses generated by the high-resolution model exhibit closer agreement with clinical measurements, and the predicted MCL values demonstrate greater consistency with clinical data. These findings indicate that improving mesh resolution enhances the accuracy of simulating individual auditory nerve responses to electrical stimulation, providing a valuable foundation for the personalized optimization and clinical application of cochlear implant modeling.
The cochlear implant FEMs were constructed based on prior research from our laboratory, component specifications provided by the manufacturer, and actual clinical measurement data. Two model versions with different mesh resolutions were developed by adjusting mesh parameters: one with approximately 6 million elements representing the standard-resolution model, and the other with around 19 million elements representing the high-resolution model. Simulations were performed on both versions to obtain corresponding ECAP responses, which were then compared against clinical measurements. In addition, the number of activated auditory nerve fibers in the models was used to estimate the most comfortable loudness level (MCL) at each electrode position, and these predictions were also compared to clinically measured values.
Simulation results show that the ECAP responses generated by the high-resolution model exhibit closer agreement with clinical measurements, and the predicted MCL values demonstrate greater consistency with clinical data. These findings indicate that improving mesh resolution enhances the accuracy of simulating individual auditory nerve responses to electrical stimulation, providing a valuable foundation for the personalized optimization and clinical application of cochlear implant modeling.
Collections
Institute of Biomedical EngineeringCite this item
Wang, Chien-Chun (Alexander Wang) (2025). Accuracy Optimization of Personalized Cochlear Implant Finite Element Models Based on Evoked Compound Action Potentials. https://thesis.lib.nycu.edu.tw/handle/20.500.14371/87014
Citation generated in APA 7th edition style; please check it before use.
Share
Permanent link: https://thesis.lib.nycu.edu.tw/handle/20.500.14371/87014