客製化彈性下顎彌補物:開發與FDA上市前評估
Customized Elastic Mandibular Prosthesis: Development and Pre-market Evaluation with FDA
摘要
口腔鱗狀細胞癌約占所有口腔惡性腫瘤之90%。腫瘤易侵入骨組織,臨床上常透過節段性切除移除病灶,進而造成下顎骨缺損。下顎骨缺損之重建對於患者外觀、咀嚼功能與生活品質具有重要影響。現行以腓骨游離皮瓣為主之重建方式,仍面臨骨不癒合、植入物失效及術後併發症等問題。現有客製化重建板多著重於固定功能,較少考量缺損間隙之彌補設計,可能影響骨癒合效果。此外,醫療器材開發需符合嚴格監管要求,若能於設計初期納入上市策略與風險管理,有助於提升產品開發效率與臨床應用可行性。
本研究建立一套包含開發與上市前評估的系統化流程,並同步開發客製化彈性下顎彌補物,其結合具結構支撐功能的鈦合金重建板,以及可彌補移植間隙並提供適合骨生長環境的熱塑性聚氨脂彈性支架。依據 ISO 14971 規範進行風險管理,針對器材潛在缺陷、手術操作及患者生理因素共14項相關風險進行評估,並透過有限元素模擬驗證設計規格,以及循環負載測試進行使用者需求之確效,最終結合比對品分析評估其法規上市可行性。
研究結果顯示,彈性支架於壓縮變形後,其孔徑維持於 0.91 至 1.12 mm,符合骨細胞生長之建議範圍 (0.30 至1.20 mm);力學測試中單次最大位移振幅為 0.1659 mm,約為骨癒合最大允許範圍 (0.5 mm) 之 33.2%,顯示支架可提供適合骨細胞生長之穩定環境。此外,重建板可承受成年人總和最大咬合力380 N與單顆臼齒最大咬力159 N,出現最大應力為 264.29 MPa,約為材料容許強度 (723 MPa) 之 36.6%,且於極端力學循環測試中未觀察到結構失效或脫位情形。風險管理結果顯示,設備與操作相關風險等級均下降一個等級,多數殘餘風險已降低至可接受範圍;而患者生理因素相關風險,則提出促進生物相容性之設計與測試方向,以作為後續驗證依據。
上市前評估方面,透過比對品分析確認產品於預期用途與技術特徵上與既有醫療器材具有相似性,具備進行實質等效性評估之基礎,並完成 FDA 510(k) 上市途徑之可行性評估。然而,針對支架採用新材料所涉及之生物相容性評估,以及滅菌確效與標準化力學測試等關鍵項目,仍需進一步補充相關驗證資料,以符合醫療器材法規要求。此外,未來若於材料、結構設計或使用條件等面向產生變更,且可能影響產品風險時,仍需重新檢視並更新相關風險評估與設計驗證。
本研究成功建立一套客製化下顎彌補物之開發與上市前評估系統化流程,並透過3D列印技術製造出彈性下顎彌補物。此流程可作為未來醫療器材開發之參考架構,有助於整合工程設計與法規要求,並提升產品之臨床應用潛力。
本研究建立一套包含開發與上市前評估的系統化流程,並同步開發客製化彈性下顎彌補物,其結合具結構支撐功能的鈦合金重建板,以及可彌補移植間隙並提供適合骨生長環境的熱塑性聚氨脂彈性支架。依據 ISO 14971 規範進行風險管理,針對器材潛在缺陷、手術操作及患者生理因素共14項相關風險進行評估,並透過有限元素模擬驗證設計規格,以及循環負載測試進行使用者需求之確效,最終結合比對品分析評估其法規上市可行性。
研究結果顯示,彈性支架於壓縮變形後,其孔徑維持於 0.91 至 1.12 mm,符合骨細胞生長之建議範圍 (0.30 至1.20 mm);力學測試中單次最大位移振幅為 0.1659 mm,約為骨癒合最大允許範圍 (0.5 mm) 之 33.2%,顯示支架可提供適合骨細胞生長之穩定環境。此外,重建板可承受成年人總和最大咬合力380 N與單顆臼齒最大咬力159 N,出現最大應力為 264.29 MPa,約為材料容許強度 (723 MPa) 之 36.6%,且於極端力學循環測試中未觀察到結構失效或脫位情形。風險管理結果顯示,設備與操作相關風險等級均下降一個等級,多數殘餘風險已降低至可接受範圍;而患者生理因素相關風險,則提出促進生物相容性之設計與測試方向,以作為後續驗證依據。
上市前評估方面,透過比對品分析確認產品於預期用途與技術特徵上與既有醫療器材具有相似性,具備進行實質等效性評估之基礎,並完成 FDA 510(k) 上市途徑之可行性評估。然而,針對支架採用新材料所涉及之生物相容性評估,以及滅菌確效與標準化力學測試等關鍵項目,仍需進一步補充相關驗證資料,以符合醫療器材法規要求。此外,未來若於材料、結構設計或使用條件等面向產生變更,且可能影響產品風險時,仍需重新檢視並更新相關風險評估與設計驗證。
本研究成功建立一套客製化下顎彌補物之開發與上市前評估系統化流程,並透過3D列印技術製造出彈性下顎彌補物。此流程可作為未來醫療器材開發之參考架構,有助於整合工程設計與法規要求,並提升產品之臨床應用潛力。
Oral squamous cell carcinoma accounts for approximately 90% of all oral malignancies. Due to tumor invasion into bone tissue, segmental resection is often required, resulting in mandibular defects. Reconstruction of mandibular defects is critical for restoring facial appearance, masticatory function, and overall quality of life. However, the current gold standard approach, fibula free flap reconstruction, is still associated with complications such as nonunion, implant failure, and postoperative morbidity. Existing patient-specific reconstruction plates primarily focus on fixation, with limited consideration of defect gap compensation, which may adversely affect bone healing. In addition, medical device development is subject to stringent regulatory requirements; therefore, integrating regulatory strategies and risk management during the early design stage may improve development efficiency and clinical feasibility.
In this study, a systematic workflow integrating device development and premarket evaluation was established, and a patient-specific elastic mandibular prosthesis was developed. The proposed design combines a load-bearing titanium reconstruction plate with a thermoplastic polyurethane (TPU) elastic scaffold to compensate for defect gaps and provide a favorable environment for bone regeneration. Risk management was conducted in accordance with ISO 14971, with a total of 14 potential risks identified and evaluated, covering device-related, surgical, and patient-related factors. Design verification was performed using finite element analysis, and validation was conducted through cyclic loading tests under simulated use conditions. In addition, predicate device analysis was carried out to assess regulatory feasibility.
The results demonstrated that the scaffold pore size remained within 0.91–1.12 mm after compression, which falls within the recommended range for bone ingrowth (0.3–1.2 mm). The maximum displacement amplitude under cyclic loading was 0.1659 mm, corresponding to 33.2% of the allowable limit for bone healing (0.5 mm), indicating that the scaffold provides a mechanically stable environment for bone regeneration. Furthermore, the reconstruction plate was able to withstand a maximum occlusal load of 380 N and a single-tooth load of 159 N, with a maximum stress of 264.29 MPa, corresponding to 36.6% of the material’s allowable strength (723 MPa). No structural failure or displacement was observed during extreme cyclic loading tests, demonstrating adequate mechanical safety. Risk management results showed that device- and surgical-related risks were reduced by at least one level, with most residual risks considered acceptable. For patient-related risks, strategies for improving biocompatibility were proposed for future validation.
Regarding premarket evaluation, predicate device analysis indicated that the proposed device shares similar intended use and technological characteristics with existing medical devices, supporting the feasibility of substantial equivalence assessment. A preliminary evaluation of the FDA 510(k) pathway was also conducted. However, additional verification, including biocompatibility testing of the scaffold material, sterilization validation, and standardized mechanical testing, is still required to meet regulatory requirements. Furthermore, future modifications in materials, structural design, or usage conditions that may affect product risk would require re-evaluation and updates to the relevant risk assessment and design verification.
In conclusion, this study established a systematic workflow for the development and premarket evaluation of a patient-specific mandibular prosthesis and successfully fabricated an elastic mandibular implant using additive manufacturing. The proposed framework provides a reference for future medical device development, facilitating the integration of engineering design and regulatory requirements, and enhancing clinical translation potential.
In this study, a systematic workflow integrating device development and premarket evaluation was established, and a patient-specific elastic mandibular prosthesis was developed. The proposed design combines a load-bearing titanium reconstruction plate with a thermoplastic polyurethane (TPU) elastic scaffold to compensate for defect gaps and provide a favorable environment for bone regeneration. Risk management was conducted in accordance with ISO 14971, with a total of 14 potential risks identified and evaluated, covering device-related, surgical, and patient-related factors. Design verification was performed using finite element analysis, and validation was conducted through cyclic loading tests under simulated use conditions. In addition, predicate device analysis was carried out to assess regulatory feasibility.
The results demonstrated that the scaffold pore size remained within 0.91–1.12 mm after compression, which falls within the recommended range for bone ingrowth (0.3–1.2 mm). The maximum displacement amplitude under cyclic loading was 0.1659 mm, corresponding to 33.2% of the allowable limit for bone healing (0.5 mm), indicating that the scaffold provides a mechanically stable environment for bone regeneration. Furthermore, the reconstruction plate was able to withstand a maximum occlusal load of 380 N and a single-tooth load of 159 N, with a maximum stress of 264.29 MPa, corresponding to 36.6% of the material’s allowable strength (723 MPa). No structural failure or displacement was observed during extreme cyclic loading tests, demonstrating adequate mechanical safety. Risk management results showed that device- and surgical-related risks were reduced by at least one level, with most residual risks considered acceptable. For patient-related risks, strategies for improving biocompatibility were proposed for future validation.
Regarding premarket evaluation, predicate device analysis indicated that the proposed device shares similar intended use and technological characteristics with existing medical devices, supporting the feasibility of substantial equivalence assessment. A preliminary evaluation of the FDA 510(k) pathway was also conducted. However, additional verification, including biocompatibility testing of the scaffold material, sterilization validation, and standardized mechanical testing, is still required to meet regulatory requirements. Furthermore, future modifications in materials, structural design, or usage conditions that may affect product risk would require re-evaluation and updates to the relevant risk assessment and design verification.
In conclusion, this study established a systematic workflow for the development and premarket evaluation of a patient-specific mandibular prosthesis and successfully fabricated an elastic mandibular implant using additive manufacturing. The proposed framework provides a reference for future medical device development, facilitating the integration of engineering design and regulatory requirements, and enhancing clinical translation potential.